Efficacy of establishment of pneumoperitoneum with the Veress needle, Hasson trocar, and modified blunt trocar (TrocDoc): A randomized study

被引:41
作者
Bemelman, WA [1 ]
Dunker, MS [1 ]
Busch, ORC [1 ]
Den Boer, KT [1 ]
De Wit, LT [1 ]
Gouma, DJ [1 ]
机构
[1] Acad Med Ctr, Dept Surg, NL-1100 DD Amsterdam, Netherlands
来源
JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES-PART A | 2000年 / 10卷 / 06期
关键词
D O I
10.1089/lap.2000.10.325
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To compare three techniques of establishment of pneumoperitoneum for efficacy: the Veress needle/first trocar, the Hasson trocar, and a newly developed modified blunt trocar, the TrocDoc. Patients and Methods: Between June and December 1999, 62 patients eligible for laparoscopic surgery were randomized. The effectiveness of installation of the pneumoperitoneum using the three techniques was assessed by time-motion analysis. Primary efficacy measures were total time and number of actions required to establish the pneumoperitoneum. Secondary efficacy measures were procedure-related complications, wound complications, and occurrence of CO2 leakage. Results: Two patients were withdrawn from inclusion. The three groups were comparable for age and body mass index. Total time was shortest using the TrocDoc rather than the Veress needle/first trocar and the Hasson trocar (respectively, 138 +/- 58 v 237 +/- 56 v 350 +/- 103 seconds), and the number of actions was lowest for the Veress needle/first trocar combination: 22 +/- 7 v 32 +/- 12 (TrocDoc) v 53 +/- 17 (Hasson). There was no morbidity related to the installation of pneumoperitoneum nor trocar wound complications. Gas leakage occurred in five of the Hasson introductions. Conclusions: Establishment of the pneumoperitoneum is more efficient using the TrocDoc compared with the Veress needle/first trocar and the Hasson trocar. The TrocDoc might replace the two alternatives because of its efficacy and open method of introduction.
引用
收藏
页码:325 / 330
页数:6
相关论文
共 20 条
  • [1] Establishment of pneumoperitoneum with a modified blunt trocar
    Bemelman, WA
    De Wit, LT
    Busch, ORC
    Den Boer, KT
    Klaase, JFM
    Grimbergen, CA
    Gouma, DJ
    [J]. JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES-PART A, 2000, 10 (04): : 217 - 218
  • [2] Open versus closed establishment of pneumoperitoneum in laparoscopic surgery
    Bonjer, HJ
    Hazebroek, EJ
    Kazemier, G
    Giuffrida, MC
    Meijer, WS
    Lange, JF
    [J]. BRITISH JOURNAL OF SURGERY, 1997, 84 (05) : 599 - 602
  • [3] Serious trocar accidents in laparoscopic surgery: A French survey of 103,852 operations
    Champault, G
    Cazacu, F
    Taffinder, N
    [J]. SURGICAL LAPAROSCOPY & ENDOSCOPY, 1996, 6 (05) : 367 - 370
  • [4] Quantitative analysis of the functionality and efficiency of three surgical dissection techniques: A time-motion analysis
    Den Boer, KT
    Straatsburg, IH
    Schellinger, AV
    De Wit, LT
    Dankelman, J
    Gouma, DJ
    [J]. JOURNAL OF LAPAROENDOSCOPIC & ADVANCED SURGICAL TECHNIQUES-PART A, 1999, 9 (05): : 389 - 395
  • [5] Peroperative time-motion analysis of diagnostic laparoscopy with laparoscopic ultrasonography
    den Boer, KT
    de Wit, LT
    Dankelman, J
    Gouma, DJ
    [J]. BRITISH JOURNAL OF SURGERY, 1999, 86 (07) : 951 - 955
  • [6] Reduction of cannula-related laparoscopic complications using a radially expanding access device
    Galen, DI
    Jacobson, A
    Weckstein, LN
    Kaplan, RA
    DeNevi, KL
    [J]. JOURNAL OF THE AMERICAN ASSOCIATION OF GYNECOLOGIC LAPAROSCOPISTS, 1999, 6 (01): : 79 - 84
  • [7] Hanney RM, 1999, J AM COLL SURGEONS, V188, P337
  • [8] Needle and trocar injury during laparoscopic surgery in Japan
    Hashizume, M
    Sugimachi, K
    [J]. SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 1997, 11 (12): : 1198 - 1201
  • [9] MODIFIED INSTRUMENT AND METHOD FOR LAPAROSCOPY
    HASSAON, HM
    [J]. AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1971, 110 (06) : 886 - &
  • [10] Kockerling F, 1996, CHIRURG, V67, P183