Evaluation of a New Disinfection Method for Ultrasound Probes Used for Regional Anesthesia Ultraviolet C Light

被引:20
作者
Bloc, Sebastien [1 ]
Mercadal, Luc [1 ]
Garnier, Thierry [1 ]
Komly, Bernard [1 ]
Leclerc, Pascal [1 ]
Morel, Bertrand [1 ]
Ecoffey, Claude [3 ]
Dhonneur, Gilles [2 ]
机构
[1] Claude Gallen Private Hosp, F-91480 Quincy Sous Senart, France
[2] Jean Verdier Univ Hosp Paris, Assistance Publ Hop Paris, Bondy, France
[3] Univ Hosp Rennes, Rennes, France
关键词
nerve; probe disinfection; regional anesthesia; ultrasound; ultraviolet C light; INFECTION; REDUCE;
D O I
10.7863/jum.2011.30.6.785
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Objectives-Ultrasound-guided regional anesthesia is commonly used for block placement. At present, the risk of cross contamination from probes is not well documented. To avoid transmission of infectious agents, several methods have been used for probe disinfection and protection. The aim of this study was to evaluate the antibacterial efficacy of a new high-level disinfection method based on ultraviolet C (UV-C) light under routine conditions after block placement with an unprotected probe. Methods-The study was after approval by the local Ethics Committee. In the first part of the study, 15 ultrasound probes were exposed to a large inoculum of 3 bacteria. Ultraviolet C disinfection consisted of cleaning the probe with dry and disinfectant-impregnated paper followed by a 90-second UV-C disinfection cycle in a decontamination chamber. A protocol was established to retrieve the probe with sterile gloves after opening the door of the chamber. In the second part, SO blocks were placed with ultrasound-guided regional anesthesia. The skin was first prepared with an antiseptic solution, and sterile gel was applied; no covers were used to protect the probes. The blocks were then disinfected with UV-C light. Bacteriologic samples were collected before and after the UV-C method and inoculated on chocolate agar plates. Results-During the first part of the study, all probes were infected after inoculation (> 150 colony-forming units) but were considered sterile (< 10 colony-forming units) after disinfection. During the second part of the study, all probes were considered sterile before and after disinfection. Conclusions-Ultraviolet C disinfection seems relevant for ultrasound-guided regional anesthesia just before block placement. It offers simple, fast, and effective high-level disinfection. Moreover, this method should obviate the use of sterile probe covers, which can improve echogenicity.
引用
收藏
页码:785 / 788
页数:4
相关论文
共 7 条
[1]   Ultrasound guided regional anaesthesia: An effective method for cleaning the probes [J].
Bloc, S. ;
Garnier, T. ;
Bounhiol, C. ;
Komly, B. ;
Machado, G. ;
Leclerc, P. ;
Roy, M. ;
Mercadal, L. ;
Demontoux, V. ;
Morel, B. ;
Ecoffey, C. ;
Dhonneur, G. .
ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 2008, 27 (12) :994-998
[2]   US probes: Risk of cross infection and ways to reduce it - Comparison of cleaning methods [J].
Fowler, C ;
McCracken, D .
RADIOLOGY, 1999, 213 (01) :299-300
[3]   Evaluation of a new disinfection procedure for ultrasound probes using ultraviolet light [J].
Kac, G. ;
Gueneret, M. ;
Rodi, A. ;
Abergel, E. ;
Grataloup, C. ;
Denarie, N. ;
Peyrard, S. ;
Chatellier, G. ;
Emmerich, J. ;
Meyer, G. ;
Podglajen, I. .
JOURNAL OF HOSPITAL INFECTION, 2007, 65 (02) :163-168
[4]  
LYTLE F, 2009, CRIT CARE S1, V13, pP290
[5]   How clean is your probe? Microbiological assessment of ultrasound transducers in routine clinical use, and cost-effective ways to reduce contamination [J].
Mullaney, P. J. ;
Munthali, P. ;
Vlachou, P. ;
Jenkins, D. ;
Rathod, A. ;
Entwisle, J. .
CLINICAL RADIOLOGY, 2007, 62 (07) :694-698
[6]   CAN ULTRASOUND PROBES AND COUPLING GEL BE A SOURCE OF NOSOCOMIAL INFECTION IN PATIENTS UNDERGOING SONOGRAPHY - AN IN-VIVO AND IN-VITRO STUDY [J].
MURADALI, D ;
GOLD, WL ;
PHILLIPS, A ;
WILSON, S .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 164 (06) :1521-1524
[7]  
Spaulding E.H., 1968, Disinfection, Sterilization, and Preservation, V1st, P517