The cardiocoil stent-artery interaction

被引:8
作者
Brand, M
Ryvkin, M [1 ]
机构
[1] Tel Aviv Univ, Fac Engn, Dept Solid Mech Mat & Syst, IL-69978 Tel Aviv, Israel
[2] Tel Aviv Univ, Fac Engn, Dept Biomed Engn, IL-69978 Tel Aviv, Israel
来源
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME | 2005年 / 127卷 / 02期
关键词
self-expanding stent; damage factor; analytic solution; parametric study;
D O I
10.1115/1.1871194
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
An analytical approach for the mechanical interaction of the self-expanding Cardiocoil stent with the stenosed artery is presented. The damage factor as the contact stress at the stent-artery interface is determined. The stent is considered as an elastic helical rod having a nonlinear pressure-displacement dependence, while the artery is modeled by an elastic cylindrical shell. An influence of a moderate relative thickness of the shell is estimated. The equations for both the stent and the artery, are presented in the stent-associated helical coordinates. The computational efficiency of the model enabled to carry out a parametric study of the damage factor Comparative examinations are conducted for the stents made of the helical rods with circular and rectangular cross sections. It was found, in particular that, under same other conditions, the damage factor for the stent with a circular cross section may be two times larger than that for a rectangular one.
引用
收藏
页码:337 / 344
页数:8
相关论文
共 25 条
  • [1] Akiyama T., 1997, Journal of the American College of Cardiology, V29, p368A
  • [2] [Anonymous], 1993, BEHAV SHELLS COMPOSE
  • [3] Auricchio F., 2001, COMPUT METHOD BIOMEC, V4, P249, DOI DOI 10.1080/10255840108908007
  • [4] SELF-EXPANDABLE NITINOL STENT FOR CARDIOVASCULAR APPLICATIONS - CANINE AND HUMAN-EXPERIENCE
    BEYAR, R
    SHOFTI, R
    GRENEDIER, E
    HENRY, M
    GLOBERMAN, O
    BEYAR, M
    [J]. CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1994, 32 (02): : 162 - 170
  • [5] De Belder A., 1998, STENT, V1, P74
  • [6] Mechanical behaviour modelling of balloon-expandable stents
    Dumoulin, C
    Cochelin, B
    [J]. JOURNAL OF BIOMECHANICS, 2000, 33 (11) : 1461 - 1470
  • [7] DUNITZ M, 1996, HDB CORONARY STENTS
  • [8] Mechanical properties of metallic stents: How do these properties influence the choice of stent for specific lesions?
    Dyet, JF
    Watts, WG
    Ettles, DF
    Nicholson, AA
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2000, 23 (01) : 47 - 54
  • [9] Mechanical properties of coronary stents determined by using finite element analysis
    Etave, F
    Finet, G
    Boivin, M
    Boyer, JC
    Rioufol, G
    Thollet, G
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (08) : 1065 - 1075
  • [10] STRENGTH, ELASTICITY, AND PLASTICITY OF EXPANDABLE METAL STENTS - IN-VITRO STUDIES WITH 3 TYPES OF STRESS
    FLUECKIGER, F
    STERNTHAL, H
    KLEIN, GE
    ASCHAUER, M
    SZOLAR, D
    KLEINHAPPL, G
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 1994, 5 (05) : 745 - 750