PET performance evaluation of MADPET4: a small animal PET insert for a 7 T MRI scanner

被引:46
作者
Omidvari, Negar [1 ]
Cabello, Jorge [1 ]
Topping, Geoffrey [1 ]
Schneider, Florian R. [1 ,4 ]
Paul, Stephan [2 ]
Schwaiger, Markus [1 ]
Ziegler, Sibylle I. [1 ,3 ]
机构
[1] Tech Univ Munich, Dept Nucl Med, Klinikum Rechts Isar, Ismaninger Str 22, D-81675 Munich, Germany
[2] Tech Univ Munich, Phys Dept E18, James Franck Str 1, D-85748 Garching, Germany
[3] Univ Hosp LMU Munich, Dept Nucl Med, Marchioninistr 15, D-81377 Munich, Germany
[4] KETEK GmbH, Hofer Str 3, D-81737 Munich, Germany
关键词
positron emission tomography (PET); PET/MRI; preclinical PET; SiPM; performance evaluation; IMAGE-RECONSTRUCTION; ITERATIVE RECONSTRUCTION; PRECLINICAL TOMOGRAPH; SPATIAL-RESOLUTION; PET/MRI SYSTEM; NEMA; MATRIX; SENSITIVITY; STANDARDS; EMISSION;
D O I
10.1088/1361-6560/aa910d
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
MADPET4 is the first small animal PET insert with two layers of individually read out crystals in combination with silicon photomultiplier technology. It has a novel detector arrangement, in which all crystals face the center of field of view transaxially. In this work, the PET performance of MADPET4 was evaluated and compared to other preclinical PET scanners using the NEMA NU 4 measurements, followed by imaging a mouse-size hot-rod resolution phantom and two in vivo simultaneous PET/MRI scans in a 7 T MRI scanner. The insert had a peak sensitivity of 0.49%, using an energy threshold of 350 keV. A uniform transaxial resolution was obtained up to 15 mm radial offset from the axial center, using filtered back-projection with single-slice rebinning. The measured average radial and tangential resolutions (FWHM) were 1.38 mm and 1.39 mm, respectively. The 1.2 mm rods were separable in the hot-rod phantom using an iterative image reconstruction algorithm. The scatter fraction was 7.3% and peak noise equivalent count rate was 15.5 kcps at 65.1 MBq of activity. The FDG uptake in a mouse heart and brain were visible in the two in vivo simultaneous PET/MRI scans without applying image corrections. In conclusion, the insert demonstrated a good overall performance and can be used for small animal multi-modal research applications.
引用
收藏
页码:8671 / 8692
页数:22
相关论文
共 50 条
  • [21] Towards quantitative small-animal imaging on hybrid PET/CT and PET/MRI systems
    Amirrashedi, Mahsa
    Zaidi, Habib
    Ay, Mohammad Reza
    CLINICAL AND TRANSLATIONAL IMAGING, 2020, 8 (04) : 243 - 263
  • [22] PET Performance Evaluation of an MR-Compatible PET Insert
    Wu, Yibao
    Catana, Ciprian
    Farrell, Richard
    Dokhale, Purushottam A.
    Shah, Kanai S.
    Qi, Jinyi
    Cherry, Simon R.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (03) : 574 - 580
  • [23] PET performance and MRI compatibility evaluation of a digital, ToF-capable PET/MRI insert equipped with clinical scintillators
    Schug, David
    Wehner, Jakob
    Dueppenbecker, Peter Michael
    Weissler, Bjoern
    Gebhardt, Pierre
    Goldschmidt, Benjamin
    Salomon, Andre
    Kiessling, Fabian
    Schulz, Volkmar
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (18) : 7045 - 7067
  • [24] Performance evaluation of RF coils integrated with an RF-penetrable PET insert for simultaneous PET/MRI
    Lee, Brian J.
    Watkins, Ronald D.
    Lee, Keum Sil
    Chang, Chen-Ming
    Levin, Craig S.
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (02) : 1434 - 1446
  • [25] Performance Assessment of a Preclinical PET Scanner with Pinhole Collimation by Comparison to a Coincidence-Based Small-Animal PET Scanner
    Walker, Matthew D.
    Goorden, Marlies C.
    Dinelle, Katherine
    Ramakers, Ruud M.
    Blinder, Stephan
    Shirmohammad, Maryam
    van der Have, Frans
    Beekman, Freek J.
    Sossi, Vesna
    JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (08) : 1368 - 1374
  • [26] Performance evaluation of an adjustable gantry PET (AGPET) for small animal PET imaging
    Song, Hankyeol
    Kang, In Soo
    Kim, Kyu Bom
    Park, Chanwoo
    Baek, Min Kyu
    Lee, Seongyeon
    Chung, Yong Hyun
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2021, 53 (08) : 2646 - 2651
  • [27] New shielding configurations for a simultaneous PET/MRI scanner at 7T
    Peng, Bo J.
    Wu, Yibao
    Cherry, Simon R.
    Walton, Jeffrey H.
    JOURNAL OF MAGNETIC RESONANCE, 2014, 239 : 50 - 56
  • [28] Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph
    Woody, C.
    Schlyer, D.
    Vaska, P.
    Tomasi, D.
    Solis-Najera, S.
    Rooney, W.
    Pratte, J. -F.
    Junnarkar, S.
    Stoll, S.
    Master, Z.
    Purschke, M.
    Park, S. -J
    Southekal, S.
    Kriplani, A.
    Krishnamoorthy, S.
    Maramraju, S.
    O'Connor, P.
    Radeka, V.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 571 (1-2) : 102 - 105
  • [29] Performance evaluation of a small animal PET scanner a high level of multiplexing and charge-signal transmission
    Lee, Sangwon
    Choi, Yong
    Jung, Jin Ho
    Jung, Jiwoong
    Kim, kyu Bom
    Choe, Hyeok-jun
    Leem, Hyun Tae
    Lee, Hwunjae
    Huh, Yong-Min
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (04)
  • [30] Evaluation Detector Module of the miniPET-3 Small Animal PET Scanner
    Imrek, J.
    Hegyesi, Gy.
    Kalinka, G.
    Kiraly, B.
    Molnar, J.
    Nagy, F.
    Valastyan, I.
    Balkay, L.
    Szabo, Zs.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 3790 - 3793