Intra-individual comparison between abdominal virtual mono-energetic spectral and conventional images using a novel spectral detector CT

被引:20
作者
Doerner, Jonas [1 ]
Wybranski, Christian [1 ]
Byrtus, Jonathan [1 ]
Houbois, Christian [1 ]
Hauger, Myriam [1 ]
Heneweer, Carola [1 ]
Siedek, Florian [1 ]
Hickethier, Tilman [1 ]
Hokamp, Nils Grosse [1 ]
Maintz, David [1 ]
Haneder, Stefan [1 ]
机构
[1] Univ Hosp Cologne, Inst Diagnost & Intervent Radiol, Cologne, Germany
来源
PLOS ONE | 2017年 / 12卷 / 08期
关键词
DUAL-ENERGY CT; KILOELECTRON VOLT SETTINGS; ANGIOGRAPHY; RECONSTRUCTION; OPTIMIZATION; QUALITY; ALGORITHM; CONTRAST;
D O I
10.1371/journal.pone.018375
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Objectives To quantitatively and qualitatively assess abdominal arterial and venous phase contrast-enhanced spectral detector computed tomography (SDCT) virtual mono-energetic (MonoE) datasets in comparison to conventional CT reconstructions provided by the same system. Materials and methods Conventional and MonoE images at 40-120 kilo-electron volt (keV) levels with a 10 keV increment as well as 160 and 200 keV were reconstructed in abdominal SDCT datasets of 55 patients. Attenuation, image noise, and contrast-/signal-to-noise ratios (CNR, SNR) of vessels and solid organs were compared between MonoE and conventional reconstructions. Two readers assessed contrast conditions, detail visualization, overall image quality and subjective image noise with both, fixed and adjustable window settings. Results Attenuation, CNR and SNR of vessels and solid organs showed a stepwise increase from high to low keV reconstructions in both contrast phases while image noise stayed stable at low keV MonoE reconstruction levels. Highest levels were found at 40 keV MonoE reconstruction (p< 0.001), respectively. Solid abdominal organs showed a stepwise decrease from low to high energy levels in regard to attenuation, CNR and SNR with significantly higher values at 40 and 50 keV, compared to conventional images. The 70 keV MonoE was comparable to conventional poly-energetic reconstruction (p >= 0.99). Subjective analysis displayed best image quality for the 70 keV MonoE reconstruction level in both phases at fixed standard window presets and at 40 keV if window settings could be adjusted. Conclusion SDCT derived low keV MonoE showed markedly increased CNR and SNR values due to constantly low image noise values over the whole energy spectrum from 40 to 200 keV.
引用
收藏
页数:15
相关论文
共 22 条
  • [1] Comprehensive Comparison of Virtual Monoenergetic and Linearly Blended Reconstruction Techniques in Third-Generation Dual-Source Dual-Energy Computed Tomography Angiography of the Thorax and Abdomen
    Albrecht, Moritz H.
    Trommer, Jesko
    Wichmann, Julian L.
    Scholtz, Jan-Erik
    Martin, Simon S.
    Lehnert, Thomas
    Vogl, Thomas J.
    Bodelle, Boris
    [J]. INVESTIGATIVE RADIOLOGY, 2016, 51 (09) : 582 - 590
  • [2] Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast
    Albrecht, Moritz H.
    Scholtz, Jan-Erik
    Huesers, Kristina
    Beeres, Martin
    Bucher, Andreas M.
    Kaup, Moritz
    Martin, Simon S.
    Fischer, Sebastian
    Bodelle, Boris
    Bauer, Ralf W.
    Lehnert, Thomas
    Vogl, Thomas J.
    Wichmann, Julian L.
    [J]. EUROPEAN RADIOLOGY, 2016, 26 (06) : 1863 - 1870
  • [3] ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY
    ALVAREZ, RE
    MACOVSKI, A
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) : 733 - 744
  • [4] Value of monoenergetic low-kV dual energy CT datasets for improved image quality of CT pulmonary angiography
    Apfaltrer, Paul
    Sudarski, Sonja
    Schneider, David
    Nance, John W., Jr.
    Haubenreisser, Holger
    Fink, Christian
    Schoenberg, Stefan O.
    Henzler, Thomas
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) : 322 - 328
  • [5] Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolation
    Bamberg, Fabian
    Dierks, Alexander
    Nikolaou, Konstantin
    Reiser, Maximilian F.
    Becker, Christoph R.
    Johnson, Thorsten R. C.
    [J]. EUROPEAN RADIOLOGY, 2011, 21 (07) : 1424 - 1429
  • [6] Blake MA, 2011, RADIOGRAPHICS, V31, P1047, DOI 10.1148/rg.314105159
  • [7] Dual-Energy CT Angiography in Peripheral Arterial Occlusive Disease
    Brockmann, Carolin
    Jochum, Susanne
    Sadick, Maliha
    Huck, Kurt
    Ziegler, Peter
    Fink, Christian
    Schoenberg, Stefan O.
    Diehl, Steffen J.
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2009, 32 (04) : 630 - 637
  • [8] Brown KM, 2012, SPIE MED IMAGING
  • [9] Assessment of a Model-Based, Iterative Reconstruction Algorithm (MBIR) Regarding Image Quality and Dose Reduction in Liver Computed Tomography
    Chang, Won
    Lee, Jeong Min
    Lee, Kyunghee
    Yoon, Jeong Hee
    Yu, Mi Hye
    Han, Joon Koo
    Choi, Byung Ihn
    [J]. INVESTIGATIVE RADIOLOGY, 2013, 48 (08) : 598 - 606
  • [10] Single- and dual-energy CT with monochromatic synchrotron x-rays
    Dilmanian, FA
    Wu, XY
    Parsons, EC
    Ren, B
    Kress, J
    Button, TM
    Chapman, LD
    Coderre, JA
    Giron, F
    Greenberg, D
    Krus, DJ
    Liang, Z
    Marcovici, S
    Petersen, MJ
    Roque, CT
    Shleifer, M
    Slatkin, DN
    Thomlinson, WC
    Yamamoto, K
    Zhong, Z
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (02) : 371 - 387