Is the novel olecranon tension plate a valid alternative to tension band wiring of olecranon fractures? A biomechanical study on cadaver bones

被引:16
作者
Gruszka, Dominik [1 ]
Arand, Charlotte [1 ]
Greenfield, Julia [1 ,2 ]
Nowak, Tobias E. [1 ]
Kuechle, Raphael [1 ]
Kuhn, Sebastian [1 ]
Rommens, Pol M. [1 ]
机构
[1] Johannes Gutenberg Univ Mainz, Dept Orthopaed & Traumatol, Univ Med Ctr, Langenbeckstr 1, D-55131 Mainz, Germany
[2] Univ Reims, GRESPI EA 4694, Fac Sport Sci, Reims, France
关键词
Tension band wiring; Olecranon complex fracture; Olecranon tension plate; Biomechanical study; Olecranon fixation; FIXATION; COMPRESSION;
D O I
10.1007/s00402-017-2760-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Tension band wiring (TBW) used for olecranon fractures is afflicted with a high rate of complications. The aim of this study was to compare the stability of the TBW versus an alternative, novel low-profile olecranon tension plate (OTP) with angular stable screws in a simulated complex fracture model. Nine fresh-frozen pairs of cadaver proximal ulnae with an oblique osteotomy and an additional wedge fragment simulating an unstable fracture were tested. The TBW and OTP were implanted pairwise. The elbow motion was simulated in a single-muscle model and ranged from full extension to 90A degrees of flexion. The pulling force of the triceps tendon ranged from 50 to 200 N. The displacement of the fracture fragments was measured with a video motion analysis system over 304 cycles. Data were assessed statistically using the Wilcoxon signed-rank test. The cyclic loading tests showed mean loosening of the fracture fragments at the articular surface of 0.56 mm using TBW (SD 0.65) and 0.15 mm for OTP (SD 0.39). There was no statistical significance in loosening between the two constructs (p = 0.31). No plate breakage or screw loosening occurred. The low-profile OTP, using the principles of pre-tensioning, lag, cortical, and angular stable screws together, demonstrated similar stability after olecranon fracture fixation when compared with the TBW technique and a lag screw in the olecranon fracture model with a third wedge fragment.
引用
收藏
页码:1651 / 1658
页数:8
相关论文
共 36 条
  • [11] Olecranon tension plating or olecranon tension band wiring? A comparative biomechanical study
    Gruszka, Dominik
    Arand, Charlotte
    Nowak, Tobias
    Dietz, Sven-Oliver
    Wagner, Daniel
    Rommens, Pol
    [J]. INTERNATIONAL ORTHOPAEDICS, 2015, 39 (05) : 955 - 960
  • [12] HUME MC, 1992, CLIN ORTHOP RELAT R, P229
  • [13] Cyclic loading of olecranon fracture fixation constructs
    Hutchinson, DT
    Horwitz, DS
    Ha, G
    Thomas, CW
    Bachus, KN
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2003, 85A (05) : 831 - 837
  • [14] Comminuted fractures of the olecranon - Management by bone graft from the iliac crest and multiple tension-band wiring
    Ikeda, M
    Fukushima, Y
    Kobayashi, Y
    Oka, Y
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (06): : 805 - 808
  • [15] Fractures of the olecranon
    Karlsson, MK
    Hasserius, R
    Karlsson, C
    Besjakov, J
    Josefsson, PO
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (403) : 205 - 212
  • [16] Kozin S H, 1996, J Shoulder Elbow Surg, V5, P442, DOI 10.1016/S1058-2746(96)80016-4
  • [17] LATERAL TENSION BAND - DEMONSTRATED FOR FRACTURES OF OLECRANON
    LABITZKE, R
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1975, 81 (03) : 193 - 198
  • [18] K-wire position in tension band wiring of the Olecranon -: A comparison of two techniques
    Mullett, JH
    Shannon, F
    Nöel, J
    Lawlor, G
    Lee, TC
    O'Rourke, SK
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 (06): : 427 - 431
  • [19] Olecranon fractures
    Newman, S. D. S.
    Mauffrey, C.
    Krikler, S.
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (06): : 575 - 581
  • [20] Nordin M, 2001, Basic biomechanics of the musculoskeletal system