A Self-Aligned 45°-Tilted Two-Axis Scanning Micromirror for Side-View Imaging

被引:8
作者
Duan, Can [1 ]
Wang, Wei [1 ,2 ]
Zhang, Xiaoyang [1 ]
Zhou, Liang [1 ]
Pozzi, Antonio [3 ,4 ]
Xie, Huikai [1 ]
机构
[1] Univ Florida, Dept Elect & Comp Engn, Gainesville, FL 32611 USA
[2] Shanghai Jiao Tong Univ, Dept Micronano Elect, Shanghai 200240, Peoples R China
[3] Univ Florida, Dept Small Anim Clin Sci, Gainesville, FL 32611 USA
[4] Univ Zurich, Small Anim Surg Clin, CH-8057 Zurich, Switzerland
基金
美国国家科学基金会;
关键词
Microelectromechanical systems (MEMS) micromirrors; electrothermal actuation; optical coherence tomography; endoscopic imaging; endomicroscopy; silicon optical bench; OPTICAL COHERENCE TOMOGRAPHY; IN-VIVO; MEMS MIRROR; PROBE; ENDOSCOPE; CATHETER; SPEED;
D O I
10.1109/JMEMS.2016.2562011
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents a two-axis electrothermal single-crystal-silicon micromirror that is tilted 45 degrees out of plane on a silicon optical bench (SiOB). The SiOB provides mechanical support and electrical wiring to the tilted two-axis scanning mirror, as well as aligned trenches for assembling other optical components, such as optical fibers and GRIN lens. The preset tilting of the mirror plate is achieved via the bending of a set of stressed bimorph cantilevers. A stopper connected on the bending bimorphs reaches and is stopped by the adjacent silicon sidewall. The tilt angle can be precisely controlled by properly choosing the distance from the stopper to the silicon sidewall and the flexure bimorph length. The fabricated mirror plate is 0.72 mm x 0.72 mm and the footprint of the entire MEMS device is 2.22 mm x 1.25 mm. The measured maximum total optical scan angle of the mirror is approximately 40 degrees in both x-axis and y-axis at only 5.5 Vdc. This technology will enable a new class of more compact microendoscopic optical imaging probes for in vivo early cancer detection. [2015-0315]
引用
收藏
页码:799 / 811
页数:13
相关论文
共 28 条
  • [11] An Electrothermal Tip-Tilt-Piston Micromirror Based on Folded Dual S-Shaped Bimorphs
    Jia, Kemiao
    Pal, Sagnik
    Xie, Huikai
    [J]. JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2009, 18 (05) : 1004 - 1015
  • [12] Three-dimensional endoscopic optical coherence tomography by use of a two-axis microelectromechanical scanning mirror
    Jung, W
    McCormick, DT
    Zhang, J
    Wang, L
    Tien, NC
    Chen, ZP
    [J]. APPLIED PHYSICS LETTERS, 2006, 88 (16)
  • [13] In vivo three-dimensional spectral domain endoscopic optical coherence tomography using a microelectromechanical system mirror
    Jung, Woonggyu
    McCormick, Daniel T.
    Ahn, Yeh-Chan
    Sepehr, Ali
    Brenner, Matt
    Wong, Brian
    Tien, Norman C.
    Chen, Zhongping
    [J]. OPTICS LETTERS, 2007, 32 (22) : 3239 - 3241
  • [14] Two-axis magnetically-driven MEMS scanning catheter for endoscopic high-speed optical coherence tomography
    Kim, Ki Hean
    Park, B. Hyle
    Maguluri, Gopi N.
    Lee, Tom W.
    Rogomentich, Fran J.
    Bancu, Mirela G.
    Bouma, Brett E.
    de Boer, Johannes F.
    Bernstein, Jonathan J.
    [J]. OPTICS EXPRESS, 2007, 15 (26): : 18130 - 18140
  • [15] Imaging needle for optical coherence tomography
    Li, XD
    Chudoba, C
    Ko, T
    Pitris, C
    Fujimoto, JG
    [J]. OPTICS LETTERS, 2000, 25 (20) : 1520 - 1522
  • [16] Rapid-scanning forward-imaging miniature endoscope for real-time optical coherence tomography
    Liu, XM
    Cobb, MJ
    Chen, YC
    Kimmey, MB
    Li, XD
    [J]. OPTICS LETTERS, 2004, 29 (15) : 1763 - 1765
  • [17] MEMS micromirror integrated endoscopic probe for optical coherence tomography bioimaging
    Mu, Xiaojing
    Sun, Winston
    Feng, Hanhua
    Yu, Aibin
    Chen, Kelvin Wei Sheng
    Fu, Chit Yaw
    Olivo, Malini
    [J]. SENSORS AND ACTUATORS A-PHYSICAL, 2011, 168 (01) : 202 - 212
  • [18] Pan Y., 2011, OPT LETT, V26, P1966
  • [19] In situ imaging of lung alveoli with an optical coherence tomography needle probe
    Quirk, Bryden C.
    McLaughlin, Robert A.
    Curatolo, Andrea
    Kirk, Rodney W.
    Noble, Peter B.
    Sampson, David D.
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (03)
  • [20] A 2.8-mm Imaging Probe Based On a High-Fill-Factor MEMS Mirror and Wire-Bonding-Free Packaging for Endoscopic Optical Coherence Tomography
    Samuelson, Sean R.
    Wu, Lei
    Sun, Jingjing
    Choe, Se-woon
    Sorg, Brian S.
    Xie, Huikai
    [J]. JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2012, 21 (06) : 1291 - 1302