Comparison of two double red cell collection settings on Fenwal Alyx apheresis instrument

被引:2
作者
Burgstaler, Edwin A. [1 ]
Duffy, Kimberly J. [1 ]
Gandhi, Manish J. [1 ]
机构
[1] Mayo Clin, Div Transfus Med, Dept Lab Med & Pathol, Rochester, MN USA
关键词
automated blood collection; Alyx; double red blood cell; RBC; RBCS;
D O I
10.1002/jca.21527
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Fenwal Alyx for collecting double red cell products has two red cell volume collection settings: fixed collection target of 360 ml (180 ml/unit) and a variable target of collecting either 400 or 360 ml (200 or 180 ml/unit), where the machine aims for the higher possible collection target. We retrospectively compared the two collection targets for the RBC content, donor time, technician time, and collection efficiency. We compared 18 fixed (F) target collections to 40 variable (V) target collections. All collections were performed as per the manufacturer's recommendations on Alyx and donors met the manufacturer's eligibility criteria. There was no significant difference in average whole blood processed (F: 963 ml, V: 1,000 ml); donor time (F: 43 min, V: 45 min) or technician time (F: 64 min, V: 64 min). There was a significant difference in unit volume (F: 283 ml, V: 300 ml); grams Hb/unit (F: 53 g, V: 57 g); ml RBC/unit (F: 157 ml, V: 167 ml); and RBC recovery (F: 87.8%, V: 88.9%). The fixed target had a significantly lower frequency of products with 51 g Hb (80.6%) than variable target (96.3%) and 153 ml RBC/unit (F: 55.6%, V: 96.3%). In conclusion, the variable target efficiently allows collections of products with higher red cell volume and hemoglobin without a significant increase in collection and processing time.
引用
收藏
页码:392 / 396
页数:5
相关论文
共 8 条
  • [1] AABB, 2014, Standards for Blood Banks and Transfusion Services, V29th
  • [2] Burgstaler E.S., 2010, APHERESIS PRINCIPLES, P71
  • [3] Double red blood cell collection: comparison of three apheresis systems
    Burgstaler, Edwin A.
    Duffy, Kimberly J.
    Gandhi, Manish J.
    [J]. TRANSFUSION, 2016, 56 (09) : 2362 - 2367
  • [4] Fenwal. Alyx, 2011, ALYX COMP COLL SYST
  • [5] Effect of increasing hemoglobin cutoff in male donors and increasing interdonation interval in whole blood donors at a hospital-based blood donor center
    Gandhi, Manish J.
    Duffy, Kimberly
    Benike, Mary
    Jenkins, Sarah
    Stubbs, James R.
    [J]. TRANSFUSION, 2012, 52 (09) : 1880 - 1888
  • [6] In vitro quality of red blood cells (RBCs) collected by multicomponent apheresis compared to manually collected RBCs during 49 days of storage
    Picker, Susanne M.
    Radojska, Stela M.
    Gathof, Birgit S.
    [J]. TRANSFUSION, 2007, 47 (04) : 687 - 696
  • [7] Prospective evaluation of double RBC collection using three different apheresis systems
    Picker, Susanne M.
    Radojska, Stela M.
    Gathof, Birgit S.
    [J]. TRANSFUSION AND APHERESIS SCIENCE, 2006, 35 (03) : 197 - 205
  • [8] Collection of two units of leukoreduced RBCs from a single donation with a portable multiple-component collection system
    Snyder, EL
    Elfath, MD
    Taylor, H
    Rugg, N
    Greenwalt, TJ
    Baril, L
    Whitley, P
    Brantigan, B
    Story, K
    [J]. TRANSFUSION, 2003, 43 (12) : 1695 - 1705