Full-field digital mammography: Dose-dependent detectability of simulated breast lesions

被引:27
作者
Obenauer, S [1 ]
Hermann, KP [1 ]
Schorn, C [1 ]
Fischer, U [1 ]
Grabbe, E [1 ]
机构
[1] Univ Gottingen Klinikum, Rontgendiagnost Abt 1, D-37075 Gottingen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2000年 / 172卷 / 12期
关键词
full-field digital mammography; flat detector; dose reduction; contrast-detail detectability; microcalcifications;
D O I
10.1055/s-2000-9222
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: The study compares contrast-detail and microcalcification detectability of a full-field digital mammography (FFDM) to a state-of-the-art conventional screen-film mammography (SFM) by using different doses in the digital system. Materials and Methods: The investigations were performed with an FFDM (Senographe 2000 D, GEMS) and an SFM system (Senographe DMR, GEMS) using a contrast-detail mammography phantom (CDMAM) and an anthropomorphic breast phantom with superimposed microcalcifications. The digital detector was exposed with standard dose of SFM and with a dose reduction of up to 75%. Contrast-detail curves and correct observation ratio (COR) were performed for the CDMAM phantom. ROC analysis with a confidence level ranging from 1 to 5 was done with the results of the anthropomorphic phantom. Results: Digital mammography with the same dose revealed at least an equivalent or even higher detectability rate than conventional mammography, COR could be increased at about 10-25%. The ROC analysis yielded better results for the FFDM system. The same lesion detectability in digital mammography as in the conventional method was reached at a dose reduction of about 25%, concerning spot views even at higher reduction. Dose reduction in the anthropomorphic phantom resulted in a linear loss of detectability. The same detectability as in conventional mammography was reached, however, by a dose reduction of about 50%. Conclusion: The results suggest that FFDM is at least equivalent to or - as far as spot views are concerned - superior to conventional SFM concerning the detectability of simulated lesions. Thus, a potential of dose reduction is suggested.
引用
收藏
页码:1052 / 1056
页数:5
相关论文
共 14 条
[1]   ANALYSIS OF VARIATIONS IN CONTRAST-DETAIL EXPERIMENTS [J].
COHEN, G ;
MCDANIEL, DL ;
WAGNER, LK .
MEDICAL PHYSICS, 1984, 11 (04) :469-473
[2]   Direct digital mammography image acquisition [J].
Cowen, AR ;
Parkin, GJS ;
Hawkridge, P .
EUROPEAN RADIOLOGY, 1997, 7 (06) :918-930
[3]  
FEIG SA, 1995, RADIOL CLIN N AM, V33, P1205
[4]   Image quality and radiation exposure in digital storage plate mammography with magnification technique. [J].
Fiedler, E ;
Aichinger, U ;
Böhner, C ;
Säbel, M ;
Schulz-Wendtland, R ;
Bautz, W .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 171 (01) :60-64
[5]  
FRIEDRICH M, 1993, RADIOLOGE, V33, P243
[6]   Magnification survey and spot view mammography with a new microfocus X-ray unit: detail resolution and radiation exposure [J].
Funke, M ;
Breiter, N ;
Hermann, KP ;
Oestmann, JW ;
Grabbe, E .
EUROPEAN RADIOLOGY, 1998, 8 (03) :386-390
[7]   DIMA-magnification mammography: use of phantoms for the investigation of image quality and radiation exposure. [J].
Hermann, KP ;
Funke, M ;
Bruggemeyer, H ;
Sperner, W ;
Gruhl, T ;
Koch, R .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 167 (03) :304-310
[8]   Digital mammography in direct magnification technique using a large-area amorphous silicon X-ray detector. [J].
Hermann, KP ;
Hundertmark, C ;
Funke, M ;
von Brenndorff, A ;
Grabbe, E .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1999, 170 (05) :503-506
[9]   SPATIAL-RESOLUTION IN DIGITAL MAMMOGRAPHY [J].
KARSSEMEIJER, N ;
FRIELING, JTM ;
HENDRIKS, JHCL .
INVESTIGATIVE RADIOLOGY, 1993, 28 (05) :413-419
[10]  
Obenauer S, 2000, ROFO-FORTSCHR RONTG, V172, P646