Development of an in-situ total phosphorus analyzer

被引:0
作者
Taylor, Elysia [1 ]
Bonner, James [1 ]
Nelson, Russell [1 ]
Fuller, Christopher [1 ]
Kirkey, William [1 ]
Cappelli, Stephanie [1 ]
机构
[1] Clarkson Univ, Dept Civil & Environm Engn, Potsdam, NY 13699 USA
来源
OCEANS 2015 - MTS/IEEE WASHINGTON | 2015年
关键词
phosphorus; in-situ; micro-fluidics; submersible; INJECTION;
D O I
暂无
中图分类号
U6 [水路运输]; P75 [海洋工程];
学科分类号
0814 ; 081505 ; 0824 ; 082401 ;
摘要
A low cost autonomous, in-situ analyzer for orthophosphate and total phosphorus was developed and tested for near real time monitoring in aquatic and marine environments. The system applies the colorimetric stannous chloride-molybdenum blue method following oxidative digestion with acidic peroxodisulfate. A (fluorinated ethylene propylene) FEB tubing reactor coil is wrapped around a germicidal UV (254nm) light to accelerate the digestion reaction. A Nickel chromium wire (NiCr) wound between reactor coil wraps provides controlled heating for digestion. Reactor temperature is monitored continuously with a small temperature probe and controlled with a PID. A sequential injection analysis (SIA) method was chosen where the analyzer operates as a sequential batch reactor. Water sample and reagents are delivered through the microfluidics using syringe pump driven by linear actuators that are controlled with a Teensy microprocessor. To equalize pressures, reagent storage and waste are contained in collapsible reagent bags maintained at hydrostatic pressure outside the submersible housing. Following digestion and colorimetric reactions the sample is transferred to flow cell (5 cm path length) where a photodiode measures sample absorbance at 600 nm. For orthophosphate phosphorus measurements, the sample is not treated with digestion reagents, heat, or UV. Potassium dihydrogen phosphate and adenosine monophosphate (AMP) were used to assess analyzer performance for measuring orthophosphate and total phosphorus, respectively. The prototype analyzer produced a linear response during bench testing to total phosphorus with a precision of +/- 4.3% over a detection range concentration range of 30-200 mu g P/L. To evaluate the performance with natural waters, dilutions of algal culture with known phosphorus concentration were measured and found to be within 90% of nominal phosphorus concentrations. A preliminary cost analysis indicates that the operating components of the system (minus submersible housing) can be purchased for under $3000 compared to $30,000 for commercially available systems. At this low cost of construction, this sensor design may be a viable option for use in nutrient monitoring programs. Considering that system components have been designed to withstand internal pressures equivalent to a water depth of 100 m with a designed duty capacity of 2 weeks suggest the possibility of applications in coastal waters and relatively deep lakes.
引用
收藏
页数:4
相关论文
共 9 条
  • [1] American Public Health Association (APHA), 1992, STAND METH EX WAT WA
  • [2] Flow analysis techniques for phosphorus:: an overview
    Estela, JM
    Cerdà, V
    [J]. TALANTA, 2005, 66 (02) : 307 - 331
  • [3] A compact portable flow analysis system for the rapid determination of total phosphorus in estuarine and marine waters
    Gentle, Brady S.
    Ellis, Peter S.
    Faber, Peter A.
    Grace, Michael R.
    McKelvie, Ian D.
    [J]. ANALYTICA CHIMICA ACTA, 2010, 674 (02) : 117 - 122
  • [4] Hach, PHOSPH WAT WAST SEAW
  • [5] Rate of phosphoantimonylmolybdenum blue complex formation in acidic persulfate digested sample matrix for total dissolved phosphorus determination: Importance of post-digestion pH adjustment
    Huang, Xiao-Lan
    Zhang, Jia-Zhong
    [J]. TALANTA, 2008, 77 (01) : 340 - 345
  • [6] A compact flow injection analysis system for surface mapping of phosphate in marine waters
    Lyddy-Meaney, AJ
    Ellis, PS
    Worsfold, PJ
    Butler, ECV
    McKelvie, ID
    [J]. TALANTA, 2002, 58 (06) : 1043 - 1053
  • [7] Pierzynski G., 2000, METHODS PHOSHORUS AN
  • [8] Gulf of Mexico hypoxia, aka "The dead zone"
    Rabalais, NN
    Turner, RE
    Wiseman, WJ
    [J]. ANNUAL REVIEW OF ECOLOGY AND SYSTEMATICS, 2002, 33 : 235 - 263
  • [9] SEQUENTIAL INJECTION - A NEW CONCEPT FOR CHEMICAL SENSORS, PROCESS ANALYSIS AND LABORATORY ASSAYS
    RUZICKA, J