Clinical implementation and commissioning of the MedAustron Particle Therapy Accelerator for non-isocentric scanned proton beam treatments

被引:28
作者
Grevillot, Loic [1 ]
Osorio Moreno, Jhonnatan [1 ]
Letellier, Virgile [1 ]
Dreindl, Ralf [1 ]
Elia, Alessio [1 ]
Fuchs, Hermann [1 ,2 ]
Carlino, Antonio [1 ]
Kragl, Gabriele [1 ]
Palmans, Hugo [1 ,3 ]
Vatnitsky, Stanislav [1 ]
Stock, Markus [1 ]
机构
[1] EBG MedAustron GmbH, Marie Curie Str 5, A-2700 Wiener Neustadt, Austria
[2] Med Univ Vienna, AKH Vienna, Dept Radiat Therapy, Vienna, Austria
[3] Natl Phys Lab, Hampton Rd, Teddington TW11 0LW, Middx, England
关键词
acceptance; beam delivery; commissioning; MedAustron; proton; quality assurance (QA); ITALIAN NATIONAL CENTER; LOW-DOSE ENVELOPE; QUALITY-ASSURANCE; DELIVERY-SYSTEM; PENCIL; CALIBRATION; DETECTORS; DESIGN;
D O I
10.1002/mp.13928
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose This paper describes the clinical implementation and medical commissioning of the MedAustron Particle Therapy Accelerator (MAPTA) for non-isocentric scanned proton beam treatments. Methods Medical physics involvement during technical commissioning work is presented. Acceptance testing procedures, including advanced measurement methods of intra-spill beam variations, are defined. Beam monitor calibration using two independent methods based on a dose-area product formalism is described. Emphasis is given to the medical commissioning work and the specificities related to non-isocentric irradiation, since a key feature of MedAustron is the routine delivery of non-isocentric scanned proton beam treatments. Results Key commissioning results and beam stability trend lines for more than 2 yr of clinical operation have been provided. Intra-spill beam range, size, and position variations were within specifications of 0.3 mm, 15%, and 0.5 mm, respectively. The agreement between two independent beam monitor calibration methods was better than 1.0%. Non-isocentric treatment delivery allowed lateral penumbra reduction of up to about 30%. Daily QA measurements of the beam range, size, position, and dose were always within 1 mm, 10%, 1 mm, and 2% from the baseline data, respectively. Conclusion Non-isocentric treatments have been successfully implemented at MedAustron for routine scanned proton beam therapy using horizontal and vertical fixed beamlines. Up to now every patient was treated in non-isocentric conditions. The presented methodology to implement a new Scanned Ion Beam Delivery (SIBD) system into clinical routine for proton therapy may serve as a guidance for other centers.
引用
收藏
页码:380 / 392
页数:13
相关论文
共 33 条
[1]   A beam model for focused proton pencil beams [J].
Almhagen, E. ;
Boersma, D. J. ;
Nystrom, H. ;
Ahnesjo, A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 52 :27-32
[2]  
Andreo P, 2006, IAEA TECH REPORTS SE, V398
[3]  
[Anonymous], 2007, SPINE J S, V7, p1S
[4]  
Bäumer C, 2015, J APPL CLIN MED PHYS, V16, P151, DOI 10.1120/jacmp.v16i6.5577
[5]   Commissioning of pencil beam and Monte Carlo dose engines for non-isocentric treatments in scanned proton beam therapy [J].
Carlino, A. ;
Boehlen, T. ;
Vatnitsky, S. ;
Grevillot, L. ;
Osorio, J. ;
Dreindl, R. ;
Palmans, H. ;
Stock, M. ;
Kragl, G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (17)
[6]   Characterization of PTW-31015 PinPoint ionization chambers in photon and proton beams [J].
Carlino, A. ;
Stock, M. ;
Zagler, N. ;
Marrale, M. ;
Osorio, J. ;
Vatnitsky, S. ;
Palmans, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (18)
[7]  
Carlino A, 2017, THESIS
[8]  
Elia A, 2019, THESIS
[9]  
Garonna A, 2016, IPAC 2016, P3176, DOI DOI 10.18429/JACOW-IPAC2016-THOAB01.
[10]   Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas MD Anderson Cancer Center, Proton Therapy Center, Houston [J].
Gillin, Michael T. ;
Sahoo, Narayan ;
Bues, Martin ;
Ciangaru, George ;
Sawakuchi, Gabriel ;
Poenisch, Falk ;
Arjomandy, Bijan ;
Martin, Craig ;
Titt, Uwe ;
Suzuki, Kazumichi ;
Smith, Alfred R. ;
Zhu, X. Ronald .
MEDICAL PHYSICS, 2010, 37 (01) :154-163