Clinical proof of practicability for an ECG device without any conductive contact

被引:10
|
作者
Czaplik, Michael [1 ]
Eilebrecht, Benjamin [2 ]
Ntouba, Alexandre [3 ]
Walter, Marian [2 ]
Schauerte, Patrick [4 ]
Leonhardt, Steffen [2 ]
Rossaint, Rolf [1 ]
机构
[1] Rhein Westfal TH Aachen, Dept Anaesthesiol, Aachen, Germany
[2] Rhein Westfal TH Aachen, Philips Chair Med Informat Technol, Aachen, Germany
[3] Practicien Hosp Anesthesie Reanimat, Amiens 1, France
[4] Rhein Westfal TH Aachen, Dept Cardiol, Aachen, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2010年 / 55卷 / 05期
关键词
capacitively coupled electrodes; ECG; electrocardiogram; non-conductive contact; premedication department; rhythm disturbances; ELECTRODES; SYSTEM; HEART;
D O I
10.1515/BMT.2010.035
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Heart rhythm disturbances are common symptoms of several heart disorders. One of the most effective screening methods is the traditional electrode-based ECG. However, this examination can be both time- and resource-consuming. Capacitive-coupling ECG (cECG) screening - working without any conductive electrical contact with the patient - might help to shorten the time required for diagnosis. In this study, we examine the practicability of employing a non-contact capacitive ECG in a clinical setting. A total of 30 volunteer patients aged over 50 years without pacemakers were included in our trial, after obtaining their written informed consent and their medical history. A cECG as well as a conventional, conductive ECG were recorded simultaneously. In addition to mathematical analysis, ECG data were manually evaluated by two clinicians blinded to the recording method and patient conditions. Data from 30 patients were collected during our study, seven of whom had experienced myocardial infarction. The obtained cECG signals showed a high correlation with the simultaneously recorded Einthoven lead II of the conventional ECG. The values for heart rate, PQ and QT time periods correlated particularly well. Significant differences were observed with regard to QRS duration. Data recorded in the supine position contained less motion artefacts and, in particular, there were fewer breathing artefacts compared to data collected from those in a sitting position. Owing to the easy and quick application of the cECG system, the feedback from the examined patients was consistently positive. In conclusion, recording cECG data in a sitting position provided sufficient quality for screening purposes. Further studies will be needed for the evaluation of cECG appropriateness in diagnosing heart disease.
引用
收藏
页码:291 / 300
页数:10
相关论文
共 50 条
  • [1] Clinical proof of practicability of a contactless ECG device
    Czaplik, M.
    Eilebrecht, B.
    Walocha, R.
    Schauerte, P.
    Rossaint, R.
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2010, 27 (01) : 65 - 66
  • [2] ECG measurement on a chair without conductive contact
    Lim, YG
    Kim, KK
    Park, KS
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (05) : 956 - 959
  • [3] Is there any clinical or experimental proof that the ovary has an internal secretion
    Morley, WH
    SURGERY GYNECOLOGY & OBSTETRICS, 1911, 13 : 28 - 31
  • [4] MOBILE ECG MONITORING DEVICE USING WEARABLE NON CONTACT ARMBAND
    Alexander, Aida
    Arun, C. S.
    PROCEEDINGS OF 2017 IEEE INTERNATIONAL CONFERENCE ON CIRCUIT ,POWER AND COMPUTING TECHNOLOGIES (ICCPCT), 2017,
  • [5] Double any-position alignment without rotatable device for SINS
    Liu Baiqi
    Fang Jiancheng
    SIGNAL ANALYSIS, MEASUREMENT THEORY, PHOTO-ELECTRONIC TECHNOLOGY, AND ARTIFICIAL INTELLIGENCE, PTS 1 AND 2, 2006, 6357
  • [6] RANDOMIZED RESPONSE TECHNIQUE WITHOUT MAKING USE OF ANY RANDOMIZING DEVICE
    TAKAHASI, K
    SAKASEGAWA, H
    ANNALS OF THE INSTITUTE OF STATISTICAL MATHEMATICS, 1977, 29 (01) : 1 - 8
  • [7] DEVICE FOR STUDY OF COMMUNICATION OF DOMINANCE WITHOUT PHYSICAL CONTACT
    WISE, LA
    ZIMMERMAN, RR
    STROBEL, DA
    SMITH, OW
    PERCEPTUAL AND MOTOR SKILLS, 1974, 39 (02) : 863 - 866
  • [8] A Novel Wearable Device for Continuous Ambulatory ECG Recording: Proof of Concept and Assessment of Signal Quality
    Steinberg, Christian
    Philippon, Francois
    Sanchez, Marina
    Fortier-Poisson, Pascal
    O'Hara, Gilles
    Molin, Franck
    Sarrazin, Jean-Francois
    Nault, Isabelle
    Blier, Louis
    Roy, Karine
    Plourde, Benoit
    Champagne, Jean
    BIOSENSORS-BASEL, 2019, 9 (01):
  • [9] Clinical Testing of the Apnea Prevention Device: Proof of Concept Data
    Zornow, Mark H.
    ANESTHESIA AND ANALGESIA, 2011, 112 (03): : 582 - 586
  • [10] Clinical Validation of a Smartphone Based, 6-lead ECG Device
    Stavrakis, Stavros
    Garabelli, Paul J.
    Smith, Landgrave
    Albert, David
    Po, Sunny S.
    CIRCULATION, 2017, 136