Multisource modeling of flattening filter free (FFF) beam and the optimization of model parameters

被引:27
作者
Cho, Woong [1 ,2 ]
Kielar, Kayla N. [3 ]
Mok, Ed [3 ]
Xing, Lei [3 ]
Park, Jeong-Hoon [1 ]
Jung, Won-Gyun [1 ]
Suh, Tae-Suk [1 ]
机构
[1] Catholic Univ Korea, Dept Biomed Engn, Coll Med, Seoul 137701, South Korea
[2] Catholic Univ Korea, Res Inst Biomed Engn, Coll Med, Seoul 137701, South Korea
[3] Stanford Univ, Dept Radiat Oncol, Stanford, CA 94305 USA
基金
美国国家科学基金会;
关键词
flattening filter free beam; beam source model; head scatter; optimization; TrueBeam; SUPERPOSITION DOSE CALCULATION; UNFLATTENED PHOTON BEAMS; MONTE-CARLO; RADIATION; FIELDS;
D O I
10.1118/1.3560426
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: With the introduction of flattening filter free (FFF) linear accelerators to radiation oncology, new analytical source models for a FFF beam applicable to current treatment planning systems is needed. In this work, a multisource model for the FFF beam and the optimization of involved model parameters were designed. Methods: The model is based on a previous three source model proposed by Yang et al. ["A three-source model for the calculation of head scatter factors," Med. Phys. 29, 2024-2033 (2002)]. An off axis ratio (OAR) of photon fluence was introduced to the primary source term to generate cone shaped profiles. The parameters of the source model were determined from measured head scatter factors using a line search optimization technique. The OAR of the photon fluence was determined from a measured dose profile of a 40 x 40 cm(2) field size with the same optimization technique, but a new method to acquire gradient terms for OARs was developed to enhance the speed of the optimization process. The improved model was validated with measured dose profiles from 3 x 3 to 40 x 40 cm(2) field sizes at 6 and 10 MV from a TrueBeam (TM) STx linear accelerator. Furthermore, planar dose distributions for clinically used radiation fields were also calculated and compared to measurements using a 2D array detector using the gamma index method. Results: All dose values for the calculated profiles agreed with the measured dose profiles within 0.5% at 6 and 10 MV beams, except for some low dose regions for larger field sizes. A slight overestimation was seen in the lower penumbra region near the field edge for the large field sizes by 1%-4%. The planar dose calculations showed comparable passing rates (>98%) when the criterion of the gamma index method was selected to be 3%/3 mm. Conclusions: The developed source model showed good agreements between measured and calculated dose distributions. The model is easily applicable to any other linear accelerator using FFF beams as the required data include only the measured PDD, dose profiles, and output factors for various field sizes, which are easily acquired during conventional beam commissioning process. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3560426]
引用
收藏
页码:1931 / 1942
页数:12
相关论文
共 32 条
[1]  
[Anonymous], 1999, SPRINGER SCI
[2]   Accelerating Monte Carlo simulations of photon transport in a voxelized geometry using a massively parallel graphics processing unit [J].
Badal, Andreu ;
Badano, Aldo .
MEDICAL PHYSICS, 2009, 36 (11) :4878-4880
[3]   Extraction of the photon spectra from measured beam parameters [J].
Bloch, P ;
McDonough, J .
MEDICAL PHYSICS, 1998, 25 (05) :752-757
[4]   The characterization of unflattened photon beams from a 6 MV linear accelerator [J].
Cashmore, Jason .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (07) :1933-1946
[5]   Assessment of the setup dependence of detector response functions for mega-voltage linear accelerators [J].
Fox, Christopher ;
Simon, Tom ;
Simon, Bill ;
Dempsey, James F. ;
Kahler, Darren ;
Palta, Jatinder R. ;
Liu, Chihray ;
Yan, Guanghua .
MEDICAL PHYSICS, 2010, 37 (02) :477-484
[6]   Delivery time comparison for intensity-modulated radiation therapy with/without flattening filter: a planning study [J].
Fu, WH ;
Dai, JR ;
Hu, YM ;
Han, DS ;
Song, YX .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) :1535-1547
[7]   Determination of penumbral widths from ion chamber measurements [J].
Herrup, D ;
Chu, J ;
Cheung, H ;
Pankuch, M .
MEDICAL PHYSICS, 2005, 32 (12) :3636-3640
[8]   DECONVOLUTION OF DETECTOR SIZE EFFECT FOR SMALL-FIELD MEASUREMENT [J].
HIGGINS, PD ;
SIBATA, CH ;
SISKIND, L ;
SOHN, JW .
MEDICAL PHYSICS, 1995, 22 (10) :1663-1666
[9]   A convolution-superposition dose calculation engine for GPUs [J].
Hissoiny, Sami ;
Ozell, Benoit ;
Despres, Philippe .
MEDICAL PHYSICS, 2010, 37 (03) :1029-1037
[10]   Fast convolution-superposition dose calculation on graphics hardware [J].
Hissoiny, Sami ;
Ozell, Benoit ;
Despres, Philippe .
MEDICAL PHYSICS, 2009, 36 (06) :1998-2005