Biomechanical Analysis of the External Fixation in a Lumbar Fracture Model: A Finite Element Study

被引:4
|
作者
Liao, Yijia [1 ]
Yan, Yiguo [1 ]
Kang, Yu [1 ,2 ]
Wang, Wenjun [1 ]
Song, Xizheng [1 ]
Peng, Wen [1 ]
Fu, Haibin [1 ]
Chen, Huarong [1 ]
Wang, Cheng [1 ]
机构
[1] Univ South China, Affiliated Hosp 1, Hengyang Med Sch, Dept Spine Surg, Hengyang 421001, Hunan, Peoples R China
[2] Anhui Med Univ, Affiliated Hosp 1, Dept Orthopedics, Hefei 230000, Anhui, Peoples R China
关键词
External fixation; Finite element analysis; Lumbar fracture; Biomechanics; THORACOLUMBAR BURST FRACTURES; ADJACENT SEGMENT DISEASE; TRANSPEDICULAR FIXATION; POSTERIOR APPROACH; SPINE; FUSION; VERTEBROPLASTY; COMPLICATIONS; EPIDEMIOLOGY; MOVEMENTS;
D O I
10.1007/s40846-022-00727-2
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose This study aimed to investigate the biomechanical characteristics of the external spinal fixation for treating lumbar fracture through finite element analysis (FEA) and provide a theoretical basis for its further application. Methods Two different models of L3 fracture fixed with the external spinal fixation and the internal fixation system respectively were constructed. The ROM, maximum stresses at L3, and the screws of the two models were measured under load control. Subsequently, the applied torque, the maximum stressed at L3, L1/2, L2/3, L3/4, L4/5 discs and the screws were analyzed under displacement control. Results Under load control, the external fixation model reserved more ROM than the internal fixation model (40.4-48.0% vs 30.5-41.0%). Compared to the internal fixation model, the maximum stresses at L3 and screws in the external fixation model were increased. Under displacement control, the external fixation model required fewer moments (N center dot mm) than the internal fixation model (flexion: 7500 vs 12,294; extension: 7500 vs 9027). Further, the maximum stresses at L3 and the screws in the external fixation model were greater than those of the internal fixation model, while the maximum stresses at the upper and lower adjacent discs of fixed segments were less than the internal fixation model. Conclusion Compared to the internal fixation system, the external fixation has a better stress distribution with the greater overall mobility. It theoretically reduces the stress concentration of the adjacent discs and the stress shielding of the fractured vertebral body.
引用
收藏
页码:469 / 478
页数:10
相关论文
共 50 条
  • [1] Biomechanical Analysis of the External Fixation in a Lumbar Fracture Model: A Finite Element Study
    Yijia Liao
    Yiguo Yan
    Yu Kang
    Wenjun Wang
    Xizheng Song
    Wen Peng
    Haibin Fu
    Huarong Chen
    Cheng Wang
    Journal of Medical and Biological Engineering, 2022, 42 : 469 - 478
  • [2] Biomechanical analysis of the tandem spinal external fixation in a multiple-level noncontiguous lumbar fractures model: a finite element analysis
    Chen, Huarong
    Kang, Yu
    Yan, Yiguo
    Wang, Hu
    Peng, Wen
    Liao, Yijia
    Zou, Mingxiang
    Xu, Zhun
    Song, Xizheng
    Wang, Wenjun
    Wang, Cheng
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2024, 12
  • [3] Biomechanical effects of vertebroplasty on thoracolumbar burst fracture with transpedicular fixation: A finite element model analysis
    Xu, G.
    Fu, X.
    Du, C.
    Ma, J.
    Li, Z.
    Ma, X.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2014, 100 (04) : 379 - 383
  • [4] Biomechanical analysis of a customized lumbar interspinous spacer based on transfacetopedicular screw fixation: A finite element study
    Zhao, Gaiping
    Wang, Xiaoyu
    Liu, Tielong
    Dong, Lianfeng
    MEDICAL ENGINEERING & PHYSICS, 2022, 107
  • [5] A novel external fixation for treating tibial fractures: a finite element and biomechanical study
    Shen Liu
    Xiangdang Liang
    Songyang Liu
    Zhanshe Guo
    Xing Wei
    Yonghui Liang
    Journal of Orthopaedic Surgery and Research, 20 (1)
  • [6] Biomechanical evaluation of lumbar spondylolysis repair with various fixation options: A finite element analysis
    Ye, Yuchen
    Jin, Shichang
    Zou, Yang
    Fang, Yuekun
    Xu, Panpan
    Zhang, Zhili
    Wu, Nan
    Zhang, Changchun
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [7] Finite Element Analysis of Fracture Fixation
    Gregory S. Lewis
    Dominic Mischler
    Hwabok Wee
    J. Spence Reid
    Peter Varga
    Current Osteoporosis Reports, 2021, 19 : 403 - 416
  • [8] Finite Element Analysis of Fracture Fixation
    Lewis, Gregory S.
    Mischler, Dominic
    Wee, Hwabok
    Reid, J. Spence
    Varga, Peter
    CURRENT OSTEOPOROSIS REPORTS, 2021, 19 (04) : 403 - 416
  • [9] Biomechanical analysis of fracture fixation with external fixator in vivo
    MING Jiang-hua
    ZHOU Pang-hu
    ZHOU Jian-peng
    ChineseJournalofTraumatology, 2006, (02) : 100 - 104
  • [10] Lumbar Finite Element Analysis and Experimental Study Based on the Biomechanical Properties
    Wang, Jianping
    Gu, Juping
    Zhao, Jian
    SYSTEMS SIMULATION AND SCIENTIFIC COMPUTING, PT I, 2012, 326 : 390 - +