Rhodium self-powered neutron detector as a suitable on-line thermal neutron flux monitor in BNCT treatments

被引:4
|
作者
Miller, Marcelo E. [1 ]
Sztejnberg, Manuel L. [1 ]
Gonzalez, Sara J. [1 ,2 ]
Thorp, Silvia I. [1 ]
Longhino, Juan M. [1 ]
Estryk, Guillermo [1 ]
机构
[1] Comis Nacl Energia Atom, RA-1429 Buenos Aires, DF, Argentina
[2] Consejo Nacl Invest Cient & Tecn, RA-1033 Buenos Aires, DF, Argentina
关键词
SPND; BNCT; on-line flux measurement; DOSIMETRY; REACTOR;
D O I
10.1118/1.3660204
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: A rhodium self-powered neutron detector (Rh SPND) has been specifically developed by the Comision Nacional de Energia Atomica (CNEA) of Argentina to measure locally and in real time thermal neutron fluxes in patients treated with boron neutron capture therapy (BNCT). In this work, the thermal and epithermal neutron response of the Rh SPND was evaluated by studying the detector response to two different reactor spectra. In addition, during clinical trials of the BNCT Project of the CNEA, on-line neutron flux measurements using the specially designed detector were assessed. Methods: The first calibration of the detector was done with the well-thermalized neutron spectrum of the CNEA RA-3 reactor thermal column. For this purpose, the reactor spectrum was approximated by a Maxwell-Boltzmann distribution in the thermal energy range. The second calibration was done at different positions along the central axis of a water-filled cylindrical phantom, placed in the mixed thermal-epithermal neutron beam of CNEA RA-6 reactor. In this latter case, the RA-6 neutron spectrum had been well characterized by both calculation and measurement, and it presented some marked differences with the ideal spectrum considered for SPND calibrations at RA-3. In addition, the RA-6 neutron spectrum varied with depth in the water phantom and thus the percentage of the epithermal contribution to the total neutron flux changed at each measurement location. Local (one point-position) and global (several points-positions) and thermal and mixed-field thermal neutron sensitivities were determined from these measurements. Thermal neutron flux was also measured during BNCT clinical trials within the irradiation fields incident on the patients. In order to achieve this, the detector was placed on patient's skin at dosimetric reference points for each one of the fields. System stability was adequate for this kind of measurement. Results: Local mixed-field thermal neutron sensitivities and global thermal and mixed-field thermal neutron sensitivities derived from measurements performed at the RA-6 were compared and no significant differences were found. Global RA-6-based thermal neutron sensitivity showed agreement with pure thermal neutron sensitivity measurements performed in the RA-3 spectrum. Additionally, the detector response proved nearly unchanged by differences in neutron spectra from real (RA-6 BNCT beam) and ideal (considered for calibration calculations at RA-3) neutron source descriptions. The results confirm that the special design of the Rh SPND can be considered as having a pure thermal response for neutron spectra with epithermal-to-thermal flux ratios up to 12%. In addition, the linear response of the detector to thermal flux allows the use of a mixed-field thermal neutron sensitivity of 1.95 +/- 0.05 x 10(-21) A n(-1).cm(2).s. This sensitivity can be used in spectra with up to 21% epithermal-to-thermal flux ratio without significant error due to epithermal neutron and gamma induced effects. The values of the measured fluxes in clinical applications had discrepancies with calculated results that were in the range of -25% to +30%, which shows the importance of a local on-line independent measurement as part of a treatment planning quality control system. Conclusions: The usefulness of the CNEA Rh SPND for the on-line local measurement of thermal neutron flux on BNCT patients has been demonstrated based on an appropriate neutron spectra calibration and clinical applications. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3660204]
引用
收藏
页码:6502 / 6512
页数:11
相关论文
共 50 条
  • [1] Comparison of Thermal Neutron Flux Measured by Uranium 235 Fission Chamber and Rhodium Self-Powered Neutron Detector in MTR
    Fourmentel, D.
    Filliatre, P.
    Barbot, L.
    Villard, J. F.
    Lyoussi, A.
    Geslot, B.
    Carcreff, H.
    Malo, J. Y.
    Reynard-Carette, C.
    2013 3RD INTERNATIONAL CONFERENCE ON ADVANCEMENTS IN NUCLEAR INSTRUMENTATION, MEASUREMENT METHODS AND THEIR APPLICATIONS (ANIMMA), 2013,
  • [2] Numerical optimization of Rhodium Self-Powered Neutron Detector
    Zhang, Qingmin
    Liu, Xinxin
    Deng, Bangjie
    Cao, Liangzhi
    Tang, Chuntao
    ANNALS OF NUCLEAR ENERGY, 2018, 113 : 519 - 525
  • [3] DETERMINATION OF NEUTRON-FLUX AND NEUTRON FLUENCE FROM SIGNAL CURRENT OF A SELF-POWERED RHODIUM DETECTOR
    HUDRITSCH, WW
    NUCLEAR TECHNOLOGY, 1973, 18 (01) : 25 - 28
  • [4] Implantable self-powered detector for on-line determination of neutron flux in patients during NCT treatment
    Miller, ME
    Mariani, LE
    Gonçalves-Carralves, MLS
    Skumanic, M
    Thorp, SI
    APPLIED RADIATION AND ISOTOPES, 2004, 61 (05) : 1033 - 1037
  • [5] SELF-POWERED DETECTOR RESPONSE TO THERMAL AND EPITHERMAL NEUTRON-FLUX
    AGU, MN
    PETITCOLAS, H
    NUCLEAR SCIENCE AND ENGINEERING, 1991, 107 (04) : 374 - 384
  • [6] Self-powered neutron detector qualification for absolute on-line in-pile neutron flux measurements in BR2
    Vermeeren, L
    Wéber, M
    REACTOR DOSIMETRY IN THE 21ST CENTURY, 2003, : 219 - 225
  • [7] Comparison of Thermal Neutron Flux Measured by 235U Fission Chamber and Rhodium Self-Powered Neutron Detector in MTR
    Fourmentel, D.
    Filliatre, P.
    Barbot, L.
    Villard, J. F.
    Lyoussi, A.
    Geslot, B.
    Carcreff, H.
    Malo, J. Y.
    Reynard-Carette, C.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2014, 61 (04) : 2285 - 2290
  • [8] THE DETERMINATION OF SELF-POWERED NEUTRON DETECTOR SENSITIVITY ON THERMAL AND EPITHERMAL NEUTRON-FLUX DENSITIES
    ERBEN, O
    JADERNA ENERGIE, 1980, 26 (02): : 47 - 52
  • [9] INSTANTANEOUS FLUX MEASUREMENTS USING THE BACKGROUND SIGNAL OF THE RHODIUM SELF-POWERED NEUTRON DETECTOR
    GURU, SV
    WEHE, DK
    ANNALS OF NUCLEAR ENERGY, 1992, 19 (04) : 203 - 215
  • [10] VVER Thermal Power Measured Using the Readings of Rhodium Self-powered Neutron Detector (SPND)
    A. Yu. Kurchenkov
    A. I. Kovel
    V. A. Mil’to
    N. V. Mil’to
    D. N. Skorohodov
    N. V. Lipin
    D. V. Vorob’eva
    V. M. Chapaev
    V. A. Khvatov
    Physics of Atomic Nuclei, 2019, 82 : 1117 - 1125