Development of a phantom to validate high-dose-rate brachytherapy treatment planning systems with heterogeneous algorithms

被引:21
作者
Moura, Eduardo S. [1 ,2 ]
Micka, John A. [1 ]
Hammer, Cliff G. [1 ]
Culberson, Wesley S. [1 ]
DeWerd, Larry A. [1 ]
Rostelato, Maria Elisa C. M. [2 ]
Zeituni, Carlos A. [2 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53705 USA
[2] CNEN SP, Inst Pesquisas Energet & Nucl, BR-05508000 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
brachytherapy; Ir-192; treatment planning system; heterogeneities; RADIOCHROMIC FILM DOSIMETRY; PARTIAL BREAST IRRADIATION; MONTE-CARLO-SIMULATION; IONIZATION-CHAMBER; GAFCHROMIC FILM; AAPM PROTOCOL; PHOTON ENERGY; IR-192; SOURCE; RECOMMENDATIONS; DISTRIBUTIONS;
D O I
10.1118/1.4914390
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This work presents the development of a phantom to verify the treatment planning system (TPS) algorithms used for high-dose-rate (HDR) brachytherapy. It is designed to measure the relative dose in a heterogeneous media. The experimental details used, simulation methods, and comparisons with a commercial TPS are also provided. Methods: To simulate heterogeneous conditions, four materials were used: Virtual Water (TM) (VM), BR50/50 (TM), cork, and aluminum. The materials were arranged in 11 heterogeneity configurations. Three dosimeters were used to measure the relative response from a HDR Ir-192 source: TLD-100 (TM), Gafchromic (R) EBT3 film, and an Exradin (TM) A1SL ionization chamber. To compare the results from the experimental measurements, the various configurations were modeled in the PENELOPE/penEasy Monte Carlo code. Images of each setup geometry were acquired from a CT scanner and imported into BrachyVision (TM) TPS software, which includes a grid-based Boltzmann solver Acuros (TM). The results of the measurements performed in the heterogeneous setups were normalized to the dose values measured in the homogeneous Virtual Water (TM) setup and the respective differences due to the heterogeneities were considered. Additionally, dose values calculated based on the American Association of Physicists in Medicine-Task Group 43 formalism were compared to dose values calculated with the Acuros (TM) algorithm in the phantom. Calculated doses were compared at the same points, where measurements have been performed. Results: Differences in the relative response as high as 11.5% were found from the homogeneous setup when the heterogeneous materials were inserted into the experimental phantom. The aluminum and cork materials produced larger differences than the plastic materials, with the BR50/50 (TM) material producing results similar to the Virtual Water (TM) results. Our experimental methods agree with the PENELOPE/penEasy simulations for most setups and dosimeters. The TPS relative differences with the Acuros (TM) algorithm were similar in both experimental and simulated setups. The discrepancy between the BrachyVision (TM), Acuros (TM), and TG-43 dose responses in the phantom described by this work exceeded 12% for certain setups. Conclusions: The results derived from the phantom measurements show good agreement with the simulations and TPS calculations, using Acuros (TM) algorithm. Differences in the dose responses were evident in the experimental results when heterogeneous materials were introduced. These measurements prove the usefulness of the heterogeneous phantom for verification of HDR treatment planning systems based on model-based dose calculation algorithms. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:1566 / 1574
页数:9
相关论文
共 59 条
  • [1] Radiochromic film dosimetry of HDR 192Ir source radiation fields
    Aldelaijan, Saad
    Mohammed, Huriyyah
    Tomic, Nada
    Liang, Li-Heng
    DeBlois, Francois
    Sarfehnia, Arman
    Abdel-Rahman, Wamied
    Seuntjens, Jan
    Devic, Slobodan
    [J]. MEDICAL PHYSICS, 2011, 38 (11) : 6074 - 6083
  • [2] Optimal beam quality selection based on contrast-to-noise ratio and mean glandular dose in digital mammography
    Aminah, M.
    Ng, K. H.
    Abdullah, B. J. J.
    Jamal, N.
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2010, 33 (04) : 329 - 334
  • [3] The effect of patient inhomogeneities in oesophageal 192Ir HDR brachytherapy:: a Monte Carlo and analytical dosimetry study
    Anagnostopoulos, G
    Baltas, D
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12) : 2675 - 2685
  • [4] An analytical dosimetry model as a step towards accounting for inhomogeneities and bounded geometries in 192Ir brachytherapy treatment planning
    Anagnostopoulos, G
    Baltas, D
    Karaiskos, P
    Pantelis, E
    Papagiannis, P
    Sakelliou, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (11) : 1625 - 1647
  • [5] On the uncertainties of photon mass energy-absorption coefficients and their ratios for radiation dosimetry
    Andreo, Pedro
    Burns, David T.
    Salvat, Francesc
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (08) : 2117 - 2136
  • [6] [Anonymous], 2010, BRACHYVISION AC ALG
  • [7] Measurement of absorbed dose-to-water for an HDR 192Ir source with ionization chambers in a sandwich setup
    Araki, Fujio
    Kouno, Tomohiro
    Ohno, Takeshi
    Kakei, Kiyotaka
    Yoshiyama, Fumiaki
    Kawamura, Shinji
    [J]. MEDICAL PHYSICS, 2013, 40 (09)
  • [8] A BrachyPhantom for verification of dose calculation of HDR brachytherapy planning system
    Austerlitz, C.
    Campos, C. A. T.
    [J]. MEDICAL PHYSICS, 2013, 40 (11)
  • [9] Technical note:: Monte-Carlo dosimetry of the HDR 12i and plus 192Ir sources
    Ballester, F
    Puchades, V
    Lluch, JL
    Serrano-Andrés, MA
    Limami, Y
    Pérez-Calatayud, J
    Casal, E
    [J]. MEDICAL PHYSICS, 2001, 28 (12) : 2586 - 2591
  • [10] Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: Current status and recommendations for clinical implementation
    Beaulieu, Luc
    Tedgren, Asa Carlsson
    Carrier, Jean-Francois
    Davis, Stephen D.
    Mourtada, Firas
    Rivard, Mark J.
    Thomson, Rowan M.
    Verhaegen, Frank
    Wareing, Todd A.
    Williamson, Jeffrey F.
    [J]. MEDICAL PHYSICS, 2012, 39 (10) : 6208 - 6236