In Vivo Trial of a Novel Atraumatic Urinary Catheter Design for Prevention of Catheter-Induced Trauma

被引:6
作者
Azar, Ramyar [1 ]
Shadpour, Pejman [2 ]
机构
[1] IUMS, Tehran, Iran
[2] IUMS, HMRC, HKC, Vanak Sq Valiasr Ave, Tehran 19697, Iran
关键词
FOLEY CATHETER; COMPLICATIONS; INFECTION; RISK;
D O I
10.1089/end.2015.0827
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Purpose: Clinically significant urinary tract infection may be the most familiar complication of urinary catheterization; however, catheter-associated trauma can entail even more significant morbidity. We have designed and patented a novel atraumatic urinary catheter (AUC) and through this study, we compare its efficacy with the conventional Foley catheter (FC) in vivo. Materials and Methods: Forty male rabbits were divided into two equal groups for FC and AUC. The animals were catheterized on day 4 for 8 hours to compare the maintenance of the tubes and sedated before standardized forcible extraction of the catheter under sedation with the balloon still inflated. They were then examined visually and by cystoscopy. Results: In a number of animals in either group, the catheters failed to drain urine effectively due to anatomical issues. Those were excluded from the study. At the extraction phase, evidence of urethral trauma was found in 13 of 15 FC rabbits, but only in 4 of 14 in the AUC group (p = 0.009). Major trauma, however, was exclusively seen in the FC group, with 12 of 15 subjects sustaining deep lacerations or urethral disruption. Conclusion: The two catheters bear similar efficacy for maintaining urine drainage; however, traumatic injury is significantly less common following forceful extraction of the AUC than a conventional Foley.
引用
收藏
页码:822 / 827
页数:6
相关论文
共 20 条
[1]  
Azar R., 2013, US Patent, Patent No. [US 8,939,962, 8939962]
[2]   Vena Cava Air Embolism After Traumatic Foley Catheter Placement [J].
Chavez, Anthony H. ;
Reilly, T. Philip ;
Bird, Erin T. .
UROLOGY, 2009, 73 (04) :748-749
[3]   Preventing Urethral Trauma from Inadvertent Inflation of Catheter Balloon in the Urethra during Catheterization: Evaluation of a Novel Safety Syringe after Correlating Trauma with Urethral Distension and Catheter Balloon Pressure [J].
Davis, Niall F. ;
Mooney, Rory O'C. ;
Cunnane, Conor V. ;
Cunnane, Eoghan M. ;
Thornhill, John A. ;
Walsh, Michael T. .
JOURNAL OF UROLOGY, 2015, 194 (04) :1138-1145
[4]  
Doherty W, 2000, Br J Nurs, V9, P2006
[5]  
Fernandez R, 2007, COCHRANE DB SYST REV, V18
[7]   Guideline for Prevention of Catheter-Associated Urinary Tract Infections 2009 [J].
Gould, Carolyn V. ;
Umscheid, Craig A. ;
Agarwal, Rajender K. ;
Kuntz, Gretchen ;
Pegues, David A. .
INFECTION CONTROL AND HOSPITAL EPIDEMIOLOGY, 2010, 31 (04) :319-326
[8]   An efficient solution to the retained Foley catheter [J].
Hamilton, Robert J. ;
Jewett, Michael A. S. ;
Finelli, Antonio .
UROLOGY, 2006, 68 (05) :1109-1111
[9]   Incidence and prevention of iatrogenic urethral injuries [J].
Kashefi, Carol ;
Messer, Karen ;
Barden, Rita ;
Sexton, Carolyn ;
Parsons, J. Kellogg .
JOURNAL OF UROLOGY, 2008, 179 (06) :2254-2257
[10]  
Lepor Herbert, 2005, Rev Urol, V7 Suppl 8, pS26