Surfactant coated graphitic carbon based stationary phases for anion-exchange chromatography

被引:15
作者
Chambers, Stuart D. [1 ]
Lucy, Charles A. [1 ]
机构
[1] Univ Alberta, Dept Chem, Gunning Lemieux Chem Ctr, Edmonton, AB T6G 2G2, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
graphitic carbon; ion chromatography; surfactant coating; carbonate/bicarbonate eluent; inorganic ions;
D O I
10.1016/j.chroma.2007.10.108
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Separations of common inorganic anions were carried out on three different surfactant coated media using carbonate[bicarbonate eluents with suppressed conductivity detection. Graphitic carbon columns (porous graphitic carbon and carbon-clad zirconia) packed with 3 mu m particles have been converted into anion-exchange stationary phases by equilibration with the cationic surfactants: didodecyldimethylammonium bromide (DDAB) cetyltrimethylammonium bromide (CTAB); and cetylpyridinium chloride (CPC). Additionally, an ethylene-bridged silica column was studied with CPC coatings. Porous graphitic carbon (PGC) columns coated with DDAB exhibited pressure increases and loss of resolution at higher capacities. CPC coatings on PGC exhibited better repeatability and efficiencies of 5.0 x 10(4) plates/m. However, CPC coatings exhibited a 15% loss in retention factor with < 1.2 x 10(3) column volumes on PGC. Conversely, the ethylene-bridged silica column showed complete failure in less than 8 h of use. As with PGC, carbon-clad zirconia coated with CPC showed an initial loss of capacity, but thereafter was stable for more than 1.7 x 10(3) column volumes (t(r) RSD < 2%). (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:178 / 184
页数:7
相关论文
共 73 条
[1]   CHROMATE AS A MOBILE-PHASE FOR THE DETERMINATION OF ANIONS USING CETYLTRIMETHYLAMMONIUM-COATED COLUMNS AND INDIRECT PHOTOMETRIC DETECTION [J].
ALBAZI, SJ ;
GHARIB, T ;
CORTES, JC .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1995, 18 (03) :489-503
[2]  
CARR PW, 2007, COMMUNICATION
[3]   DYNAMICALLY COATED COLUMNS FOR THE SEPARATION OF METAL-IONS AND ANIONS BY ION CHROMATOGRAPHY [J].
CASSIDY, RM ;
ELCHUK, S .
ANALYTICAL CHEMISTRY, 1982, 54 (09) :1558-1563
[4]   Effect of buffers on silica-based column stability in reversed-phase high-performance liquid chromatography [J].
Claessens, HA ;
vanStraten, MA ;
Kirkland, JJ .
JOURNAL OF CHROMATOGRAPHY A, 1996, 728 (1-2) :259-270
[5]   MODIFIED CARBON-BLACK IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
COLIN, H ;
EON, C ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHY, 1976, 119 (APR28) :41-54
[6]   DEVELOPMENT AND USE OF CARBON ADSORBENTS IN HIGH-PERFORMANCE LIQUID-SOLID CHROMATOGRAPHY .2. REPRODUCIBILITY OF CARBON ADSORBENTS AND INFLUENCE OF GRAPHITIZATION ON THEIR PERFORMANCE [J].
COLIN, H ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHY, 1977, 137 (01) :19-33
[7]   REVERSED-PHASE LIQUID-SOLID CHROMATOGRAPHY ON MODIFIED CARBON-BLACK [J].
COLIN, H ;
EON, C ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHY, 1976, 122 (JUL7) :223-242
[8]  
Conder J.R.C.L. Young., 1979, PHYSICOCHEMICAL MEAS
[9]   Rapid, low pressure, and simultaneous ion chromatography of common inorganic anions and cations on short permanently coated monolithic columns [J].
Connolly, D ;
Victory, D ;
Paull, B .
JOURNAL OF SEPARATION SCIENCE, 2004, 27 (10-11) :912-920
[10]   Fast ion chromatography of common inorganic anions on a short ODS column permanently coated with didodecyldimethylammonium bromide [J].