Potential adverse effects of high-intensity focused ultrasound exposure on blood vessels in vivo

被引:125
作者
Hynynen, K [1 ]
Chung, AH [1 ]
Colucci, V [1 ]
Jolesz, FA [1 ]
机构
[1] HARVARD UNIV,SCH MED,BOSTON,MA 02115
关键词
ultrasound; bioeffects; surgery; blood vessels; therapy; minimally invasive surgery;
D O I
10.1016/0301-5629(95)02044-6
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The aim of the study was to evaluate the potential adverse effects of high intensity ultrasound exposure on blood vessels during noninvasive focused ultrasound surgery. A hydraulic MR-compatible positioning device was used to manipulate a focused ultrasound transducer (frequency 1.49 MHz, f-number = 0.8) in an MRI scanner. The system was used to sonicate a branch of the femoral artery and vein of 19 rabbits (26 thighs) in vivo at intensity levers above the threshold for transient cavitation; i.e., between 4400 and 8800 W cm(-2) with multiple 1 s pulses stepped across the vessels (step size = 0.7 mm). The vessels were located and followed by MR angiography. In 13 rabbits, x-ray angiograms were also performed after the animals were euthanized. The results demonstrated that the 1 s high-intensity exposures caused the arteries to constrict at all exposure levels tested. At the intensity of 5800 W cm(-2) and above, the MRI angiogram immediately after the sonications showed no flow. The x-ray angiograms (1-2 h later) showed that the blood vessels were open, but constricted to about 50% or less of their diameter. Both the MR and x-ray angiograms showed that the vessel diameters relaxed toward their initial diameter during the first week after sonication. In five cases, hemorrhage or vessel rupture was caused by the sonication. This study demonstrates that short, high-intensity focused ultrasound exposure can cause vessel spasm and hemorrhage when transient cavitation is present. This condition should be avoided during noninvasive focused ultrasound surgery.
引用
收藏
页码:193 / 201
页数:9
相关论文
共 29 条
  • [1] RATE-PROCESS MODEL FOR ARTERIAL TISSUE THERMAL-DAMAGE - IMPLICATIONS ON VESSEL PHOTOCOAGULATION
    AGAH, R
    PEARCE, JA
    WELCH, AJ
    MOTAMEDI, M
    [J]. LASERS IN SURGERY AND MEDICINE, 1994, 15 (02) : 176 - 184
  • [2] [Anonymous], 2012, ESSENTIALS MED ULTRA
  • [3] EFFECT OF BLOOD PERFUSION ON THE ABLATION OF LIVER PARENCHYMA WITH HIGH-INTENSITY FOCUSED ULTRASOUND
    CHEN, L
    TERHAAR, G
    HILL, CR
    DWORKIN, M
    CARNOCHAN, P
    YOUNG, H
    BENSTED, JPM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (11) : 1661 - 1673
  • [4] FOCUSED US SYSTEM FOR MR IMAGING-GUIDED TUMOR ABLATION
    CLINE, HE
    HYNYNEN, K
    WATKINS, RD
    ADAMS, WJ
    SCHENCK, JF
    ETTINGER, RH
    FREUND, WR
    VETRO, JP
    JOLESZ, FA
    [J]. RADIOLOGY, 1995, 194 (03) : 731 - 737
  • [5] MR TEMPERATURE MAPPING OF FOCUSED ULTRASOUND SURGERY
    CLINE, HE
    HYNYNEN, K
    HARDY, CJ
    WATKINS, RD
    SCHENCK, JF
    JOLESZ, FA
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (06) : 628 - 636
  • [6] COLEMAN DJ, 1985, OPHTHALMOLOGY, V92, P339
  • [7] ONLINE MONITORING OF ULTRASONIC SURGERY WITH MR-IMAGING
    DARKAZANLI, A
    HYNYNEN, K
    UNGER, EC
    SCHENCK, JF
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (03): : 509 - 514
  • [8] THE EFFECTS OF TISSUE HETEROGENEITIES AND LARGE BLOOD-VESSELS ON THE THERMAL EXPOSURE INDUCED BY SHORT HIGH-POWER ULTRASOUND PULSES
    DORR, LN
    HYNYNEN, K
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1992, 8 (01) : 45 - 59
  • [9] FALLON JT, 1972, ARCH PATHOL, V94, P380
  • [10] FOSTER RS, 1993, EUR UROL, V23, P29