Radiation exposure in full-field digital mammography with a flat-panel x-ray detector based on amorphous silicon in comparison with conventional screen-film mammography.

被引:18
|
作者
Hermann, KP [1 ]
Obenauer, S [1 ]
Grabbe, E [1 ]
机构
[1] Univ Gottingen Klinikum, Zentrum Radiol, Abt Rontgendiagnost 1, D-37075 Gottingen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2000年 / 172卷 / 11期
关键词
mammography; digital mammography; flat panel detector; radiation exposure; image quality;
D O I
10.1055/s-2000-8370
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Comparison of radiation exposure between a digital amorphous silicon and a screen-film based mammography system. Evaluation of a possible potential of full-field digital mammography in order to decrease the radiation dose. Methods: The average glandular dose for phantom thicknesses from 30 to 60 mm was calculated from experimentally determined entrance surface air kerma for a digital and a conventional mammography system. The effect of reducing the detector dose and of changing the radiation quality on radiation exposure and on image quality were investigated. Results: By using the delivered settings of the automatic exposure control (AEC) devices, both mammographic systems needed nearly the same doses. Regulations and guidelines on radiation doses were complied. With the digital system, a reduction of radiation exposure of up to 40% by using a higher radiation quality and decreasing slightly the detector dose without loss of diagnostic image quality, might be possible. Conclusion: The potential of fullfield digital mammography for radiation dose reduction, as shown in the present phantom study, needs however, a careful examination under clinical conditions.
引用
收藏
页码:940 / 945
页数:6
相关论文
共 50 条
  • [1] Radiation exposure in full-field digital mammography with a selenium flat-panel detector
    Gosch, D.
    Jendrass, S.
    Scholz, M.
    Kahn, T.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2006, 178 (07): : 693 - 697
  • [2] Full-field digital mammography in comparison to screen-film mammography - A phantom study
    Obenauer, S
    Schorn, C
    Funke, M
    Fischer, U
    Grabbe, E
    Hermann, KP
    IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 499 - 503
  • [3] Clinical performance comparison of full-field digital mammography to screen-film mammography
    Berns, EA
    Hendrick, RE
    RADIOLOGY, 2001, 221 : 283 - 283
  • [4] Comparison of full-field digital and screen-film mammography dose
    Schueler, B.
    MEDICAL PHYSICS, 2006, 33 (06) : 2005 - 2005
  • [5] Characterization of a direct, full-field flat-panel digital mammography detector
    Zhao, B
    Zhao, W
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 157 - 167
  • [6] Performance comparison of full-field digital mammography to screen-film mammography in clinical practice
    Berns, EA
    Hendrick, RE
    Cutter, GR
    MEDICAL PHYSICS, 2002, 29 (05) : 830 - 834
  • [7] Clinical comparison of full-field digital mammography and screen-film mammography of breast cancer
    Lewin, JM
    D'Orsi, CJ
    Hendrick, RE
    Moss, LJ
    Isaacs, PK
    Karellas, A
    Cutter, GR
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (03) : 671 - 677
  • [8] Breast mass classification on full-field digital mammography and screen-film mammography
    Shi, Jiazheng
    Sahiner, Berkman
    Chan, Heang-Ping
    Hadjiiski, Lubomir M.
    Ge, Jun
    Wei, Jun
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2008, 5116 : 371 - 377
  • [9] Detection of simulated chest lesions with normal and reduced radiation dose -: Comparison of conventional screen-film radiography and a flat-panel x-ray detector based on amorphous silicon
    Strotzer, M
    Gmeinwieser, JK
    Völk, M
    Fründ, R
    Seitz, J
    Feuerbach, S
    INVESTIGATIVE RADIOLOGY, 1998, 33 (02) : 98 - 103
  • [10] Screen-film mammography and soft-copy full-field digital mammography: Comparison in the patients with microcalcifications
    Kim, HS
    Han, BK
    Choo, KS
    Jeon, YH
    Kim, JH
    Choe, YH
    KOREAN JOURNAL OF RADIOLOGY, 2005, 6 (04) : 214 - 220