Long-term stability of liquid ionization chambers with regard to their qualification as local reference dosimeters for low dose-rate absorbed dose measurements in water

被引:7
作者
Bahar-Gogani, J [1 ]
Grindborg, JE
Johansson, BE
Wickman, G
机构
[1] Umea Univ, Dept Radiat Phys, S-90185 Umea, Sweden
[2] Swedish Radiat Protect Inst, S-17116 Stockholm, Sweden
关键词
D O I
10.1088/0031-9155/46/3/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The long-term sensitivity and calibration stability of liquid ionization chambers (LICs) has been studied at a local and a secondary standards dosimetry laboratory over a period of 3 years. The chambers were transported several times by mail between the two laboratories for measurements. The LICs used in this work are designed For absorbed dose measurements in the dose rate region of 0.1-100 mGy min(-1) and have a liquid layer thickness of 1 mm and a sensitive volume of 16.2 mm(3). The liquids used as sensitive media in the chambers are mixtures of isooctane (C8H18) and tetramethylsilane (Si(CH3)(4)) in different proportions (about 2 to 1). Operating at a polarizing voltage of 300 V the leakage current of the chambers was stable acid never exceeded 3% of the observable current at a dose rate of about 1 mGy min(-1). The volume sensitivity of the chambers was measured to be of the order of 10(-9) C Gy(-1) mm(-3). No systematic changes in the absorbed dose to water calibration was observed for any of the chambers during the test period (sigma < 0.2%). Variations in chamber dose response with small changes in the polarizing voltage as well as sensitivity changes with accumulated absorbed dose were also investigated. Measurements showed that the LIC response varies by 0.15% per 1% change in applied voltage around 300 V. No significant change could be observed in the LIC sensitivity after a single absorbed dose of 15 kCy. The results indicate that the LIC can be made to serve as a calibration transfer instrument and a reference detector for absorbed dose to water determinations providing good precision and long-term reproducibility.
引用
收藏
页码:729 / 740
页数:12
相关论文
共 20 条
[2]  
*INT AT EN AG, 1994, IAEA TECHN REP SER, V374, P15
[3]  
JAFFE G, 1906, J PHYSIQUE, V4, P262
[4]   PROPERTIES OF LIQUID IONIZATION CHAMBERS AT LDR BRACHYTHERAPY DOSE-RATES [J].
JOHANSSON, B ;
WICKMAN, G ;
HOLMSTROM, T .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) :575-587
[5]   General collection efficiency for liquid isooctane and tetramethylsilane used as sensitive media in a parallel-plate ionization chamber [J].
Johansson, B ;
Wickman, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (01) :133-145
[6]   General collection efficiency for liquid isooctane and tetramethylsilane in pulsed radiation [J].
Johansson, B ;
Wickman, G ;
BaharGogani, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (10) :1929-1938
[7]   General collection efficiency in liquid iso-octane and tetramethylsilane used as sensitive media in a thimble ionization chamber [J].
Johansson, BE ;
Bahar-Gogani, J ;
Wickman, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (09) :2161-2170
[8]  
Mie G, 1904, ANN PHYS-BERLIN, V13, P857
[9]   EFFECT OF AN EXTERNAL ELECTRIC-FIELD ON YIELD OF FREE IONS .2. INITIAL DISTRIBUTION OF ION-PAIRS IN LIQUID HYDROCARBONS [J].
MOZUMDER, A .
JOURNAL OF CHEMICAL PHYSICS, 1974, 60 (11) :4305-4310
[10]  
NILSSON U, 1997, 9714 SSI