Implant Material, Type of Fixation at the Shaft, and Position of Plate Modify Biomechanics of Distal Femur Plate Osteosynthesis

被引:29
作者
Kandemir, Utku [1 ]
Augat, Peter [2 ,3 ]
Konowalczyk, Stefanie [4 ]
Wipf, Felix [4 ]
von Oldenburg, Geert [5 ]
Schmidt, Ulf [6 ]
机构
[1] Univ Calif San Francisco, Dept Orthoped Surg, San Francisco, CA 94143 USA
[2] Paracelsus Med Univ, Inst Biomech, Salzburg, Austria
[3] Berufsgenossenschaftl Unfallklin, Inst Biomech, Murnau, Germany
[4] Stryker, Selzach, Switzerland
[5] Stryker, Kiel, Germany
[6] Krankenhaus Barmherzigen Schwestern Ried, Dept Trauma Orthopaed Surg, Schlossberg 1, A-4910 Ried Im Innkreis, Austria
关键词
implant; material; distal femur; biomechanics; osteosynthesis; ANGLED BLADE PLATE; LCP CONDYLAR PLATE; FEMORAL FRACTURES; OSTEOPOROTIC BONE; INTERNAL-FIXATION; LOCKING PLATES; INTRAMEDULLARY NAIL; EPIDEMIOLOGY; STABILITY; FORCES;
D O I
10.1097/BOT.0000000000000860
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To investigate whether (1) the type of fixation at the shaft (hybrid vs. locking), (2) the position of the plate (offset vs. contact) and (3) the implant material has a significant effect on (a) construct stiffness and (b) fatigue life in a distal femur extraarticular comminuted fracture model using the same design of distal femur periarticular locking plate. Methods: An extraarticular severely comminuted distal femoral fracture pattern (OTA/AO 33-A3) was simulated using artificial bone substitutes. Ten-hole distal lateral femur locking plates were used for fixation per the recommended surgical technique. At the distal metaphyseal fragment, all possible locking screws were placed. For the proximal diaphyseal fragment, different types of screws were used to create 4 different fixation constructs: (1) stainless steel hybrid (SSH), (2) stainless steel locked (SSL), (3) titanium locked (TiL), and (4) stainless steel locked with 5-mm offset at the diaphysis (SSLO). Six specimens of each construct configuration were tested. First, each specimen was nondestructively loaded axially to determine the stiffness. Then, each specimen was cyclically loaded with increasing load levels until failure. Results: Construct Stiffness: The fixation construct with a stainless steel plate and hybrid fixation (SSH) had the highest stiffness followed by the construct with a stainless steel plate and locking screws (SSL) and were not statistically different from each other. Offset placement (SSLO) and using a titanium implant (TiL) significantly reduced construct stiffness. Fatigue Failure: The stainless steel with hybrid fixation group (SSH) withstood the most number of cycles to failure and higher loads, followed by the stainless steel plate and locking screw group (SSL), stainless steel plate with locking screws and offset group (SSLO), and the titanium plate and locking screws group (TiL) consecutively. Offset placement (SSLO) as well as using a titanium implant (TiL) reduced cycles to failure. Conclusions: Using the same plate design, the study showed that implant material, screw type, and position of the plate affect the construct stiffness and fatigue life of the fixation construct. With this knowledge, the surgeon can decide the optimal construct based on a given fracture pattern, bone strength, and reduction quality.
引用
收藏
页码:E241 / E246
页数:6
相关论文
共 44 条
[1]  
ARNESON TJ, 1988, CLIN ORTHOP RELAT R, P188
[2]   Shear movement at the fracture site delays healing in a diaphyseal fracture model [J].
Augat, P ;
Burger, J ;
Schorlemmer, S ;
Henke, T ;
Peraus, M ;
Claes, L .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2003, 21 (06) :1011-1017
[3]   Mechanics and mechano-biology of fracture healing in normal and osteoporotic bone [J].
Augat, Peter ;
Simon, Ulrich ;
Liedert, Astrid ;
Claes, Lutz .
OSTEOPOROSIS INTERNATIONAL, 2005, 16 (Suppl 2) :S36-S43
[4]   Management of Distal Femur Fractures With Modern Plates and Nails: State of the Art [J].
Beltran, Michael J. ;
Gary, Joshua L. ;
Collinge, Cory A. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (04) :165-172
[5]   Comparison of the LISS and a retrograde-inserted supracondylar intramedullary nail for fixation of a periprosthetic distal femur fracture proximal to a total knee arthroplasty [J].
Bong, MR ;
Egol, KA ;
Koval, KJ ;
Kummer, FJ ;
Su, ET ;
Iesaka, K ;
Bayer, J ;
Di Cesare, PE .
JOURNAL OF ARTHROPLASTY, 2002, 17 (07) :876-881
[6]   Far Cortical Locking Can Improve Healing of Fractures Stabilized with Locking Plates [J].
Bottlang, Michael ;
Lesser, Maren ;
Koerber, Julia ;
Doornink, Josef ;
von Rechenberg, Brigitte ;
Augat, Peter ;
Fitzpatrick, Daniel C. ;
Madey, Steven M. ;
Marsh, J. Lawrence .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (07) :1652-1660
[7]   Epidemiology of adult fractures: A review [J].
Court-Brown, Charles M. ;
Caesar, Ben .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (08) :691-697
[8]  
Cusick R P, 2000, Am J Orthop (Belle Mead NJ), V29, P695
[9]   Internal forces and moments in the femur during walking [J].
Duda, GN ;
Schneider, E ;
Chao, EYS .
JOURNAL OF BIOMECHANICS, 1997, 30 (09) :933-941
[10]   MECHANICS OF RETROGRADE NAIL VERSUS PLATE FIXATION FOR SUPRACONDYLAR FEMUR FRACTURES [J].
FIROOZBAKHSH, K ;
BEHZADI, K ;
DECOSTER, TA ;
MONEIM, MS ;
NARAGHI, FF .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1995, 9 (02) :152-157