Pachymetric measurements with a new Scheimpflug photography-based system Intraobserver repeatability and agreement with optical coherence tomography pachymetry

被引:67
作者
Milla, Myriam [1 ]
Pinero, David P. [1 ,2 ]
Amparo, Francisco [1 ]
Alio, Jorge L. [1 ,3 ]
机构
[1] Vissum Corp, Inst Oftalmol Alicante, Alicante 03016, Spain
[2] Univ Alicante, Dept Opt, E-03080 Alicante, Spain
[3] Univ Miguel Hernandez, Div Ophthalmol, Alicante, Spain
关键词
CORNEAL THICKNESS MEASUREMENTS; IN-SITU KERATOMILEUSIS; ULTRASOUND PACHYMETRY; ANTERIOR SEGMENT; OCULUS-PENTACAM; ORBSCAN-II; REPRODUCIBILITY; VISANTE; TOPOGRAPHY; MYOPIA;
D O I
10.1016/j.jcrs.2010.08.038
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the intraobserver repeatability of pachymetric measurements obtained with a new Scheimpflug photography-based system and the agreement of the measurements with those obtained by optical coherence tomography (OCT). SETTING: Vissum Corporation, Alicante, Spain. DESIGN: Evaluation of diagnostic technology. METHODS: Corneal thickness was measured at different locations (center; 2.5 mm and 4.0 mm nasal, temporal, superior, and inferior) in 18 eyes of 18 patients with no ocular pathology or previous surgery with a Scheimpflug photography-based system (Sirius). Three consecutive measurements at each location were obtained with the system to evaluate intraobserver repeatability. Afterward, pachymetric measurements were obtained at the same locations with an OCT system (Visante) to evaluate the agreement between techniques using the Bland-Altman method. RESULTS: The intraclass correlation coefficient of repeated measures ranged from 0.990 to 0.997. The coefficient of variation was lower than 1% and the standard deviations of the repeated measurements (S-w) were below 6 mu m at all corneal locations. Statistically significant differences were found between the coefficient of variation and S-w, values of the repeated measurements corresponding to the 2.5 mm and 4.0 mm locations (P <=.037). The agreement with OCT pachymetric measurements was poor, with ranges of agreement larger than 20 mu m at all corneal locations. Statistically significant differences were found between standard deviations of the differences between the 2 devices at the 2.5 mm and 4.0 mm locations (P <=.017). CONCLUSION: Scheimpflug photography-based technology provided repeatable pachymetric measurements that were not equivalent to those obtained with OCT.
引用
收藏
页码:310 / 316
页数:7
相关论文
共 22 条
  • [1] Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography
    Amano, S
    Honda, N
    Amano, Y
    Yamagami, S
    Miyai, T
    Samejima, T
    Ogata, M
    Miyata, K
    [J]. OPHTHALMOLOGY, 2006, 113 (06) : 937 - 941
  • [2] Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry
    Barkana, Y
    Gerber, Y
    Elbaz, U
    Schwartz, S
    Ken-Dror, G
    Avni, I
    Zadok, D
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) : 1729 - 1735
  • [3] Measurement error and correlation coefficients
    Bland, JM
    Altman, DG
    [J]. BRITISH MEDICAL JOURNAL, 1996, 313 (7048) : 41 - 42
  • [4] Bland JM, 1996, BRIT MED J, V313, P744
  • [5] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [6] Cheng ACK, 2008, J REFRACT SURG, V24, P361, DOI 10.3928/1081597X-20080401-08
  • [7] Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia
    Ho, Thomas
    Cheng, Arthur C. K.
    Rao, Srinivas K.
    Lau, Silvania
    Leung, Chris K. S.
    Lam, Dennis S. C.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (07) : 1177 - 1182
  • [8] Repeatability, reproducibility, and agreement characteristics of rotating Scheimpflug photography and scanning-slit corneal topography for corneal power measurement
    Kawamorita, Takushi
    Uozato, Hiroshi
    Kamiya, Kazutaka
    Bax, Leon
    Tsutsui, Kenta
    Aizawa, Daisuke
    Shimizu, Kimiya
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) : 127 - 133
  • [9] Comparison of central corneal thickness using anterior segment optical coherence tomography vs ultrasound pachymetry
    Kim, Hanna Y.
    Budenz, Donald L.
    Lee, Pak S.
    Feuer, William J.
    Barton, Keith
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (02) : 228 - 232
  • [10] Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound
    Lackner, B
    Schmidinger, G
    Pieh, S
    Funovics, MA
    Skorpik, C
    [J]. OPTOMETRY AND VISION SCIENCE, 2005, 82 (10) : 892 - 899