Median of patient results as a tool for assessment of analytical stability

被引:17
作者
Jorgensen, Lars Monster [1 ]
Hansen, Steen Ingemann [1 ]
Petersen, Per Hyltoft [1 ,2 ]
Soletormos, Gyorgy [1 ]
机构
[1] Univ Copenhagen, Nordsjaellands Hosp Hillerod, Dept Clin Biochem, DK-3400 Hillerod, Denmark
[2] Univ Bergen, Sect Gen Practice, Norwegian Qual Improvement Primary Care Labs NOKL, Bergen, Norway
关键词
Analytical stability; Monthly patient median; Performance goals for bias; Common analytes; Frozen reference serum; EXTERNAL QUALITY ASSESSMENT; OF-AMERICAN-PATHOLOGISTS; LONG-TERM STABILITY; REFERENCE INTERVALS; ASSESSMENT SCHEMES; PERFORMANCE; GOALS; STANDARDIZATION; TRACEABILITY; AVERAGE;
D O I
10.1016/j.cca.2015.04.024
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: In spite of the well-established external quality assessment and proficiency testing surveys of analytical quality performance in laboratory medicine, a simple tool to monitor the long-term analytical stability as a supplement to the internal control procedures is often needed. Method: Patient data from daily internal control schemes was used for monthly appraisal of the analytical stability. This was accomplished by using the monthly medians of patient results to disclose deviations from analytical stability, and by comparing divergences with the quality specifications for allowable analytical bias based on biological variation. Results: Seventy five percent of the twenty analytes achieved on two COBASs INTEGRA 800 instruments performed in accordance with the optimum and with the desirable specifications for bias. Discussion: Patient results applied in analytical quality performance control procedures are the most reliable sources of material as they represent the genuine substance of the measurements and therefore circumvent the problems associated with non-commutable materials in external assessment. Conclusion: Patient medians in the monthly monitoring of analytical stability in laboratory medicine are an inexpensive, simple and reliable tool to monitor the steadiness of the analytical practice. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:186 / 191
页数:6
相关论文
共 29 条
[1]   A NORDIC REFERENCE SERUM SUITABLE FOR USE AS TRUENESS CONTROL IN THE CLINICAL ROUTINE LABORATORY [J].
BLAABJERG, O ;
ELG, P ;
GERHARDT, W ;
HELLSING, K ;
OLAFSDOTTIR, E ;
PENTTILA, I ;
PETERSEN, PH ;
STEENSLAND, H ;
ULDALL, A .
UPSALA JOURNAL OF MEDICAL SCIENCES, 1993, 98 (03) :405-412
[2]  
BULL BS, 1974, AM J CLIN PATHOL, V61, P473
[3]  
CEMBROWSKI GS, 1985, AM J CLIN PATHOL, V83, P337, DOI 10.1093/ajcp/83.3.337
[4]   Proposals for setting generally applicable quality goals solely based on biology [J].
Fraser, CG ;
Petersen, PH ;
Libeer, JC ;
Ricos, C .
ANNALS OF CLINICAL BIOCHEMISTRY, 1997, 34 :8-12
[5]   ANALYTICAL GOALS FOR THE ACCEPTANCE OF COMMON REFERENCE INTERVALS FOR LABORATORIES THROUGHOUT A GEOGRAPHICAL AREA [J].
GOWANS, EMS ;
PETERSEN, PH ;
BLAABJERG, O ;
HORDER, M .
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1988, 48 (08) :757-764
[6]  
HOFFMANN RG, 1965, AM J CLIN PATHOL, V43, P134
[7]   A category 1 EQA scheme for comparison of laboratory performance and method performance: An international pilot study in the framework of the Calibration 2000 project [J].
Jansen, Rob ;
Jassam, Nuthar ;
Thomas, Annette ;
Perich, Carmen ;
Fernandez-Calle, Pilar ;
Faria, Ana Paula ;
Correia, Helena ;
Barth, Julian H. ;
Weykamp, Cas ;
Cobbaert, Christa ;
Thelen, Marc ;
Ricos, Carmen .
CLINICA CHIMICA ACTA, 2014, 432 :90-98
[8]  
Kenny D, 1999, SCAND J CLIN LAB INV, V59, P585
[9]   Establishment of Outcome-Related Analytic Performance Goals [J].
Klee, George G. .
CLINICAL CHEMISTRY, 2010, 56 (05) :714-722
[10]   Clinical interpretation of reference intervals and reference limits. A plea for assay harmonization [J].
Klee, GG .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2004, 42 (07) :752-757