摘要
Background and purpose Customized femoral stems are designed to have a perfect fit and fill in the femur in order to achieve physiological load transfer and minimize stress shielding. Dual-energy X-ray absorptiometry (DXA) is regarded as an accurate method for detection of small alterations in bone mineral density (BMD) around hip prostheses. We present medium-term DXA results from a randomized study comparing a customized and an anatomical femoral stem. Methods 100 hips were randomized to receive either the anatomical ABG-I stem or the Unique customized femoral stem, both uncemented. DXA measurements were conducted postoperatively and after 3, 6, 12, 24, 36, and 60 months, and BMD was computed for each of the 7 Gruen zones in the proximal femur. Results Results from 87 patients were available for analysis. 78 completed the 5-year follow-up: 35 patients in the ABG group and 43 patients in the Unique group. In both groups, we found the greatest degree of bone loss in the proximal Gruen zones. In zone 1, there was 15% reduction in BMD in the ABG-I group and 14% reduction in the Unique group. In zone 7, the reduction was 28% in the ABG-I group and 27% in the Unique group. The only statistically significant difference between the groups was found in Gruen zone 4, which is distal to the tip of the stem, with 1.6% reduction in BMD in the ABG-I group and 9.7% reduction in the Unique group (p = 0.003). Interpretation 5-year DXA results showed that because of stress-shielding, proximal bone loss could not be avoided-either for the anatomical ABG-I stem or for the customized Unique stem.
引用
收藏
页码:410 / 416
页数:7
相关论文
共 39 条
[1]   Changes in proximal femoral strain after insertion of uncemented standard and customised femoral stems [J].
Aamodt, A ;
Lund-Larsen, J ;
Eine, J ;
Andersen, E ;
Benum, P ;
Husby, OS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (06) :921-929
[2]   Determination of the Hounsfield value for CT-based design of custom femoral stems [J].
Aamodt, A ;
Kvistad, KA ;
Andersen, E ;
Lund-Larsen, J ;
Eine, J ;
Benum, P ;
Husby, OS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01) :143-147
[3]   Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems 37 patients followed for 3 years [J].
Albanese, Carlina V. ;
Santori, Francesco S. ;
Pavan, Laura ;
Learmonth, Ian D. ;
Passariello, Roberto .
ACTA ORTHOPAEDICA, 2009, 80 (03) :291-297
[4]   Periprosthetic bone remodelling after cementless total hip replacement - A prospective comparison of two different implant designs [J].
Ang, KC ;
De, SD ;
Goh, JCH ;
Low, SL ;
Bose, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (04) :675-679
[5]  
BARGAR WL, 1989, CLIN ORTHOP RELAT R, P73
[6]   Uncemented custom femoral components in hip arthroplasty A prospective clinical study of 191 hips followed for at least 7 years [J].
Benum, Pal ;
Aamodt, Arild .
ACTA ORTHOPAEDICA, 2010, 81 (04) :427-435
[7]  
BOBYN JD, 1992, CLIN ORTHOP RELAT R, P79
[8]   ACCURACY OF DEXA MEASUREMENT OF BONE-MINERAL DENSITY AFTER TOTAL HIP-ARTHROPLASTY [J].
COHEN, B ;
RUSHTON, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (03) :479-483
[9]   FUNCTIONAL RESULTS OF HIP ARTHROPLASTY WITH ACRYLIC PROSTHESIS [J].
DAUBIGNE, RM ;
POSTEL, M .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1954, 36-A (03) :451-475
[10]  
ENGH CA, 1988, CLIN ORTHOP RELAT R, P7