Biomechanical effects of interbody cage height on adjacent segments in patients with lumbar degeneration: a 3D finite element study

被引:7
|
作者
Lu, Xiao [1 ]
Li, Dachuan [1 ]
Wang, Hongli [1 ]
Xia, Xinlei [1 ]
Ma, Xiaosheng [1 ]
Lv, Feizhou [1 ]
Zou, Fei [1 ]
Jiang, Jianyuan [1 ]
机构
[1] Fudan Univ, Huashan Hosp, Dept Orthoped, 12 Middle Wulumuqi Rd, Shanghai 200040, Peoples R China
基金
中国国家自然科学基金;
关键词
TLIF; Interbody cage; Intervertebral height; Degenerative lumbar spine; Biomechanics; INTERVERTEBRAL DISC DEGENERATION; ANTERIOR CERVICAL DISKECTOMY; LOW-BACK-PAIN; MECHANICAL-BEHAVIOR; FUSION; DISEASE; SPINE; DISTRACTION; COMPRESSION; LEVEL;
D O I
10.1186/s13018-022-03220-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective To investigate the biomechanical effects of interbody cage height on adjacent segments in patients with lumbar degeneration undergoing transforaminal lumbar interbody fusion (TLIF) surgery, so as to provide references for selection of interbody cage. Methods The finite element model of normal lower lumbar spine (L3-S1) was built and validated, then constructed three different degenerative segments in L3-L4, and the cages with different height (8, 10, 12, 14 mm) were implanted into L4-L5 disc. All the twelve models were loaded with pure moment of 7.5 N m to produce flexion, extension, lateral bending and axial rotation motions on lumbar spine, and the effects of cage height on range of motion (RoM) and intervertebral pressure in lumbar spine were investigated. Results The RoM of adjacent segments and the maximum stress of intervertebral discs increased with the increase in cage height, but this trend was not obvious in mild and moderate degeneration groups. After implantation of four different height cages (8, 10, 12, 14 mm), the RoM of L3/L4 segment reached the maximum during extension. The RoM of mild degeneration group was 2.07 degrees, 2.45 degrees, 2.48 degrees, 2.54 degrees, that of moderate degeneration group was 1.79 degrees, 1.97 degrees, 2.05 degrees, 2.05 degrees, and that of severe degeneration group was 1.43 degrees, 1.66 degrees, 1.74 degrees, 1.74 degrees. The stress of L3-L4 intervertebral disc reached the maximum during flexion. The maximum stress of L3-L4 intervertebral disc was 20.16 MPa, 20.28 MPa, 20.31 MPa and 20.33 MPa in the mild group, 20.58 MPa, 20.66 MPa, 20.71 MPa and 20.75 MPa in the moderate group, and 21.27 MPa, 21.40 MPa, 21.50 MPa and 21.60 MPa in the severe group. Conclusion For patients with mild-to-moderate lumbar degenerative disease who need to undergo TLIF surgery, it is recommended that the height of fusion cage should not exceed the original intervertebral space height by 2 mm, while for patients with severe degeneration, a fusion cage close to the original intervertebral height should be selected as far as possible, and the intervertebral space should not be overstretched.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Comparison of posterior versus transforaminal lumbar interbody fusion using finite element analysis. Influence on adjacent segmental degeneration
    Tang, Shujie
    SAUDI MEDICAL JOURNAL, 2015, 36 (08) : 993 - 996
  • [42] Study of mechanical effects of lumbar disc arthroplasty on facet joints at the index level/adjacent levels by using a validated finite element analysis
    Zot, Francois
    Ben-Brahim, Estelle
    Severyns, Mathieu
    Ledoux, Yann
    Mesnard, Michel
    Caille, Laetitia
    Swennen, Cecile
    Teyssedou, Simon
    Moufid, Abdollah-Yassine
    Germaneau, Arnaud
    Vendeuvre, Tanguy
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [43] The biomechanical effects of foraminoplasty of different areas under lumbar percutaneous endoscopy on intervertebral discs A 3D finite element analysis
    Xie, YiZhou
    Zhou, Qun
    Wang, Xinling
    Jian, Qiang
    Fan, Xiaohong
    Yu, Yang
    Gu, Dangwei
    Wu, WeiDong
    MEDICINE, 2020, 99 (17) : E19847
  • [44] Comparison of cage application modality in posterior lumbar interbody fusion with posterior instrumentation-A finite element study
    Tsuang, Yang-Hwei
    Chiang, Yueh-Feng
    Hung, Chih-Yi
    Wei, Hung-Wen
    Huang, Chang-Hung
    Cheng, Cheng-Kung
    MEDICAL ENGINEERING & PHYSICS, 2009, 31 (05) : 565 - 570
  • [45] A novel narrow surface cage for full endoscopic oblique lateral lumbar interbody fusion: A finite element study
    Ling, Qinjie
    He, Erxing
    Zhang, Huanliang
    Lin, Hui
    Huang, Wenhua
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2019, 24 (06) : 991 - 998
  • [46] Influence of cement-augmented pedicle screws with different volumes of polymethylmethacrylate in osteoporotic lumbar vertebrae over the adjacent segments: a 3D finite element analysis
    Guo, Hui-zhi
    Zhang, Shun-cong
    Guo, Dan-qing
    Ma, Yan-huai
    Yuan, Kai
    Li, Yong-xian
    Peng, Jian-cheng
    Li, Jing-lan
    Liang, De
    Tang, Yong-chao
    BMC MUSCULOSKELETAL DISORDERS, 2020, 21 (01)
  • [47] The Morphological Changes in Adjacent Segments Amongst Patients Receiving Anterior and Oblique Lumbar Interbody Fusion: A Retrospective Study
    Tung, Kuan-Kai
    Hsu, Fang-Wei
    Ou, Hsien-Che
    Chen, Kun-Hui
    Pan, Chien-Chou
    Lu, Wen-Xian
    Chin, Ning-Chien
    Shih, Cheng-Min
    Wu, Yun-Che
    Lee, Cheng-Hung
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (23)
  • [48] The biomechanical effects of treating double-segment lumbar degenerative diseases with unilateral fixation through interlaminar fenestration interbody fusion surgery: a three-dimensional finite element study
    Feng, Kai-Kai
    Xiang, Xiao-Bing
    Li, Cheng-Peng
    Gao, Kai-Yin
    Zhang, Wei-Wei
    Nie, Zhen-Yu
    Liao, Wen-Xin
    Zhao, Dong
    Cao, Guang-Ru
    BMC MUSCULOSKELETAL DISORDERS, 2025, 26 (01)
  • [49] Biomechanical investigation on the influence of the regional material degeneration of an intervertebral disc in a lower lumbar spinal unit: A finite element study
    Masni-Azian
    Tanaka, Masao
    COMPUTERS IN BIOLOGY AND MEDICINE, 2018, 98 : 26 - 38
  • [50] Biomechanical Effects of Zero-P Height on Anterior Cervical Discectomy and Fusion: A Finite Element Study
    Huang, Cheng-yi
    He, Jun-bo
    Wang, Xing-jing
    Wu, Ting-kui
    Wang, Bei-yu
    Xu, Jin
    Liu, Hao