Studying the biomechanical effects of zirconia implant geometry and bone quality on the bone-implant interface using finite element analysis

被引:0
|
作者
Velmurugan, D. [1 ]
Pridhar, T. [2 ]
Yuyaraj, T. [1 ]
Srinivasan, R. [2 ]
Babu, B. Suresh [2 ]
机构
[1] Muthayammal Engn Coll Autonomous, Dept Mech Engn, Rasipuram 637408, Tamil Nadu, India
[2] Sri Krishna Coll Technol, Dept Mech Engn, Coimbatore 641042, Tamil Nadu, India
来源
JOURNAL OF OPTOELECTRONIC AND BIOMEDICAL MATERIALS | 2021年 / 13卷 / 02期
关键词
Bone quality; Dental implants; Stress analysis; Zirconia; OSSEOINTEGRATED IMPLANTS; ORAL IMPLANTS; OVERDENTURES; PROSTHESES; FIXTURES; 15-YEAR; SUCCESS; DESIGN; LOAD;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
The success of dental implants is influenced by several biomechanical factors such as implant design, bone quality, length of the implant, and the load transferred to the implant. The present study aims to determine the effects of bone quality, implant length, and bone type on stress/strain distribution in bone and implant for a zirconia implant. The study was performed using the three dimensional finite element analysis and four different lengths, two types of implants, and four different bone qualities. The Elastic modulus of a cancellous bone was varied to represent the four different bone qualities. A load of 100 N was applied at the center of the abutment. The result of this current work shows that a bone is subjected to the maximum equivalent stress and strain when the cancellous bone density decreased. The screw type implant induced lesser strain than the cylinder type load. The study also confirms that longer implants produced lesser stain than shorter implants, and type I, and II bones induced lesser stresses than type III, and IV bones. The importance of bone quality has been confirmed from this study, and zirconia dental implants have induced lesser stresses.
引用
收藏
页码:67 / 75
页数:9
相关论文
共 50 条
  • [41] Finite element analysis of the biomechanical effects of PEEK dental implants on the peri-implant bone
    Schwitalla, A. D.
    Abou-Emara, M.
    Spintig, T.
    Lackmann, J.
    Mueller, W. D.
    JOURNAL OF BIOMECHANICS, 2015, 48 (01) : 1 - 7
  • [42] Influence of bone quality on the mechanical interaction between implant and bone: A finite element analysis
    Azcarate-Velazquez, Francisco
    Castillo-Oyague, Raquel
    Oliveros-Lopez, Luis-Guillermo
    Torres-Lagares, Daniel
    Martinez-Gonzalez, Alvaro-Jose
    Perez-Velasco, Andrea
    Lynch, Christopher D.
    Gutierrez-Perez, Jose-Luis
    Serrera-Figallo, Maria-Angeles
    JOURNAL OF DENTISTRY, 2019, 88
  • [43] The effects of PTH, loading and surgical insult on cancellous bone at the bone-implant interface in the rabbit
    Fahlgren, Anna
    Yang, Xu
    Ciani, Cesare
    Ryan, James A.
    Kelly, Natalie
    Ko, Frank C.
    van der Meulen, Marjolein C. H.
    Bostrom, Mathias P. G.
    BONE, 2013, 52 (02) : 718 - 724
  • [44] Nonlinear Finite Element Analysis of Bone-Implant Contact in Three Short Dental Implant Models with Varying Osseointegration Percentages
    Alemayehu, Dawit Bogale
    Todoh, Masahiro
    Huang, Song-Jeng
    ORAL, 2024, 4 (04): : 505 - 524
  • [45] Evaluation of the glenoid implant survival using a biomechanical finite element analysis: Influence of the implant design, bone properties, and loading location
    Mansat, P.
    Briot, J.
    Mansat, M.
    Swider, P.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (03) : 79S - 83S
  • [46] Evaluate the effect of bone density variation on stress distribution at the bone-implant interface using numerical analysis
    Hindurao, Bhargav
    Gujare, Aditya
    Jadhav, Harshavardhan
    Dhatrak, Pankaj
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2024, 238 (05) : 463 - 470
  • [47] Bone-Implant Interface with Simulated Insertion Stress Around an Immediately Loaded Dental Implant in the Anterior Maxilla: A Three-Dimensional Finite Element Analysis
    Lee, Joon-Seok
    Cho, In-Ho
    Kim, Yung-Soo
    Heo, Seong-Joo
    Kwon, Ho-Beom
    Lim, Young-Jun
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2012, 27 (02) : 295 - 302
  • [48] The effects of particulate polyethylene at a weight-bearing bone-implant interface
    Karahan, M
    Esemenli, T
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02): : 340 - 340
  • [49] In vivo monitoring of bone-implant bond strength by microCT and finite element modelling
    Stadelmann, Vincent A.
    Conway, Carl M.
    Boyd, Steven K.
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2013, 16 (09) : 993 - 1001
  • [50] ELECTRON-MICROSCOPIC STUDY OF THE BONE-IMPLANT INTERFACE
    ROLAND, K
    DUFIEF, MP
    SHERIDAN, R
    HEUSON, JA
    MICRON AND MICROSCOPICA ACTA, 1987, 18 (03): : 209 - 209