Experimental determination of the effective point of measurement for various detectors used in photon and electron beam dosimetry

被引:33
作者
Looe, Hui Khee [1 ,2 ]
Harder, Dietrich [3 ]
Poppe, Bjoern [1 ,2 ]
机构
[1] Pius Hosp, Clin Radiat Therapy, Oldenburg, Germany
[2] Carl von Ossietzky Univ Oldenburg, WG Med Radiat Phys, Oldenburg, Germany
[3] Univ Gottingen, Gottingen, Germany
关键词
REPLACEMENT CORRECTION FACTORS; IONIZATION CHAMBERS; ION CHAMBERS; PERTURBATION FACTORS;
D O I
10.1088/0031-9155/56/14/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The subject of this study is the 'shift of the effective point of measurement', Delta z, well known as a method of correction compensating for the 'displacement effect' in photon and electron beam dosimetry. Radiochromic EBT 1 films have been used to measure the 'true' TPR curves of 6 and 15 MV photons and 6 and 9 MeV electrons in the solid water-equivalent material RW3. For the Roos and Markus chambers, the cylindrical 'PinPoint', 'Semiflex' and 'Rigid-Stem' chambers, the 2D-Array and the E-type silicon diode (all from PTW-Freiburg), the positions of the effective points of measurement have been determined by direct or indirect comparison between their TPR curves and those of the EBT 1 film. Both for the Roos and Markus chambers, we found Delta z = (0.4 +/- 0.1) mm, which confirms earlier experimental and Monte Carlo results, but means a shortcoming of the 'water-equivalent window thickness' formula. For the cylindrical chambers, the ratio Delta z/r was observed to increase with r, confirming a recent Monte Carlo prediction by Tessier (2010 E2-CN-182, Paper no 147, IDOS, Vienna) as well as the experimental observations by Johansson et al (1978 IAEA Symp. Proc. (Vienna) IAEA-SM-222/35 pp 243-70). According to a theoretical consideration, the shift of the effective point of measurement from the reference point of the detector is caused by a gradient of the fluence of the ionizing particles. As the experiments have shown, the value of Delta z depends on the construction of the detector, but remains invariant under changes of radiation quality and depth. Other disturbances, which do not belong to the class of 'gradient effects', are not corrected by shifting the effective point of measurement.
引用
收藏
页码:4267 / 4290
页数:24
相关论文
共 40 条
[1]  
[Anonymous], 2007, HDB RADIOTHERAPY PHY, DOI DOI 10.1201/9781420012026
[2]  
[Anonymous], 2008, 68002 DIN
[3]   Monte Carlo calculations of correction factors for plane-parallel ionization chambers in clinical electron dosimetry [J].
Araki, Fujio .
MEDICAL PHYSICS, 2008, 35 (09) :4033-4040
[4]   A NEW LOOK AT DISPLACEMENT FACTOR AND POINT OF MEASUREMENT CORRECTIONS IN IONIZATION-CHAMBER DOSIMETRY [J].
AWSCHALOM, M ;
ROSENBERG, I ;
TENHAKEN, RK .
MEDICAL PHYSICS, 1983, 10 (03) :307-313
[5]  
BJARNGARD BE, 1985, MED PHYS, V12, P785, DOI 10.1118/1.595665
[6]   Ionization chamber gradient effects in nonstandard beam configurations [J].
Bouchard, Hugo ;
Seuntjens, Jan ;
Carrier, Jean-Francois ;
Kawrakow, Iwan .
MEDICAL PHYSICS, 2009, 36 (10) :4654-4663
[7]  
BRUGGMOSER G, 2007, DREIL MED 2007 BERN
[8]   Wall correction factors, Pwall, for parallel-plate ionization chambers [J].
Buckley, Lesley A. ;
Rogers, D. W. O. .
MEDICAL PHYSICS, 2006, 33 (06) :1788-1796
[9]  
Harder D, 1966, Strahlentherapie Sonderb, V62, P254
[10]  
HARDER D, 1965, S HIGH ENERGY ELECTR, P26