The influence of cornea properties on rebound tonometry

被引:117
作者
Chui, Wan-sang [1 ]
Lam, Andrew [1 ]
Chen, Davie [1 ]
Chiu, Roger [1 ]
机构
[1] Hong Kong Polytech Univ, Sch Optometry, Hong Kong, Hong Kong, Peoples R China
关键词
D O I
10.1016/j.ophtha.2007.03.061
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To determine the influence of corneal thickness, corneal hysteresis (CH), and corneal resistance factor (CRF) on intraocular pressure (IOP) measurements obtained using the iCARE rebound tonometer. Design: Cross-sectional study. Participants: One hundred twenty-five eyes of 125 normal healthy subjects. Methods: Intraocular pressure was determined in all subjects using the iCARE at both central and temporal corneal regions. Corneal thickness at these 2 regions was determined by ultrasonic pachymetry. Corneal hysteresis and corneal resistance factor were measured using an ocular response analyzer. Goldmann applanation tonometry (GAT) was also performed on all eyes. Main Outcome Measures: Intraocular pressure, corneal thickness, corneal hysteresis, and corneal resistance factor. Results: Although the peripheral cornea was significantly thicker than the central cornea, IOPs obtained by the iCARE at the 2 regions were similar and significantly higher than GAT measurements. There were an overestimation at higher IOPs and an underestimation at lower IOPs relative to GAT measurements. Multiple regression analysis showed a significant correlation between the iCARE and CH (partial correlation r = -0.67, P<0.01) and CRF (partial correlation r = 0.82, P<0.01). The correlation between the iCARE and central corneal thickness was not significant. Conclusions: Rebound tonometry is affected by corneal properties including CH and CRF but not corneal thickness. Ophthalmology 2008; 115:80-84 (c) 2008 by the American Academy of Ophthalmology.
引用
收藏
页码:80 / 84
页数:5
相关论文
共 32 条
[1]  
ABRAHAM LM, IN PRESS EYE
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Collagen fibrils appear more closely packed in the prepupillary cornea: Optical and biomechanical implications [J].
Boote, C ;
Dennis, S ;
Newton, RH ;
Puri, H ;
Meek, KM .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2003, 44 (07) :2941-2948
[4]   Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients [J].
Brusini, P ;
Salvetat, ML ;
Zeppieri, M ;
Tosoni, C ;
Parisi, L .
JOURNAL OF GLAUCOMA, 2006, 15 (03) :213-217
[5]   Clinical evaluation of rebound tonometer [J].
Davies, LN ;
Bartlett, H ;
Mallen, EAH ;
Wolffsohn, JS .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2006, 84 (02) :206-209
[6]   Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach [J].
Doughty, MJ ;
Zaman, ML .
SURVEY OF OPHTHALMOLOGY, 2000, 44 (05) :367-408
[7]   Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings [J].
Feltgen, N ;
Leifert, D ;
Funk, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2001, 85 (01) :85-87
[8]   Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population [J].
Fernandes, P ;
Díaz-Rey, JA ;
Queirós, A ;
Gonzalez-Meijome, JM ;
Jorge, J .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (05) :436-440
[9]  
Friedman DS, 2004, ARCH OPHTHALMOL-CHIC, V122, P532
[10]   Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes [J].
García-Resúa, C ;
González-Meijome, JM ;
Gilino, J ;
Yebra-Pimentel, E .
OPTOMETRY AND VISION SCIENCE, 2006, 83 (02) :102-107