PET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner

被引:46
作者
Omidvari, Negar [1 ]
Cabello, Jorge [1 ]
Topping, Geoffrey [1 ]
Schneider, Florian R. [1 ,4 ]
Paul, Stephan [2 ]
Schwaiger, Markus [1 ]
Ziegler, Sibylle I. [1 ,3 ]
机构
[1] Tech Univ Munich, Dept Nucl Med, Klinikum Rechts Isar, Ismaninger Str 22, D-81675 Munich, Germany
[2] Tech Univ Munich, Phys Dept E18, James Franck Str 1, D-85748 Garching, Germany
[3] Univ Hosp LMU Munich, Dept Nucl Med, Marchioninistr 15, D-81377 Munich, Germany
[4] KETEK GmbH, Hofer Str 3, D-81737 Munich, Germany
关键词
positron emission tomography (PET); PET/MRI; preclinical PET; SiPM; performance evaluation; IMAGE-RECONSTRUCTION; ITERATIVE RECONSTRUCTION; PRECLINICAL TOMOGRAPH; SPATIAL-RESOLUTION; PET/MRI SYSTEM; NEMA; MATRIX; SENSITIVITY; STANDARDS; EMISSION;
D O I
10.1088/1361-6560/aa910d
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
MADPET4 is the first small animal PET insert with two layers of individually read out crystals in combination with silicon photomultiplier technology. It has a novel detector arrangement, in which all crystals face the center of field of view transaxially. In this work, the PET performance of MADPET4 was evaluated and compared to other preclinical PET scanners using the NEMA NU 4 measurements, followed by imaging a mouse-size hot-rod resolution phantom and two in vivo simultaneous PET/MRI scans in a 7 T MRI scanner. The insert had a peak sensitivity of 0.49%, using an energy threshold of 350 keV. A uniform transaxial resolution was obtained up to 15 mm radial offset from the axial center, using filtered back-projection with single-slice rebinning. The measured average radial and tangential resolutions (FWHM) were 1.38 mm and 1.39 mm, respectively. The 1.2 mm rods were separable in the hot-rod phantom using an iterative image reconstruction algorithm. The scatter fraction was 7.3% and peak noise equivalent count rate was 15.5 kcps at 65.1 MBq of activity. The FDG uptake in a mouse heart and brain were visible in the two in vivo simultaneous PET/MRI scans without applying image corrections. In conclusion, the insert demonstrated a good overall performance and can be used for small animal multi-modal research applications.
引用
收藏
页码:8671 / 8692
页数:22
相关论文
共 50 条
[41]   Performance evaluation of the microPET R4 PET scanner for rodents [J].
Knoess, C ;
Siegel, S ;
Smith, A ;
Newport, D ;
Richerzhagen, R ;
Winkeler, A ;
Jacobs, A ;
Goble, RN ;
Graf, R ;
Wienhard, K ;
Heiss, WD .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (05) :737-747
[42]   Attenuation correction for the NIH ATLAS small animal PET scanner [J].
Yao, RT ;
Seidel, J ;
Liow, JS ;
Green, MV .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (03) :664-668
[43]   Development of a small animal PET scanner using DOI detectors [J].
Yamada, R. ;
Watanabe, M. ;
Omura, T. ;
Sato, N. ;
Shimizu, K. ;
Takahashi, M. ;
Ote, K. ;
Katabe, A. ;
Moriya, T. ;
Sakai, K. ;
Yamashita, T. ;
Tanaka, E. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2008, 55 (03) :906-911
[44]   Raytest ClearPET™, a new generation small animal PET scanner [J].
Roldan, P. Sempere ;
Chereul, E. ;
Dietzel, O. ;
Magnier, L. ;
Pautrot, C. ;
Rbah, L. ;
Sappey-Marinier, D. ;
Wagner, A. ;
Zimmer, L. ;
Janier, M. ;
Tarazona, V. ;
Dietzel, G. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) :498-501
[45]   Temperature Dependence of the LabPET Small-Animal PET Scanner [J].
Keereman, Vincent ;
Vanhove, Christian ;
Van Holen, Roel ;
Vandenberghe, Stefaan .
2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, :2896-2901
[46]   A Preclinical PET/MR Insert for a Human 3T MR Scanner [J].
Schulz, Volkmar ;
Solf, Torsten ;
Weissler, Bjoern ;
Gebhardt, Pierre ;
Fischer, Peter ;
Ritzert, Michael ;
Mlotok, Viacheslav ;
Piemonte, Claudio ;
Zorzi, Nicola ;
Melchiorri, Mirko ;
Vandenberghe, Stefaan ;
Keereman, Vincent ;
Schaeffter, Tobias ;
Marsden, Paul K. .
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2009, :2577-+
[47]   Performance evaluation of the Trans-PET® BioCaliburn® LH system: a large FOV small-animal PET system [J].
Wang, Luyao ;
Zhu, Jun ;
Liang, Xiao ;
Niu, Ming ;
Wu, Xiaoke ;
Kao, Chien-Min ;
Kim, Heejong ;
Xie, Qingguo .
PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (01) :137-150
[48]   Performance evaluation of a four-layer LSO detector for a small animal DOI PET scanner: jPET-RD [J].
Tsuda, T ;
Murayama, H ;
Kitamura, K ;
Inadama, N ;
Yamaya, T ;
Yoshida, E ;
Nishikido, F ;
Hamamoto, M ;
Kawai, H ;
Ono, Y .
2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, :3321-3324
[49]   Performance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4-2008 Standards [J].
Miyake, Kanae K. ;
Matsumoto, Keiichi ;
Inoue, Mika ;
Nakamoto, Yuji ;
Kanao, Shotaro ;
Oishi, Tae ;
Kawase, Shigeto ;
Kitamura, Keishi ;
Yamakawa, Yoshiyuki ;
Akazawa, Ayako ;
Kobayashi, Tetsuya ;
Ohi, Junichi ;
Togashi, Kaori .
JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (07) :1198-1203
[50]   NEMA NU-04-based Performance Characteristics of the LabPET-8™ Small Animal PET Scanner [J].
Prasad, Rameshwar ;
Ratib, Osman ;
Zaidi, Habib .
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, :3654-3660