Trial of new vascular clips for aortic anastomosis in a canine model

被引:3
作者
Masuda, Shinya
Saiki, Yoshikatsu
Kawatsu, Satoshi
Yoshioka, Ichiro
Fujiwara, Hidenori
Kawamoto, Shunsuke
Sai, Sadahiro
Iguchi, Atsushi
Sakamoto, Naoya
Ohashi, Toshiro
Sato, Masaaki
Tabayashi, Koichi
机构
[1] Tohoku Univ, Grad Sch Med, Dept Cardiovasc Surg, Aoba Ku, Sendai, Miyagi 9808574, Japan
[2] Tohoku Univ, Sch Mech Engn, Sendai, Miyagi 9808574, Japan
关键词
D O I
10.1016/j.jtcvs.2007.04.048
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: We created a new vascular clip designed for aortic surgery. The purposes of this investigation were to examine surgical applicability in a clinically relevant aortic replacement model and to assess biomechanical strength of the clipped anastomosis and serial histologic changes in the clipped anastomotic site. Methods: Twenty-one beagles underwent descending thoracic aortic replacement. Distal anastomosis was performed with the new clips, mimicking the cuffed anastomosis technique, and proximal anastomosis was carried out by conventional suture anastomosis. Tissue specimens of the anastomotic sites were harvested at 1, 3, 6, and 12 months postoperatively for examination. Results: There was no significant difference in the time required to carry out clip anastomosis ( 12.2 +/- 1.3 minutes) and suture anastomosis ( 13.7 +/- 0.9 minutes; P = .38). Neither type of anastomotic site was disrupted by raising the intraluminal pressure to 280 mm Hg. Microscopically, the areas of aortic wall compressed by vascular clips appeared as hyalinized areas adjacent to surrounding collagen fibers, with no significant infiltration of inflammatory cells. Identical histologic changes were observed at the site of the sutured anastomosis. The media at the clipped anastomosis site was significantly thinner than that at the sutured anastomosis site at 1 month after the operation. However, there was no significant difference in the thickness of the media at 3 months. Conclusions: The new vascular clips were effective in this clinically relevant model, with appropriate biomechanical strength, and the anastomotic sites underwent similar histologic changes to those observed after suture anastomosis.
引用
收藏
页码:723 / 730
页数:8
相关论文
共 30 条
[1]  
Ashizawa K, 1991, Nihon Geka Gakkai Zasshi, V92, P874
[2]  
BEVER MB, 1986, IRON BASED BIOMECHAN, V3, P2430
[3]   Stephen Jay Gould (1941-2002): A wonderful life [J].
Carroll, SB .
DEVELOPMENTAL CELL, 2002, 3 (01) :21-23
[4]  
Elkouri Stephane, 2004, Vasc Endovascular Surg, V38, P321, DOI 10.1177/153857440403800403
[5]  
GRIEPP RB, 1975, J THORAC CARDIOV SUR, V70, P1051
[6]   Sutureless coronary artery bypass with biologic glued anastomoses: Preliminary in vivo and in vitro results [J].
Gundry, SR ;
Black, K ;
Izutani, H .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 120 (03) :473-477
[7]   A prospective randomized trial of hand-sutured versus mechanically stapled anastomoses for gastroduodenostomy after distal gastrectomy [J].
Hori S. ;
Ochiai T. ;
Gunji Y. ;
Hayashi H. ;
Suzuki T. .
Gastric Cancer, 2004, 7 (1) :24-30
[8]  
HPHF RP, 1962, ANN SURG, V156, P805
[9]   NONPENETRATING CLIPS SUCCESSFULLY REPLACING SUTURES IN BASE OF SKULL SURGERY [J].
KIRSCH, WM ;
ZHU, YH ;
HARDESTY, RA ;
PETTI, G ;
FURNAS, D .
SKULL BASE SURGERY, 1993, 3 (04) :171-181
[10]  
KUMADA J, 1972, J ARTIFICIAL ORGAN, V1, P128