Close-packed arrays of transition-edge x-ray microcalorimeters with high spectral resolution at 5.9 keV

被引:42
作者
Iyomoto, N. [1 ,2 ]
Bandler, S. R. [1 ,3 ]
Brekosky, R. P. [1 ,4 ]
Brown, A. -D. [1 ]
Chervenak, J. A. [1 ]
Finkbeiner, F. M. [1 ,5 ]
Kelley, R. L. [1 ]
Kilbourne, C. A. [1 ]
Porter, F. S. [1 ]
Sadleir, J. E. [1 ]
Smith, S. J. [1 ]
Figueroa-Feliciano, E. [6 ]
机构
[1] NASA, Goddard Space Flight Ctr, Greenbelt, MD 20771 USA
[2] Johns Hopkins Univ, Baltimore, MD 21218 USA
[3] Univ Maryland, College Pk, MD 20742 USA
[4] Northrop Grumman Corp, Lanham, MD 20706 USA
[5] RS Informat Syst Inc, Mclean, VA 22102 USA
[6] MIT, Dept Phys, Cambridge, MA 02139 USA
基金
美国国家航空航天局;
关键词
D O I
10.1063/1.2830665
中图分类号
O59 [应用物理学];
学科分类号
摘要
We present measurements of high fill-factor arrays of superconducting transition-edge x-ray microcalorimeters designed to provide rapid thermalization of the x-ray energy. We designed an x-ray absorber that is cantilevered over the sensitive part of the thermometer itself, making contact only at normal-metal features. With absorbers made of electroplated gold, we have demonstrated an energy resolution between 2.4 and 3.1 eV at 5.9 keV on 13 separate pixels. We have determined the thermal and electrical parameters of the devices throughout the superconducting transition and, using these parameters, have modeled all aspects of the detector performance. (c) 2008 American Institute of Physics.
引用
收藏
页数:3
相关论文
共 9 条
[1]   K alpha(1,2) and K beta(1,3) x-ray emission lines of the 3d transitions metals [J].
Holzer, G ;
Fritsch, M ;
Deutsch, M ;
Hartwig, J ;
Forster, E .
PHYSICAL REVIEW A, 1997, 56 (06) :4554-4568
[2]  
HOLZER G, COMMUNICATION
[3]   Thermodynamics of nonlinear bolometers near equilibrium [J].
Irwin, KD .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 559 (02) :718-720
[4]   X-ray detection using a superconducting transition-edge sensor microcalorimeter with electrothermal feedback [J].
Irwin, KD ;
Hilton, GC ;
Wollman, DA ;
Martinis, JM .
APPLIED PHYSICS LETTERS, 1996, 69 (13) :1945-1947
[5]   High-density arrays of x-ray microcalorimeters for Constellation-X [J].
Kilbourne, Caroline A. ;
Bandler, Simon R. ;
Brown, Ari D. ;
Chervenak, James A. ;
Figueroa-Feliciano, Enectali ;
Finkbeiner, Fred M. ;
Iyomoto, Naoko ;
Kelley, Richard L. ;
Porter, F. Scott ;
Saab, Tarek ;
Sadleir, John ;
White, Jennifer .
SPACE TELESCOPES AND INSTRUMENTATION II: ULTRAVIOLET TO GAMMA RAY, PTS 1 AND 2, 2006, 6266
[6]   Impedance measurements and modeling of a transition-edge-sensor calorimeter [J].
Lindeman, MA ;
Bandler, S ;
Brekosky, RP ;
Chervenak, JA ;
Figueroa-Feliciano, E ;
Finkbeiner, FM ;
Li, MJ ;
Kilbourne, CA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2004, 75 (05) :1283-1289
[7]   ADVANCES TOWARD HIGH SPECTRAL RESOLUTION QUANTUM X-RAY CALORIMETRY [J].
MOSELEY, SH ;
KELLEY, RL ;
SCHOELKOPF, RJ ;
SZYMKOWIAK, AE ;
MCCAMMON, D ;
ZHANG, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :59-64
[8]   Optimized transition-edge x-ray microcalorimeter with 2.4 eV energy resolution at 5.9 keV [J].
Ullom, JN ;
Beall, JA ;
Doriese, WB ;
Duncan, WD ;
Ferreira, L ;
Hilton, GC ;
Irwin, KD ;
Reintsema, CD ;
Vale, LR .
APPLIED PHYSICS LETTERS, 2005, 87 (19) :1-3
[9]   Characterization and reduction of unexplained noise in superconducting transition-edge sensors [J].
Ullom, JN ;
Doriese, WB ;
Hilton, GC ;
Beall, JA ;
Deiker, S ;
Duncan, WD ;
Ferreira, L ;
Irwin, KD ;
Reintsema, CD ;
Vale, LR .
APPLIED PHYSICS LETTERS, 2004, 84 (21) :4206-4208