Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy

被引:34
|
作者
Fix, MK
Keller, H
Rüegsegger, P
Born, EJ
机构
[1] Univ Zurich, Inst Biomed Engn, CH-8044 Zurich, Switzerland
[2] Swiss Fed Inst Technol, Swiss Fed Inst Technol, CH-8044 Zurich, Switzerland
[3] Univ Bern, Inselspital, Clin Radiooncol, Div Med Radiat Phys, CH-3045 Bern, Switzerland
关键词
Monte Carlo simulation; photon dose calculation; beam model;
D O I
10.1118/1.1318220
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Monte Carlo (code GEANT) produced 6 anti 15 MV phase space (PS) data were used to define several simple photon beam models. For creating the PS data the energy of starting electrons hitting the target was tuned to get correct depth dose data compared to measurements. The modeling process used the full PS information within the geometrical boundaries of the beam including all scattered radiation of the accelerator head. Scattered radiation outside the boundaries was neglected. Photons and electrons were assumed to be radiated from point sources. Four different models were investigated which involved different ways to determine the energies and locations of beam particles in the output plane. Depth dose curves, profiles, and relative output factors were calculated with these models for six field sizes from 5x5 to 40x40 cm(2) and compared to measurements. Model 1 uses a photon energy spectrum independent of location in the PS plane and a constant photon fluence in this plane. Model 2 takes into account the spatial particle fluence distribution in the PS plane. A constant fluence is used again in model 3, but the photon energy spectrum depends upon the off axis position. Model 4, finally uses the spatial particle fluence distribution and off axis dependent photon energy spectra in the PS plane. Depth dose curves and profiles for field sizes up to 10 x 10 cm(2) were not model sensitive. Good agreement between measured and calculated depth dose curves and profiles for all field sizes was reached for model 4. However, increasing deviations were found for increasing field sizes for models 1-3. Large deviations resulted for the profiles of models 2 and 3. This is due to the fact that these models overestimate and underestimate the energy fluence at large off axis distances. Relative output factors consistent with measurements resulted only for model 4. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00911-1].
引用
收藏
页码:2739 / 2747
页数:9
相关论文
共 50 条
  • [41] COMPARISON OF ONETRAN CALCULATIONS OF ELECTRON BEAM DOSE PROFILES WITH MONTE CARLO AND EXPERIMENT.
    Garth, J.C.
    Woolf, S.
    IEEE Transactions on Nuclear Science, 1987, NS-34 (06)
  • [42] Analysis of algorithms for photon beam dose calculations
    Traneus, E.
    Ahnesjo, A.
    Failla, G.
    Wareing, T.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S221 - S221
  • [43] Integral dose in external beam photon radiotherapy
    Mackie, T
    Kissick, M
    Jeraj, R
    Tome, W
    Fenwick, J
    Olivera, G
    Ritter, M
    MEDICAL PHYSICS, 2004, 31 (06) : 1721 - 1721
  • [44] Are neutrons responsible for the dose discrepancy between Monte Carlo calculations and measurements in a high-energy photon beam at the buildup region?
    Ding, GG
    Duzenli, C
    Kalach, N
    MEDICAL PHYSICS, 2002, 29 (06) : 1233 - 1233
  • [45] Depth Dose Enhancement on Flattening-Filter-Free Photon Beam: A Monte Carlo Study in Nanoparticle-Enhanced Radiotherapy
    Chow, James C. L.
    APPLIED SCIENCES-BASEL, 2020, 10 (20): : 1 - 10
  • [46] Monte Carlo calculations of correction factors for plastic phantoms in clinical photon and electron beam dosimetry
    Araki, Fujio
    Hanyu, Yuji
    Fukuoka, Miyoko
    Matsumoto, Kenji
    Okumura, Masahiko
    Oguchi, Hiroshi
    MEDICAL PHYSICS, 2009, 36 (07) : 2992 - 3001
  • [47] Accuracy of Monte Carlo Dose Calculations with Kilovoltage Photon Beams
    Bazalova, M.
    Graves, E.
    MEDICAL PHYSICS, 2011, 38 (06) : 3871 - +
  • [48] Absolute dose calculations for Monte Carlo simulations of radiotherapy beams
    Popescu, IA
    Shaw, CP
    Zavgorodni, SF
    Beckham, WA
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (14): : 3375 - 3392
  • [49] A Monte Carlo evaluation of RapidArc dose calculations for oropharynx radiotherapy
    Gagne, I. M.
    Ansbacher, W.
    Zavgorodni, S.
    Popescu, C.
    Beckham, W. A.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (24): : 7167 - 7185
  • [50] Evaluation of PENFAST - A fast Monte Carlo code for dose calculations in photon and electron radiotherapy treatment planning
    Habib, B.
    Poumarede, B.
    Tola, F.
    Barthe, J.
    PHYSICA MEDICA, 2010, 26 (01) : 17 - 25