Safety of endoscopic procedures with monopolar versus bipolar instruments in an ex vivo porcine model

被引:16
|
作者
Shinmura, Kensuke [1 ]
Ikematsu, Hiroaki [1 ]
Kojima, Motohiro [2 ,3 ]
Nakamura, Hiroshi [1 ]
Osera, Shozo [1 ]
Yoda, Yusuke [1 ]
Hori, Keisuke [1 ]
Oono, Yasuhiro [1 ]
Ochiai, Atsushi [2 ,3 ]
Yano, Tomonori [1 ]
机构
[1] Natl Canc Ctr Hosp East, Dept Gastroenterol & Endoscopy, 6-5-1 Kashiwanoha, Kashiwa, Chiba 2778577, Japan
[2] Natl Canc Ctr Hosp East, Dept Pathol, 6-5-1 Kashiwanoha, Kashiwa, Chiba 2778577, Japan
[3] Natl Canc Ctr Hosp East, Clin Labs, 6-5-1 Kashiwanoha, Kashiwa, Chiba 2778577, Japan
关键词
Colorectal endoscopic resection; Monopolar instrument; Bipolar instrument; SMALL COLORECTAL POLYPS; COLONOSCOPIC POLYPECTOMY; MUCOSAL RESECTION; COLD SNARE; COAGULATION; FORCEPS; COMPLICATIONS; HOT; ELECTROCOAGULATION; PREVENTION;
D O I
10.1186/s12876-020-1176-9
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background Monopolar instruments are generally used in colorectal endoscopic mucosal resection (EMR). Bipolar instruments have previously been reported to be as safe as monopolar instruments. We sought to compare the safety of the monopolar and bipolar snare and hemostatic forceps in an animal model. Methods We created 5-mm, 10-mm, and 15-mm target lesions on an ex vivo porcine rectum. Two lesions of each size were resected via monopolar polypectomy (M-P), monopolar EMR (M-E), bipolar polypectomy (B-P), and bipolar EMR (B-E). We performed a pathological evaluation of the conditions of perforation and the effects of burning on the tissues. In addition, we burned the muscularis propria covered with submucosal layer using monopolar and bipolar hemostatic forceps and performed pathological evaluations. Results Polypectomy and EMR were performed in a total of 24 target lesions. A perforation was found on histology in one case of M-P and one case of M-E after removing target lesions of 15 mm in diameter. There were no perforations during endoscopic resection using the bipolar snare. The thermal denaturation in B-P did not reach the muscularis propria layer regardless of the size of the target lesion. Although thermal damage after using monopolar hemostatic forceps was extensive, thermal denaturation was only seen on the surface of the submucosal layer when bipolar hemostatic forceps were used. Conclusions Bipolar instruments cause less damage to the tissue than monopolar instruments. Our results also suggest that bipolar instruments may be safer than monopolar instruments in endoscopic procedures for colorectal lesions.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Safety of endoscopic procedures with monopolar versus bipolar instruments in an ex vivo porcine model
    Kensuke Shinmura
    Hiroaki Ikematsu
    Motohiro Kojima
    Hiroshi Nakamura
    Shozo Osera
    Yusuke Yoda
    Keisuke Hori
    Yasuhiro Oono
    Atsushi Ochiai
    Tomonori Yano
    BMC Gastroenterology, 20
  • [2] The Evaluation of Safety Regarding Endoscopic Resection Using Monopolar Versus Bipolar snare in Porcine Rectum
    Shinmura, Kensuke
    Ikematsu, Hiroaki
    Kojima, Motohiro
    Nakamura, Hiroshi
    Kato, Tomoji
    Osera, Shozo
    Yoda, Yusuke
    Oono, Yasuhiro
    Yano, Tomonori
    Ochiai, Atsushi
    Kaneko, Kazuhiro
    GASTROINTESTINAL ENDOSCOPY, 2015, 81 (05) : AB287 - AB287
  • [3] COMPARISON OF MICROWAVE WITH MONOPOLAR AND BIPOLAR COAGULATION IN A PORCINE MODEL
    Tsiamoulos, Z. P.
    Hancock, C.
    Sibbons, P. D.
    Saunders, B. P.
    GUT, 2014, 63 : A153 - A154
  • [4] EFFICACY AND SAFETY OF A MAGNET TRACTION DEVICE FOR ENDOSCOPIC SUBMUCOSAL DISSECTION: EX-VIVO STUDY IN A PORCINE MODEL
    Dobashi, Akira
    Song, Louis M. Wong Kee
    Gostout, Christopher
    Rajan, Elizabeth
    GASTROINTESTINAL ENDOSCOPY, 2018, 87 (06) : AB236 - AB237
  • [5] Safety testing of indocyanine green in an ex vivo porcine retina model
    Saikia, Parykshit
    Maisch, Tim
    Kobuch, Karin
    Jackson, Timothy L.
    Baeumler, Wolfgang
    Szeimies, Rolf-Markus
    Gabel, Veit-Peter
    Hillenkamp, Jost
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (11) : 4998 - 5003
  • [6] Efficacy and safety of an internal magnet traction device for endoscopic submucosal dissection: ex vivo study in a porcine model (with video)
    Dobashi, Akira
    Storm, Andrew C.
    Song, Louis M. Wong Kee
    Gostout, Christopher J.
    Deters, Jodie L.
    Miller, Charles A.
    Knipschield, Mary A.
    Rajan, Elizabeth
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2019, 33 (02): : 663 - 668
  • [7] Efficacy and safety of an internal magnet traction device for endoscopic submucosal dissection: ex vivo study in a porcine model (with video)
    Akira Dobashi
    Andrew C. Storm
    Louis M. Wong Kee Song
    Christopher J. Gostout
    Jodie L. Deters
    Charles A. Miller
    Mary A. Knipschield
    Elizabeth Rajan
    Surgical Endoscopy, 2019, 33 : 663 - 668
  • [8] Histological evaluation of monopolar and bipolar radiofrequency microneedling treatment in a porcine model
    Wang, Hao
    Hamblin, Michael R.
    Zhang, Yi
    Xu, Yidan
    Wen, Xiang
    LASERS IN SURGERY AND MEDICINE, 2024, 56 (03) : 288 - 297
  • [9] Ex vivo comparison between thyroid-dedicated bipolar and monopolar radiofrequency electrodes
    Yoon, Ra Gyoung
    Baek, Jung Hwan
    Chung, Sae Rom
    Choi, Young Jun
    Lee, Jeong Hyun
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2018, 34 (05) : 624 - 630
  • [10] Feasibility and safety of third-space endoscopic full-thickness resection in ex vivo and in vivo porcine models
    Goto, Osamu
    Sasaki, Motoki
    Akimoto, Teppei
    Tatsuguchi, Atsushi
    Kaise, Mitsuru
    Iwakiri, Katsuhiko
    Yahagi, Naohisa
    ENDOSCOPY INTERNATIONAL OPEN, 2019, 7 (04) : E471 - E476