In vitro aerodynamic characterization of the dose emitted during nebulization of tobramycin high strength solution by novel and jet nebulizer delivery systems

被引:8
作者
Mashat, M. [1 ]
Clark, B. J. [2 ,3 ]
Assi, K. H. [2 ,3 ]
Chrystyn, H. [4 ]
机构
[1] Makkah Toxicol & Drug Res Ctr, Riyadh, Saudi Arabia
[2] Univ Bradford, Sch Pharm, Bradford BD7 1DP, W Yorkshire, England
[3] Univ Bradford, Inst Pharmaceut Innovat, Bradford BD7 1DP, W Yorkshire, England
[4] Univ Huddersfield, Sch Appl Sci, Huddersfield HD1 3DH, W Yorkshire, England
关键词
Cystic fibrosis; High strength tobramycin solution (HSTS); Nebulizer; CEN method; CYSTIC-FIBROSIS; AEROSOLIZED TOBRAMYCIN; INHALED TOBRAMYCIN; ADULT PATIENTS; DEPOSITION; INHALATION; PHARMACOKINETICS; BRAMITOB((R)); COMPRESSOR; MANAGEMENT;
D O I
10.1016/j.pupt.2015.12.003
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
Background: Chronic infections with Pseudomonas aeruginosa are a leading cause of morbidity in patients with cystic fibrosis (CF). The aim of tobramycin inhalation therapy in CF patients with chronic pulmonary infection is to deliver high amounts of drug directly to the site of infection. TOBI (R) is a tobramycin nebulizer solution (300 mg/5 ml) approved by FDA for maintenance therapy for patient with CF. The 20% tobramycin sulfate solution was reported as the optimal and maximal concentration. Methods: Nebulization of high strength tobramycin solution (20% tobramycin sulfate) (HSTS) has been assessed in this study by using different selected high performance nebulizer delivery systems: two different designs of jet nebulizers, and three new nebulizers based on vibrating mesh technology. The aerosol particle size distribution and output characteristics were measured for in vitro performance assessment of the nebulizer systems. The methodology was adapted from the current European standard, EN 13544-1:2001E. Results: The particle size distribution characteristic measurements showed that all tested nebulizers may be suitable for inhalation of HSTS. The mean (SD) of highest percentage of fine particles (<5 mu m) was 77.64 (2.3) % for Sidestream (R), at flow rate 16 L/min. The highest respirable inhaled mass was for Pad LC Plus combined with PariBoyN (R) compressor, with mean (SD) 90.85 (8.6) mg. The mean (SD) of highest drug wastage percentage was 63.9 (3.9) % for Sidestream (R) jet nebulizer combined with compressed air cylinder at flow rate 16 L/min, while the lowest was 2.3 (0.26) % for NE-U22 Omron (R) (high frequency). Conclusions: The HSTS can be nebulized by all tested nebulisers but the high frequency NE-U22 Omron (R) and Aeroneb Go (R) are more efficient. When the HSTS compared to TOBI (R), the respirable inhaled dose was increased to more than 73%. (C) 2016 Published by Elsevier Ltd.
引用
收藏
页码:37 / 42
页数:6
相关论文
共 37 条
[1]  
American Association for Respiratory Care, 1991, RESP CARE, V36, P916
[2]   Distribution and elimination of tobramycin administered in single or multiple daily doses in adult patients with cystic fibrosis [J].
Aminimanizani, A ;
Beringer, PM ;
Kang, J ;
Tsang, L ;
Jelliffe, RW ;
Shapiro, BJ .
JOURNAL OF ANTIMICROBIAL CHEMOTHERAPY, 2002, 50 (04) :553-559
[3]  
Bisgaard H., 2002, DRUG DELIV LUNG, P105
[4]   Comparison of lung deposition of colomycin using the HaloLite and the Pari LC Plus nebulisers in patients with cystic fibrosis [J].
Byrne, NM ;
Keavey, PM ;
Perry, JD ;
Gould, FK ;
Spencer, DA .
ARCHIVES OF DISEASE IN CHILDHOOD, 2003, 88 (08) :715-718
[5]  
Byron P., 1987, PHARM TECH, V11, P42
[6]   DRUG-DELIVERY VIA THE RESPIRATORY-TRACT [J].
BYRON, PR ;
PATTON, JS .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 1994, 7 (01) :49-75
[7]   A formulation of aerosolized tobramycin (Bramitob®) in the treatment of patients with cystic fibrosis and Pseudomonas aeruginosa infection: A double-blind, placebo-controlled, multicenter study [J].
Chuchalin A. ;
Csiszér E. ;
Gyurkovics K. ;
Bartnicka M.T. ;
Sands D. ;
Kapranov N. ;
Varoli G. ;
Preti P.A.M. ;
Mazurek H. .
Pediatric Drugs, 2007, 9 (Suppl 1) :21-31
[8]   FACTORS INFLUENCING THE SIZE DISTRIBUTION OF AEROSOLS FROM JET NEBULIZERS [J].
CLAY, MM ;
PAVIA, D ;
NEWMAN, SP ;
CLARKE, SW .
THORAX, 1983, 38 (10) :755-759
[9]   The choice of jet nebulizer, nebulizing flow, and addition of albuterol affects the output of tobramycin aerosols [J].
Coates, AL ;
MacNeish, CF ;
Meisner, D ;
Kelemen, S ;
Thibert, R ;
MacDonald, J ;
Vadas, E .
CHEST, 1997, 111 (05) :1206-1212
[10]   A comparison of the availability of tobramycin for inhalation from vented vs unvented nebulizers [J].
Coates, AL ;
MacNeish, CF ;
Lands, LC ;
Meisner, D ;
Kelemen, S ;
Vadas, EB .
CHEST, 1998, 113 (04) :951-956