Validation of LensHooke® X1 PRO and Computer-Assisted Semen Analyzer Compared with Laboratory-Based Manual Semen Analysis

被引:14
作者
Agarwal, Ashok [1 ]
Selvam, Manesh Kumar Panner [1 ]
Ambar, Rafael F. [1 ,2 ,3 ]
机构
[1] Cleveland Clin, Amer Ctr Reprod Med, Mail Code X-11,10681 Carnegie Ave, Cleveland, OH 44195 USA
[2] Ideia Fertil Inst Human Reprod, Urol Dept, Ctr Univ Saude ABC, Androl Grp, Santo Andre, SP, Brazil
[3] Hope Clin Human Reprod, Sao Paulo, Brazil
关键词
Computer assisted semen analyzer; Lensllooke (R) X1 PRO; IVOS CASA; Semen analysis; MALE-FERTILITY; SPERM QUALITY; SQA-V; FREQUENCY; SYSTEMS;
D O I
10.5534/wjmh.200185
中图分类号
R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
摘要
Purpose: To compare two automated semen quality analysis systems (lerisllooke (R) X1 PRO [X1 PRO]) and IVOS CASA) for accuracy, precision and agreement with laboratory-based manual semen analysis (MSA). Materials and Methods: Semen samples (n=31) were obtained from normozoospermic healthy male volunteers and infertile men with a minimum abstinence period between 2-3 days. After complete liquefaction, 101 seminal aliquots were prepared and tested according to WHO 5th Edition (2010) guidelines. The results obtained by X1 PRO and IVOS CASA were compared with that of MSA. Additionally, 10 samples were used to evaluate the intra- and inter-rater agreement for X1 PRO and MSA. Results: The semen parameters (sperm concentration, total, and progressive motility) showed strong correlation and agreement for both automated semen analyzers and MSA (Spearman's rank correlation >= 0.92, p<0.0001). X1 PRO and IVOS CLASA were able to differentiate samples with abnormal concentration with a positive predictive value (PPV) of 100%. Furthermore, the PPV for X1 PRO (86.5%) was higher than that for IVOS CASA (71.7%) in differentiating samples with abnormal motility. The X1 PRO device showed a high PPV (97.7%) in identifying normal sperm forms compared to MSA. Semen parameters evaluated showed a high inter-rater (kappa >0.91) and intra-rater (kappa >0.92) agreement for X1 PRO compared with MSA. Conclusions: Both automated semen analyzers demonstrated a high level of concordance and their performance was comparable with MSA analysis. Furthermore, high-levels of inter-and intra-rater reliability for semen analysis indicate that the new X1 PRO can be used in a clinical laboratory to offer accurate and quick test results.
引用
收藏
页码:496 / 505
页数:10
相关论文
共 27 条
  • [1] Agarwal A, 2016, IEEE INT POWER ELEC, P75
  • [2] Agarwal A., 2016, Andrological Evaluation of Male Infertility: A Laboratory Guide, P47, DOI [10.1007/978-3-319-26797-5_5, DOI 10.1007/978-3-319-26797-5_5]
  • [3] Agarwal A., 2016, Andrological Evaluation of Male Infertility 109-112, DOI DOI 10.1007/978-3-319-26797-5_8
  • [4] Automation is the key to standardized semen analysis using the automated SQA-V sperm quality analyzer
    Agarwal, Ashok
    Sharma, Rakesh K.
    [J]. FERTILITY AND STERILITY, 2007, 87 (01) : 156 - 162
  • [5] Automation of human semen analysis using a novel artificial intelligence optical microscopic technology
    Agarwal, Ashok
    Henkel, Ralf
    Huang, Chun-Chia
    Lee, Maw-Sheng
    [J]. ANDROLOGIA, 2019, 51 (11)
  • [6] Agarwal A, 2019, WORLD J MENS HEALTH, V37, P296
  • [7] Usefulness of Sperm Quality Analyzer-V (SQA-V) for the assessment of sperm quality in infertile men
    Akashi, T
    Mizuno, I
    Okumura, A
    Fuse, H
    [J]. ARCHIVES OF ANDROLOGY, 2005, 51 (06): : 437 - 442
  • [8] [Anonymous], 2010, WHO laboratory manual for the examination and processing of human semen, P287
  • [9] [Anonymous], 2004, The DHS Wealth Index
  • [10] Diagnostic value of routine semen analysis in clinical andrology
    Baskaran, Saradha
    Finelli, Renata
    Agarwal, Ashok
    Henkel, Ralf
    [J]. ANDROLOGIA, 2021, 53 (02)