The fixation tines of the Micra™ leadless pacemaker are atraumatic to the tricuspid valve

被引:8
作者
Mattson, Alexander R. [1 ,2 ,3 ,4 ]
Sanchez, Jorge D. Zhingre [1 ,2 ,3 ]
Iaizzo, Paul A. [1 ,2 ,3 ]
机构
[1] Univ Minnesota, Dept Biomed Engn, Minneapolis, MN USA
[2] Univ Minnesota, Dept Surg, Box 242 UMHC, Minneapolis, MN 55455 USA
[3] Univ Minnesota, Inst Engn Med, Minneapolis, MN USA
[4] Medtronic, Mounds View, MN USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2018年 / 41卷 / 12期
关键词
biomechanical failure properties; cardiac anatomy; extraction; Micra pacemaker; retrieval; tricuspid valve device interactions; MANAGEMENT; RETRIEVAL; SYSTEM;
D O I
10.1111/pace.13529
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Today, there is no manufacturer-supplied retrieval tool for the Micra (TM) pacemaker (Medtronic, Minneapolis, MN, USA); therefore, off-the-shelf catheters have been employed for retrievals. The proximal retrieval feature of the Micra (TM) can be snared and the device is then retracted from the myocardium, pulling the device through the tricuspid valve. This study characterizes the potential risks of Micra (TM) nitinol tine engagement with the tricuspid sub-valvular apparatus. Methods Fresh human hearts nonviable for transplant (n = 10) were obtained from our regional organ procurement agency (LifeSource, Minneapolis, MN, USA). Micra (TM) fixation tines were affixed to a linear force transducer. Tines were then engaged in tricuspid chordae tendineae to conduct a constant velocity tensile test. Each test was run until tines disengaged from the chordae tendineae or until they released from the valve apparatus. Subsequently, biomechanical failure properties of the valve apparatus and isolated chordae tendineae were determined using a series of uniaxial tensile tests. Results There were no chordal ruptures observed during our Micra (TM) tine extraction testing. Chordal failure required 15.0 times the force of extracting a single engaged tine, and 9.0 times the force of extracting two engaged tines. The uniaxial stresses required for isolated chordal failure averaged 17.4 N/mm(2); failure strains exceeded 150% resting chordal length. Conclusions The forces required to rupture tricuspid chordae tendineae significantly exceeded the forces potentially imposed on the chordae during Micra (TM) device retrievals. We conclude that the fixation tines of the Micra (TM) device are unlikely to damage the tricuspid apparatus during either implant or retrieval.
引用
收藏
页码:1606 / 1610
页数:5
相关论文
共 9 条
[1]   Techniques for successful early retrieval of the Micra transcatheter pacing system: A worldwide experience [J].
Afzal, Muhammad R. ;
Daoud, Emile G. ;
Cunnane, Ryan ;
Mulpuru, Shiva K. ;
Koay, Alan ;
Hussain, Azlan ;
Omar, Razali ;
Wei, Koh Kok ;
Amin, Anish ;
Kidwell, Gregory ;
Patel, Nirav ;
Love, Charles ;
Lloyd, Michael ;
Sterlinski, Maciej ;
Goldbarg, Seth ;
Leal, Miguel A. ;
Gabriels, James ;
Patel, Apoor ;
Jadonath, Ram ;
Grubman, Eric ;
Crossley, George ;
Pepper, Chris ;
Lakkireddy, Dhanunjaya ;
Okabe, Toshimasa ;
Hummel, John D. ;
Augostini, Ralph S. .
HEART RHYTHM, 2018, 15 (06) :841-846
[2]  
Beurskens NEG, 2017, ARRHYTH ELECTROPHYSI, V6, P129, DOI 10.15420/aer.2017:16:1
[3]   Isolated four-chamber working swine heart model [J].
Chinchoy, E ;
Soule, CL ;
Houlton, AJ ;
Gallagher, WJ ;
Hjelle, MA ;
Laske, TG ;
Morissette, J ;
Iaizzo, PA .
ANNALS OF THORACIC SURGERY, 2000, 70 (05) :1607-1614
[4]   Retrieval of Medtronic Micra Transcatheter Pacing System after tether removal [J].
Gerdes, Christian ;
Nielsen, Jens Cosedis .
EUROPACE, 2016, 18 (08) :1202-1202
[5]   To retrieve, or not to retrieve: System revisions with the Micra transcatheter pacemaker [J].
Grubman, Eric ;
Ritter, Philippe ;
Ellis, Christopher R. ;
Giocondo, Michael ;
Augostini, Ralph ;
Neuzil, Petr ;
Ravindran, Bipin ;
Patel, Anshul M. ;
Omdahl, Pamela ;
Pieper, Karen ;
Stromberg, Kurt ;
Hudnall, J. Harrison ;
Reynolds, Dwight .
HEART RHYTHM, 2017, 14 (12) :1801-1806
[6]   A worldwide experience of the management of battery failures and chronic device retrieval of the Nanostim leadless pacemaker [J].
Lakkireddy, Dhanunjaya ;
Knops, Reinoud ;
Atwater, Brett ;
Neuzil, Petr ;
Ip, John ;
Gonzalez, Elkin ;
Friedman, Paul ;
Defaye, Pascal ;
Exner, Derek ;
Aonuma, Kazutaka ;
Doshi, Rahul ;
Sperzel, Johannes ;
Reddy, Vivek .
HEART RHYTHM, 2017, 14 (12) :1756-1763
[7]   Retrieval of the Leadless Cardiac Pacemaker A Multicenter Experience [J].
Reddy, Vivek Y. ;
Miller, Marc A. ;
Knops, Reinoud E. ;
Neuzil, Petr ;
Defaye, Pascal ;
Jung, Werner ;
Doshi, Rahul ;
Castellani, Mark ;
Strickberger, Adam ;
Mead, R. Hardwin ;
Doppalapudi, Harish ;
Lakkireddy, Dhanunjaya ;
Bennett, Matthew ;
Sperzel, Johannes .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2016, 9 (12)
[8]   A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry [J].
Roberts, Paul R. ;
Clementy, Nicolas ;
Al Samadi, Faisal ;
Garweg, Christophe ;
Martinez-Sande, Jose Luis ;
Iacopino, Saverio ;
Johansen, Jens Brock ;
Prat, Xavier Vinolas ;
Kowal, Robert C. ;
Klug, Didier ;
Mont, Lluis ;
Steffel, Jan ;
Li, Shelby ;
Van Osch, Dirk ;
El-Chami, Mikhael F. .
HEART RHYTHM, 2017, 14 (09) :1375-1379
[9]  
Vatterott Pierce J, 2018, HeartRhythm Case Rep, V4, P167, DOI 10.1016/j.hrcr.2017.11.014