Laboratory evaluation of the optical properties of two extended-depth-of-focus intraocular lenses

被引:29
作者
Chae, Sue Hey [1 ]
Son, Hyeck Soo [2 ,3 ]
Khoramnia, Ramin [2 ,3 ]
Lee, Kyung Heon [1 ]
Choi, Chul Young [4 ]
机构
[1] Sungmo Eye Hosp, Busan, South Korea
[2] Heidelberg Univ, David J Apple Int Lab Ocular Pathol, Dept Ophthalmol, Heidelberg, Germany
[3] Heidelberg Univ, IVCRC, Dept Ophthalmol, Heidelberg, Germany
[4] Sungkyunkwan Univ, Kangbuk Samsung Hosp, Dept Ophthalmol, Sch Med, Seoul, South Korea
关键词
Optical quality; Intraocular lenses; Extended-depth-of-focus; Pupil size; IMAGE QUALITY; PERFORMANCE; ABERRATION;
D O I
10.1186/s12886-020-1332-6
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background To experimentally compare the optical performance of two different Extended-Depth-of-Focus (EDOF) intraocular lenses (IOLs) using a standardized optical bench set-up. Methods In this experimental study, following IOLs were assessed: the TECNIS (R) Symfony ZXR00 (Johnson&Johnson, Santa Ana, USA) and the AT LARA 829MP (Carl Zeiss Meditec, Jena, Germany) IOLs. The through-focus modulation transfer function (MTF) values were measured at a spatial frequency of 50 lp/mm and at aperture sizes of 2, 3, and 4.5 mm. Each IOL was measured while centered using ISO 11979-2 Model 1 (aberration-free) and Model 2 (+ 0.28 mu m spherical aberration) corneas. United States Air Force (USAF) target images were also recorded for a qualitative evaluation. Results At 2 mm pupil with ISO1 cornea, the primary and secondary foci of both IOLs appeared to merge, providing an elongated depth of focus. At 3 and 4.5 mm pupil sizes, the through-focus MTF curves of both IOLs showed a bifocal-like V-pattern. While the Symfony IOL showed an overall superior MTF values when measured with the ISO2 cornea, the opposite propensity could be observed with the AT LARA IOL. This optical behavior could be qualitatively confirmed by the USAF target images. Conclusions Although the two EDOF IOLs share similarities in their optical properties, the main difference lies in their optical design and performance with respect to spherical aberration. Such characteristics should be taken into account during IOL and patient selection.
引用
收藏
页数:7
相关论文
共 20 条
  • [1] Image quality with multifocal intraocular lenses and the effect of pupil size -: Comparison of refractive and hybrid refractive-diffractive designs
    Artigas, Jose M.
    Menezo, Jose L.
    Peris, Cristina
    Felipe, Adelina
    Diaz-Llopis, Manuel
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (12) : 2111 - 2117
  • [2] Breyer DRH, 2017, ASIA-PAC J OPHTHALMO, V6, P339, DOI 10.22608/APO.2017186
  • [3] In vitro optical quality comparison between the Mini WELL Ready progressive multifocal and the TECNIS Symfony
    Dominguez-Vicent, Alberto
    Juan Esteve-Taboada, Jose
    Del Aguila-Carrasco, Antonio J.
    Ferrer-Blasco, Teresa
    Montes-Mico, Robert
    [J]. GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2016, 254 (07) : 1387 - 1397
  • [4] Through-Focus Vision Performance and Light Disturbances of 3 New Intraocular Lenses for Presbyopia Correction
    Escandon-Garcia, Santiago
    Ribeiro, Filomena J.
    McAlinden, Colm
    Queiros, Antonio
    Gonzalez-Meijome, Jose M.
    [J]. JOURNAL OF OPHTHALMOLOGY, 2018, 2018
  • [5] Clinically Relevant Optical Properties of Bifocal, Trifocal, and Extended Depth of Focus Intraocular Lenses
    Gatinel, Damien
    Loicq, Jerome
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2016, 32 (04) : 273 - +
  • [6] Comparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench
    Gatinel, Damien
    Houbrechts, Yvette
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2013, 39 (07) : 1093 - 1099
  • [7] Objective evaluation of through-focus optical performance of presbyopia-correcting intraocular lenses using an optical bench system
    Kim, Myoung Joon
    Zheleznyak, Len
    MacRae, Scott
    Tchah, Hungwon
    Yoon, Geunyoung
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (07) : 1305 - 1312
  • [8] Longitudinal Chromatic Aberration and Polychromatic Image Quality Metrics of Intraocular Lenses
    Labuz, Grzegorz
    Papadatou, Eleni
    Khoramnia, Ramin
    Auffarth, Gerd U.
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2018, 34 (12) : 832 - +
  • [9] Multifocal versus monofocal intraocular lenses in cataract surgery
    Leyland, M
    Zinicola, E
    [J]. OPHTHALMOLOGY, 2003, 110 (09) : 1789 - 1798
  • [10] Performance of presbyopia-correcting intraocular lenses in distance optical bench tests
    Maxwell, W. Andrew
    Lane, Stephen S.
    Zhou, Fan
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (01) : 166 - 171