Catastrophic wear in a metal-on-ceramic total hip arthroplasty

被引:16
作者
Ramon Valenti, Juan [1 ]
Del Rio, Javier [1 ]
Amillo, Santiago [1 ]
机构
[1] Univ Navarra Clin, Dept Orthoped Surg & Traumatol, Navarra, Spain
关键词
wear; bearing surface; total hip arthroplasty; ceramic-on-metal; metal-on-metal;
D O I
10.1016/j.arth.2006.06.008
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A 51-year-old woman came to our clinic 6 months after a right total hip arthroplasty. She had noticed a slowly growing mass in the proximal thigh and referred progressive pain in the right groin. Plain radiography revealed premature acetabular cup aseptic loosening, and in the computed tomography study, a 14-cm-diameter mass was observed. Revision surgery was performed, showing a metal-on-ceramic bearing surface. The histologic analysis of surrounding tissues was reported as massive metallosis. Although occasionally chosen for primary or revision hip arthroplasty, there is little information available about the in vivo wear behavior of this combination. This important fact should be taken into account before considering such a surface alternative.
引用
收藏
页码:920 / 922
页数:3
相关论文
共 13 条
[1]  
Campbell P, 2004, CLIN ORTHOP RELAT R, P98
[2]   A novel low wearing differential hardness, ceramic-on-metal hip joint prosthesis [J].
Firkins, PJ ;
Tipper, JL ;
Ingham, E ;
Stone, MH ;
Farrar, R ;
Fisher, J .
JOURNAL OF BIOMECHANICS, 2001, 34 (10) :1291-1298
[3]   Metal sensitivity in patients with orthopaedic implants [J].
Hallab, N ;
Merritt, K ;
Jacobs, JJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (03) :428-436
[4]   Can metal levels be used to monitor metal-on-metal hip arthroplasties? [J].
Jacobs, JJ ;
Skipor, AK ;
Campbell, PA ;
Hallab, NJ ;
Urban, RM ;
Amstutz, HC .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :59-65
[5]   Massive wear of an incompatible metal-on-metal articulation in total hip arthroplasty [J].
Keel, JB ;
Kuster, MS .
JOURNAL OF ARTHROPLASTY, 2004, 19 (05) :638-642
[6]  
KNAHR K, 2001, WORLD TRIBOLOGY FORU, P67
[7]   In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty [J].
Komistek, RD ;
Dennis, DA ;
Ochoa, JA ;
Haas, BD ;
Hammill, C .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10) :1836-1841
[8]   Four-year study of cobalt and chromium blood levels in patients managed with two different metal-on-metal total hip replacements [J].
Lhotka, C ;
Szekeres, T ;
Steffan, I ;
Zhuber, K ;
Zweymüller, K .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2003, 21 (02) :189-195
[9]   Microseparation of the centers of alumina-alumina artificial hip joints during simulator testing produces clinically relevant wear rates and patterns [J].
Nevelos, J ;
Ingham, E ;
Doyle, C ;
Streicher, R ;
Nevelos, A ;
Walter, W ;
Fisher, J .
JOURNAL OF ARTHROPLASTY, 2000, 15 (06) :793-795
[10]   Severe wear and fracture of zirconia heads against alumina inserts in hip simulator studies with microseparation [J].
Stewart, TD ;
Tipper, JL ;
Insley, G ;
Streicher, RM ;
Ingham, E ;
Fisher, J .
JOURNAL OF ARTHROPLASTY, 2003, 18 (06) :726-734