Findings of an observational investigation of pure remote follow-up of pacemaker patients: is the in-clinic device check still needed?

被引:10
作者
Facchin, D. [1 ]
Baccillieri, M. S. [2 ]
Gasparini, G. [3 ]
Zoppo, F. [4 ]
Allocca, G. [5 ]
Brieda, M. [6 ]
Verlato, R. [2 ]
Proclemer, A. [1 ]
机构
[1] Azienda Sanitaria Univ Integrata Udine, Udine, Italy
[2] Osped P Cosma, Camposampiero, Italy
[3] Osped Angelo, Venice, Italy
[4] Azienda Osped Mirano, Mirano, Italy
[5] Osped Conegliano, Conegliano, Italy
[6] Osped Santa Maria Angeli, Pordenone, Italy
关键词
Pacemakers; Remote follow-up; Remote monitoring; Remote interrogation; IMPLANTABLE ELECTRONIC DEVICES; CARDIAC RESYNCHRONIZATION THERAPY; CARDIOVERTER-DEFIBRILLATORS; ATRIAL-FIBRILLATION; EXPERT CONSENSUS; CIEDS; TRIAL; MANAGEMENT; WIRELESS; EVENTS;
D O I
10.1016/j.ijcard.2016.06.162
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Device follow-up is mandatory in the care of patients with a pacemaker. However, in most cases, device checks appear to be mere technical, time-consuming procedures. The aim of this research is to evaluate whether remote follow-up can replace in-clinic device checks by assessing clinical outcomes for pacemaker patients followed only via remote follow-up. Methods and results: Consecutive pacemaker patients followed with remote monitoring were prospectively included by 6 Italian cardiology centers in an observational investigation. The workflow for remote monitoring included an initial assessment by nursing staff and, when necessary, by a responsible physician for medical decisions. No in-person visits were scheduled after the start of remote monitoring. One-thousand and two-hundred and fifty one patients (30% female, 75 +/- 11 years old) were followed for a median observation period of 15 months. Out of 4965 remote transmissions, 1882 (38%) had at least one clinically relevant event to be investigated further, but, only after 137 transmissions (2.8%), the patients were contacted for an in-clinic visit or hospitalization. Sixty-nine patients died and 124 were hospitalized for various reasons. Atrial fibrillation episodes were the most common clinical events discovered by remote transmissions, occurring in 1339 (26%) transmissions and 471 (38%) patients. Conclusions: Our experience shows that remote monitoring in a pacemaker population can safely replace in-clinic follow-up, avoiding unnecessary in-hospital device follow-up. (C) 2016 Published by Elsevier Ireland Ltd.
引用
收藏
页码:781 / 786
页数:6
相关论文
共 20 条
[1]   AF burden is important - fact or fiction? [J].
Boriani, G. ;
Diemberger, I. ;
Ziacchi, M. ;
Valzania, C. ;
Gardini, B. ;
Cimaglia, P. ;
Martignani, C. ;
Biffi, M. .
INTERNATIONAL JOURNAL OF CLINICAL PRACTICE, 2014, 68 (04) :444-452
[2]   2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy [J].
Brignole, Michele ;
Auricchio, Angelo ;
Baron-Esquivias, Gonzalo ;
Bordachar, Pierre ;
Boriani, Giuseppe ;
Breithardt, Ole-A ;
Cleland, John ;
Deharo, Jean-Claude ;
Delgado, Victoria ;
Elliott, Perry M. ;
Gorenek, Bulent ;
Israel, Carsten W. ;
Leclercq, Christophe ;
Linde, Cecilia ;
Mont, Llus ;
Padeletti, Luigi ;
Sutton, Richard ;
Vardas, Panos E. ;
Zamorano, Jose Luis ;
Achenbach, Stephan ;
Baumgartner, Helmut ;
Bax, Jeroen J. ;
Bueno, Hctor ;
Dean, Veronica ;
Deaton, Christi ;
Erol, Cetin ;
Fagard, Robert ;
Ferrari, Roberto ;
Hasdai, David ;
Hoes, Arno W. ;
Kirchhof, Paulus ;
Knuuti, Juhani ;
Kolh, Philippe ;
Lancellotti, Patrizio ;
Linhart, Ales ;
Nihoyannopoulos, Petros ;
Piepoli, Massimo F. ;
Ponikowski, Piotr ;
Sirnes, Per Anton ;
Tamargo, Juan Luis ;
Tendera, Michal ;
Torbicki, Adam ;
Wijns, William ;
Windecker, Stephan ;
Kirchhof, Paulus ;
Blomstrom-Lundqvist, Carina ;
Badano, Luigi P. ;
Aliyev, Farid ;
Bnsch, Dietmar ;
Baumgartner, Helmut .
EUROPACE, 2013, 15 (08) :1070-1118
[3]   Monitored atrial fibrillation duration predicts arterial embolic events in patients suffering from bradycardia and atrial fibrillation implanted with antitachycardia pacemakers [J].
Capucci, A ;
Santini, M ;
Padeletti, L ;
Gulizia, M ;
Botto, G ;
Boriani, G ;
Ricci, R ;
Favale, S ;
Zolezzi, F ;
Di Belardino, N ;
Molon, G ;
Drago, F ;
Villani, GQ ;
Mazzini, E ;
Vimercati, M ;
Grammatico, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (10) :1913-1920
[4]   The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial The Value of Wireless Remote Monitoring With Automatic Clinician Alerts [J].
Crossley, George H. ;
Boyle, Andrew ;
Vitense, Holly ;
Chang, Yanping ;
Mead, R. Hardwin .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 57 (10) :1181-1189
[5]   Clinical Benefits of Remote Versus Transtelephonic Monitoring of Implanted Pacemakers [J].
Crossley, George H. ;
Chen, Jane ;
Choucair, Wassim ;
Cohen, Todd J. ;
Gohn, Douglas C. ;
Ben Johnson, W. ;
Kennedy, Eleanor E. ;
Mongeon, Luc R. ;
Serwer, Gerald A. ;
Qiao, Hongyan ;
Wilkoff, Bruce L. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (22) :2012-2019
[6]   ISHNE/EHRA Expert Consensus on Remote Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs) [J].
Dubner, Sergio ;
Auricchio, Angelo ;
Steinberg, Jonathan S. ;
Vardas, Panos ;
Stone, Peter ;
Brugada, Josep ;
Piotrowicz, Ryszard ;
Hayes, David L. ;
Kirchhof, Paulus ;
Breithardt, Guenter ;
Zareba, Wojciech ;
Schuger, Claudio ;
Aktas, Mehmet K. ;
Chudzik, Michal ;
Mittal, Suneet ;
Varma, Niraj .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2012, 17 (01) :36-56
[7]   The Relationship Between Daily Atrial Tachyarrhythmia Burden From Implantable Device Diagnostics and Stroke Risk The TRENDS Study [J].
Glotzer, Taya V. ;
Daoud, Emile G. ;
Wyse, D. George ;
Singer, Daniel E. ;
Ezekowitz, Michael D. ;
Hilker, Christopher ;
Miller, Clayton ;
Qi, Dongfeng ;
Ziegler, Paul D. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2009, 2 (05) :474-480
[8]   Costs of remote monitoring vs. ambulatory follow-ups of implanted cardioverter defibrillators in the randomized ECOST study [J].
Guedon-Moreau, Laurence ;
Lacroix, Dominique ;
Sadoul, Nicolas ;
Clementy, Jacques ;
Kouakam, Claude ;
Hermida, Jean-Sylvain ;
Aliot, Etienne ;
Kacet, Salem .
EUROPACE, 2014, 16 (08) :1181-1188
[9]   Follow-up of patients with new cardiovascular implantable electronic devices: Is adherence to the experts' recommendations associated with improved outcomes? [J].
Hess, Paul L. ;
Mi, Xiaojuan ;
Curtis, Lesley H. ;
Wilkoff, Bruce L. ;
Hegland, Donald D. ;
Al-Khatib, Sana M. .
HEART RHYTHM, 2013, 10 (08) :1127-1133
[10]   Long-term risk of recurrent atrial fibrillation as documented by an Implantable monitoring device -: Implications for optimal patient care [J].
Israel, CW ;
Grönefeld, G ;
Ehrlich, JR ;
Li, YG ;
Hohnloser, SH .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 43 (01) :47-52