Three-Year Outcomes From the Harmony Native Outflow Tract Early Feasibility Study

被引:49
作者
Benson, Lee N. [1 ]
Gillespie, Matthew J. [2 ]
Bergersen, Lisa [3 ]
Cheatham, Sharon L. [4 ]
Hor, Kan N. [4 ]
Horlick, Eric M. [5 ]
Weng, Shicheng [6 ]
Mchenry, Brian T. [6 ]
Osten, Mark D. [5 ]
Powell, Andrew J. [3 ]
Cheatham, John P. [4 ]
机构
[1] Hosp Sick Children, Labatt Family Heart Ctr, Toronto, ON, Canada
[2] Childrens Hosp Philadelphia, Philadelphia, PA 19104 USA
[3] Boston Childrens Hosp, Boston, MA USA
[4] Nationwide Childrens Hosp, Columbus, OH USA
[5] Univ Hlth Network, Toronto Gen Hosp, Toronto, ON, Canada
[6] Medtronic, Mounds View, MN USA
关键词
congenital heart disease; echocardiography; pulmonary valve; stent; Tetralogy of Fallot; tomography; PULMONARY VALVE-REPLACEMENT; RIGHT-VENTRICLE; DACRON CONDUIT;
D O I
10.1161/CIRCINTERVENTIONS.119.008320
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The Harmony transcatheter pulmonary valve (TPV) was designed for treatment of postoperative pulmonary valve regurgitation in patients with repaired right ventricular outflow tracts. Methods: The Native TPV EFS (Early Feasibility Study) is a prospective, multicenter, nonrandomized feasibility study. Three-year outcomes are reported. Results: Of 20 implanted patients, 17 completed 3-year follow-up (maximum: 4.1 years). There were no deaths and 2 early explants. One patient did not complete the 3-year visit. In patients with available 3-year echocardiographic data, 1 had a mild paravalvular leak and the rest had none/trace; 1 patient had mild pulmonary valve regurgitation and the remainder had none/trace. The 3-year mean right ventricular outflow tract echocardiographic gradient was 15.7 +/- 5.5 mm Hg. Radiographically, no late frame fractures or erosions were identified. At 2 years, 2 patients presented with an increased echocardiographic outflow gradient (1 mixed lesion with moderate/severe pulmonary valve regurgitation). Computed tomography scans identified neointimal tissue ingrowth within the stent frame in both patients, and they were treated successfully with a transcatheter valve-in-valve procedure (Melody TPV). Additional follow-up computed tomography scans performed at 3.2 +/- 0.5 years after implant were obtained in 16 patients and revealed luminal tissue thickening at the inflow and outflow portion of the frame with no significant alteration of the valve housing. Conclusions: Three-year results from the Native TPV EFS revealed stable Harmony TPV device position, good valve function in most, and the absence of moderate/severe paravalvular leak and significant late frame fractures. Two patients developed significant neointimal proliferation requiring valve-in-valve treatment, while all others had no clinically significant right ventricular outflow tract obstruction. VISUAL OVERVIEW: A visual overview is available for this article.
引用
收藏
页数:10
相关论文
共 17 条
  • [1] Alkashkari W, 2018, J STRUCT HEART DIS, V4, P1
  • [2] [Anonymous], 2019, MEDTRONIC HARMONY TR
  • [3] The Performance of Hancock Porcine-Valved Dacron Conduit for Right Ventricular Outflow Tract Reconstruction
    Belli, Emre
    Salihoglu, Ece
    Leobon, Bertrand
    Roubertie, Francois
    Ly, Mohammed
    Roussin, Regine
    Serraf, Alain
    [J]. ANNALS OF THORACIC SURGERY, 2010, 89 (01) : 152 - 158
  • [4] Harmony Feasibility Trial Acute and Short-Term Outcomes With a Self-Expanding Transcatheter Pulmonary Valve
    Bergersen, Lisa
    Benson, Lee N.
    Gillespie, Matthew J.
    Cheatham, Sharon L.
    Crean, Andrew M.
    Hor, Kan N.
    Horlick, Eric M.
    Lung, Te-Hsin
    McHenry, Brian T.
    Osten, Mark D.
    Powell, Andrew J.
    Cheatham, John P.
    [J]. JACC-CARDIOVASCULAR INTERVENTIONS, 2017, 10 (17) : 1763 - 1773
  • [5] Percutaneous replacement of pulmonary valve in a right-ventricle to pulmonary-artery prosthetic conduit with valve dysfunction
    Bonhoeffer, P
    Boudjemline, Y
    Saliba, Z
    Merckx, J
    Aggoun, Y
    Bonnet, D
    Acar, P
    Le Bidois, J
    Sidi, D
    Kachaner, J
    [J]. LANCET, 2000, 356 (9239) : 1403 - 1405
  • [6] Remodelling of the right ventricle after early pulmonary valve replacement in children with repaired tetralogy of Fallot: assessment by cardiovascular magnetic resonance
    Buechel, ERV
    Dave, HH
    Kellenberger, CJ
    Dodge-Khatami, A
    Pretre, R
    Berger, F
    Bauersfeld, U
    [J]. EUROPEAN HEART JOURNAL, 2005, 26 (24) : 2721 - 2727
  • [7] Biventricular response after pulmonary valve replacement for right ventricular outflow tract dysfunction - Is age a predictor of outcome?
    Frigiola, Alessandra
    Tsang, Victor
    Bull, Catherine
    Coats, Louise
    Khambadkone, Sachin
    Derrick, Graham
    Mist, Bryan
    Walker, Fiona
    van Doorn, Carin
    Bonhoeffer, Philipp
    Taylor, Andrew M.
    [J]. CIRCULATION, 2008, 118 (14) : S182 - S190
  • [8] Pressure Recovery Explains Doppler Overestimation of Invasive Pressure Gradient across Segmental Vascular Stenosis
    Giardini, Alessandro
    Tacy, Theresa A.
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2010, 27 (01): : 21 - 31
  • [9] Patient Selection Process for the Harmony Transcatheter Pulmonary Valve Early Feasibility Study
    Gillespie, Matthew J.
    Benson, Lee N.
    Bergersen, Lisa
    Bacha, Emile A.
    Cheatham, Sharon L.
    Crean, Andrew M.
    Eicken, Andreas
    Ewert, Peter
    Geva, Tal
    Hellenbrand, William E.
    Hor, Kan N.
    Horlick, Eric M.
    Jones, Thomas K.
    Mayer, John
    McHenry, Brian T.
    Osten, Mark D.
    Powell, Andrew J.
    Zahn, Evan M.
    Cheatham, John P.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2017, 120 (08) : 1387 - 1392
  • [10] Aortic root compression during transcatheter pulmonary valve replacement
    Lindsay, Ian
    Aboulhosn, Jamil
    Salem, Morris
    Levi, Daniel
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2016, 88 (05) : 814 - 821