Evaluation of high resolution X-ray monochromatic wavelength dispersive X-ray fluorescence instrument for determining Pu and U in nuclear reprocessing streams

被引:7
作者
McIntosh, K. G. [1 ]
Havrilla, G. J. [1 ]
Gilmore, R. F., Jr. [2 ]
Holland, M. K. [2 ]
机构
[1] Los Alamos Natl Lab, Los Alamos, NM 87545 USA
[2] Savannah River Natl Lab, Aiken, SC 29808 USA
关键词
MWDXRF; hiRX; Elemental analysis; Reprocessing; Accountancy; FUEL; PLUTONIUM; URANIUM;
D O I
10.1016/j.sab.2018.04.019
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
An alpha-prototype benchtop instrument was developed and tested for measuring Pu and U in spent fuel dis-solver tanks in a nuclear fuel reprocessing plant. The high resolution X-ray (hiRX) instrument was designed with the aim of providing a new capability to measure samples with high radiation background with selectivity, accuracy, and precision using ultra-low volumes (7 mu L.) of liquid. The instrument was tested with specimens from the operational reprocessing facility at Savannah River National Laboratory. The results showed that bias was <6% for Pu nitrate process solution and < 18% for Pu in U/HNO3 matrix. Instrumental precision was <1% RSD while between microcell repeatability was <6% RSD. The limit of detection for Pu in HNO3 matrix was 3 x 10(-5) g/L (100 s, 35 W). This study demonstrates that hiRX technology has the potential to provide rapid and simultaneous quantification of Pu and U without sample preparation, making it suitable for process monitoring and materials control and accountability applications. (C) 2018 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 17 条
[1]   ANALYTICAL-CHEMISTRY IN NUCLEAR TECHNOLOGY [J].
ACHE, HJ .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 343 (11) :852-862
[2]  
[Anonymous], INT SAF CHALL OPP 21
[3]  
[Anonymous], 2010, STR-368 International Target Values 2010 for measurement uncertainties in safeguarding nuclear material
[4]  
[Anonymous], 24609 JRC EUR EN
[5]  
[Anonymous], ANLNE157
[6]  
[Anonymous], SAND201417297C SNL
[7]  
[Anonymous], 2008, EV MEAS DAT GUID EXP, DOI DOI 10.1373/CLINCHEM.2003.030528
[8]  
[Anonymous], SRNLL4600201600017
[9]  
[Anonymous], SRNLL41102016000001
[10]   Low-level sulfur in fuel determination using monochromatic WD XRF-ASTM D 7039-04 [J].
Chen, ZW ;
Wei, FZ ;
Radley, I ;
Beumer, B .
Elemental Analysis of Fuels and Lubricants: Recent Advances and Future Prospects, 2005, 1468 :116-127