Intra-individual comparison between abdominal virtual mono-energetic spectral and conventional images using a novel spectral detector CT

被引:20
作者
Doerner, Jonas [1 ]
Wybranski, Christian [1 ]
Byrtus, Jonathan [1 ]
Houbois, Christian [1 ]
Hauger, Myriam [1 ]
Heneweer, Carola [1 ]
Siedek, Florian [1 ]
Hickethier, Tilman [1 ]
Hokamp, Nils Grosse [1 ]
Maintz, David [1 ]
Haneder, Stefan [1 ]
机构
[1] Univ Hosp Cologne, Inst Diagnost & Intervent Radiol, Cologne, Germany
来源
PLOS ONE | 2017年 / 12卷 / 08期
关键词
DUAL-ENERGY CT; KILOELECTRON VOLT SETTINGS; ANGIOGRAPHY; RECONSTRUCTION; OPTIMIZATION; QUALITY; ALGORITHM; CONTRAST;
D O I
10.1371/journal.pone.018375
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives To quantitatively and qualitatively assess abdominal arterial and venous phase contrast-enhanced spectral detector computed tomography (SDCT) virtual mono-energetic (MonoE) datasets in comparison to conventional CT reconstructions provided by the same system. Materials and methods Conventional and MonoE images at 40-120 kilo-electron volt (keV) levels with a 10 keV increment as well as 160 and 200 keV were reconstructed in abdominal SDCT datasets of 55 patients. Attenuation, image noise, and contrast-/signal-to-noise ratios (CNR, SNR) of vessels and solid organs were compared between MonoE and conventional reconstructions. Two readers assessed contrast conditions, detail visualization, overall image quality and subjective image noise with both, fixed and adjustable window settings. Results Attenuation, CNR and SNR of vessels and solid organs showed a stepwise increase from high to low keV reconstructions in both contrast phases while image noise stayed stable at low keV MonoE reconstruction levels. Highest levels were found at 40 keV MonoE reconstruction (p< 0.001), respectively. Solid abdominal organs showed a stepwise decrease from low to high energy levels in regard to attenuation, CNR and SNR with significantly higher values at 40 and 50 keV, compared to conventional images. The 70 keV MonoE was comparable to conventional poly-energetic reconstruction (p >= 0.99). Subjective analysis displayed best image quality for the 70 keV MonoE reconstruction level in both phases at fixed standard window presets and at 40 keV if window settings could be adjusted. Conclusion SDCT derived low keV MonoE showed markedly increased CNR and SNR values due to constantly low image noise values over the whole energy spectrum from 40 to 200 keV.
引用
收藏
页数:15
相关论文
共 22 条
  • [21] Dual-Energy CT: Vascular Applications
    Vlahos, Ioannis
    Chung, Raymond
    Nair, Arjun
    Morgan, Robert
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (05) : S87 - S97
  • [22] Dual-Energy CT-Based Monochromatic Imaging
    Yu, Lifeng
    Leng, Shuai
    McCollough, Cynthia H.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 199 (05) : S9 - S15