Measurement of total and unbound bictegravir concentrations in plasma and cerebrospinal fluid by UHPLC-MS/MS

被引:17
作者
Rigo-Bonnin, Raul [1 ]
Manuel Tiraboschi, Juan [2 ]
Alvarez-Alvarez, Marta [1 ]
Ainara Perez-Fernandez, Gloria [1 ]
Sanjuas-Iglesias, Mercedes [1 ]
Scevola, Sofia [2 ]
Niubo, Jordi [3 ]
Videla, Sebastian [4 ]
Podzamczer, Daniel [2 ]
机构
[1] Hosp Univ Bellvitge IDIBELL, Lab Clin Dept, Barcelona, Spain
[2] Hosp Univ Bellvitge IDIBELL, Dept Infect Dis, HIV & STI Unit, Barcelona, Spain
[3] Hosp Univ Bellvitge, IDIBELL, Microbiol Dept, Barcelona, Spain
[4] Univ Barcelona, Hosp Univ Bellvitge IDIBELL, Clin Pharmacol Dept, Clin Res Support Unit, Barcelona, Spain
关键词
Cerebrospinal fluid; plasma; total bictegravir; UHPLC-MS/MS; ultrafiltration; unbound bictegravir; PROTEIN BINDING; ULTRAFILTRATION;
D O I
10.1016/j.jpba.2020.113250
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Bictegravir is a novel integrase strand transfer inhibitor, administrated in co-formulation with tenofovir alafenamide and emtricitabine (Biktarvy (R)), indicated in the management of HIV-1 infection in patients not previously treated with antiretroviral therapy. Bictegravir is highly bound to plasma proteins, and this significantly determines its clearance, solubility, and activity. These characteristics are crucial determinants of bictegravir penetration into human body compartments, as the central nervous system. We developed and validated UHPLC-MS/MS procedures to measure total and unbound bictegravir concentrations in plasma and cerebrospinal fluid. Simple protein precipitation with acetonitrile was implemented to prepare plasma and cerebrospinal fluid samples. Sample preparation was preceded by ultrafiltration for measuring unbound bictegravir concentrations. Chromatographic separations were achieved on an Acquity (R) UHPLC (R) BEH (TM) (2.1 x 100 mm id, 1.7 mu m) reverse-phase C-18 column using an iso-cratic mobile phase 20:80 (v/v) water/acetonitrile with 0.1% formic. Bictegravir and its internal standard (bictegravir-N-15 d(2)) were detected by electrospray ionization mass spectrometry in positive and multiple reaction monitoring modes, using transitions of 450.2 -> 289.2/145.4 and 453.2 -> 289.2, respectively. Ultrafiltration procedures presented non-specific bindings of (8.6 +/- 1.2) % for bictegravir in plasma and (26.6 +/- 3.1) % for bictegravir in cerebrospinal fluid. Linearity was observed between (10.70-8560) mu g/L, (1.07-856.0) mu g/L for total and unbound bictegravir in plasma, and 0.107-26.75 mu g/L for total and unbound bictegravir in cerebrospinal fluid. Imprecisions, absolute relative biases, normalized-matrix factors, and normalized-recoveries were <= 14.4%, <= 13.8%, (97.4-102.5) %, and (99.8-105.1) %, respectively. No significant interferences and carry-over were observed. The validated UHPLC-MS/MS procedures could be useful for pharmacokinetic and pharmacodynamic studies. (C) 2020 Elsevier B.V. All rights reserved.
引用
收藏
页数:11
相关论文
共 20 条
  • [1] [Anonymous], 2018, BIOAN METH VAL GUID
  • [2] [Anonymous], [No title captured]
  • [3] [Anonymous], 2003, EP06A CLSI
  • [4] [Anonymous], [No title captured]
  • [5] Clinical and Laboratory Standard Institute, 2014, C62A CLSI
  • [6] Bictegravir/Emtricitabine/Tenofovir Alafenamide: A Review in HIV-1 Infection
    Deeks, Emma D.
    [J]. DRUGS, 2018, 78 (17) : 1817 - 1828
  • [7] European Medicines Agency E., 2011, Science Medicines Health, Guideline on bioanalytical method validation.
  • [8] Fayet A, 2008, THER DRUG MONIT, V30, P511, DOI 10.1097/FTD.0b013e3181817318
  • [9] Gallant J, 2017, LANCET, V390, P2063, DOI [10.1016/s0140-6736(17)32299-7, 10.1016/S0140-6736(17)32299-7]
  • [10] Challenges regarding analysis of unbound fraction of highly bound protein antiretroviral drugs in several biological matrices: lack of harmonisation and guidelines
    Illamola, Silvia M.
    Hirt, Deborah
    Treluyer, Jean M.
    Urien, Saik
    Benaboud, Sihem
    [J]. DRUG DISCOVERY TODAY, 2015, 20 (04) : 466 - 474