Difference in Postoperative Periprosthetic Bone Mineral Density Changes Between 3 Major Designs of Uncemented Stems: A 3-Year Follow-Up Study

被引:39
作者
Inaba, Yutaka [1 ]
Kobayashi, Naomi [1 ]
Oba, Masatoshi [1 ]
Ike, Hiroyuki [1 ]
Kubota, So [1 ]
Saito, Tomoyuki [1 ]
机构
[1] Yokohama City Univ, Dept Orthopaed Surg, Yokohama, Kanagawa, Japan
关键词
total hip arthroplasty; bone mineral density; stem design; stress shielding; uncemented arthroplasty; TOTAL-HIP-ARTHROPLASTY; FEMORAL STEM; LONG-TERM; COMPONENTS; REPLACEMENTS; FIXATION; DEXA;
D O I
10.1016/j.arth.2016.02.009
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Although few studies have examined the direct effect of stress shielding on clinical outcomes, periprosthetic bone loss due to stress shielding is still an issue of concern, especially when physicians perform uncemented total hip arthroplasty (THA) in younger patients. Differences in femoral stem design may affect the degree of postoperative stress shielding. Therefore, the characteristics of the behavior for stress shielding of each type of femoral stem should be determined. This study compares differences in bone mineral density (BMD) change in the femur after primary THA between 3 major types of uncemented stems. Methods: Among a total of 89 hips, 26 hips received THA with a fit-and-fill type stem (VerSys Fiber Metal MidCoat; Zimmer, Inc, Warsaw, IN), 32 hips received a tapered rectangular Zweymuller-type stem (SL-Plus; Smith & Nephew Inc, Memphis, TN), and 31 received a tapered wedge-type stem (Accolade TMZF; Stryker Orthopaedics, Mahwah, NJ). BMD measurements were performed with a HOLOGIC Discovery device (Hologic Inc, Waltham, MA). Results: BMD in the medial-proximal femur was maintained for 3 years after THA in the group with the tapered wedge-type stem. BMD in the lateral-proximal femur was maintained for 3 years after THA in the group with the Zweymuller-type stem. There were no significant differences in the Harris Hip Score among the 3 stem groups preoperatively and 1, 2, and 3 years after surgery. Conclusion: There are clear differences in postoperative BMD loss of the proximal femur among these 3 commonly used uncemented stems. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:1836 / 1841
页数:6
相关论文
共 27 条
[1]   A concise minimum eight year follow-up of proximally porous-coated tapered titanium femoral stem in primary total hip arthroplasty [J].
Beecher, Benjamin ;
Glassner, Philip ;
Malchau, Henrik ;
Kwon, Young-Min .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (08) :1561-1565
[2]   Twenty-five-year survivorship of two thousand consecutive primary Charnley total hip replacements - Factors affecting survivorship of acetabular and femoral components [J].
Berry, DJ ;
Harmsen, WS ;
Cabanela, ME ;
Morrey, BF .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (02) :171-177
[3]   Continuous bone loss around a tapered uncemented femoral stem -: A long-term evaluation with DEXA [J].
Boden, Henrik S. G. ;
Skoldenberg, Olof G. ;
Salemyr, Mats F. ;
Lundberg, Hans-Jerker ;
Adolphson, Per Y. .
ACTA ORTHOPAEDICA, 2006, 77 (06) :877-885
[4]   Primary Total Hip Arthroplasty With an Uncemented Femoral Component Five- to Nine-Year Results [J].
Casper, David S. ;
Kim, Gregory K. ;
Restrepo, Camilo ;
Parvizi, Javad ;
Rothman, Richard H. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (06) :838-841
[5]   The accuracy and reproducibility of radiographic assessment of stress-shielding - A postmortem analysis [J].
Engh, CA ;
McAuley, JP ;
Sychterz, CJ ;
Sacco, ME ;
Engh, CA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (10) :1414-1420
[6]   POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS [J].
ENGH, CA ;
BOBYN, JD ;
GLASSMAN, AH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01) :45-55
[7]   Bone graft for revision hip arthroplasty [J].
Gamradt, SC ;
Lieberman, JR .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) :183-194
[8]   Ti based biomaterials, the ultimate choice for orthopaedic implants - A review [J].
Geetha, M. ;
Singh, A. K. ;
Asokamani, R. ;
Gogia, A. K. .
PROGRESS IN MATERIALS SCIENCE, 2009, 54 (03) :397-425
[9]   New femoral designs - Do they influence stress shielding? [J].
Glassman, A. H. ;
Bobyn, J. D. ;
Tanzer, M. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2006, (453) :64-74
[10]  
GRUEN TA, 1979, CLIN ORTHOP RELAT R, P17