Biomechanical Comparison of Dual and Posterior Locking Plates in an Ex Vivo Comminuted Olecranon Fracture Model

被引:2
作者
Sobel, Andrew D. [1 ,2 ,3 ]
Babu, Jacob M. [2 ]
Blood, Travis D. [2 ]
Paxton, E. Scott [2 ]
机构
[1] St Lukes Univ Hlth Network, Dept Orthopaed, Bethlehem, PA USA
[2] Brown Univ, Dept Orthopaed, Warren Alpert Med Sch, Providence, RI USA
[3] 501 Cetronia Rd,Suite 125, Allentown, PA 18104 USA
来源
JOURNAL OF HAND SURGERY-AMERICAN VOLUME | 2022年 / 47卷 / 08期
关键词
Biomechanics; dual plating; elbow; olecranon fracture; SURGICAL-TREATMENT; FIXATION; MANAGEMENT; SUTURE;
D O I
10.1016/j.jhsa.2021.07.029
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Comminuted olecranon fractures are commonly fixed with posterior locking plates (PLPs). Though biomechanically validated, this method comes with risks of symptomatic implant prominence and wound breakdown. Dual locking plates (DLPs) placed on the medial and lateral surfaces of the olecranon theoretically avoid these risks and allow for fixation of proximal fracture fragments in multiple planes. A biomechanical comparison of DLP and PLP fixation would help to validate the use of DLPs in comminuted olecranon fractures. Methods Seven matched pairs of cadaveric upper extremities were evaluated with quantitative computed tomography scans to evaluate bone mineral density (BMD). Osteotomies simu-lating comminuted olecranon fractures (Mayo Type IIB) were created and the specimens were fixed either with variable angle PLPs or variable angle DLPs. The specimens were then cyclically loaded and loaded to failure. The ultimate strength, fracture displacement, and mechanism of failure were recorded and compared across groups. The correlation between BMD and ultimate strength was analyzed.Results The mean total BMD was 0.79 g/cm2 (SD, 0.14 g/cm2). No specimen failed during cyclic testing. Five of 7 PLP specimens and 5 of 7 DLP specimens failed by fracture through the proximal screws though in different planes (sagittal vs axial splits, respectively). The mean ultimate strengths of the PLP (1077 N [SD, 462 N]) and DLP (1241 N [SD, 506 N]) groups were similar. There was a linear relationship between ultimate strength and BMD (R2 1/4 0.33). Conclusion Dual locking plates display biomechanical properties that suggest that they can be used in the fixation of comminuted olecranon fractures. Catastrophic failure of the fixation constructs occurs around stress risers at the proximal screws rather than due to displacement at the fracture itself.Clinical relevance Dual locking plate constructs can be considered for use in the fixation of comminuted olecranon fractures. (J Hand Surg Am. 2022;47(8):796.e1-e6. Copyright (c) 2022 by the American Society for Surgery of the Hand. All rights reserved.)
引用
收藏
页码:796.e1 / 796.e6
页数:6
相关论文
共 24 条
[1]   A biomechanical comparison of multidirectional nail and locking plate fixation in unstable olecranon fractures [J].
Argintar, Evan ;
Martin, Benjamin D. ;
Singer, Andrea ;
Hsieh, Adam H. ;
Edwards, Scott .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2012, 21 (10) :1398-1405
[2]   Outcome of plate fixation of olecranon fractures [J].
Bailey, CS ;
MacDermid, J ;
Patterson, SD ;
King, GJW .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2001, 15 (08) :542-548
[3]   Clinical Evaluation of Locking Compression Plate Fixation for Comminuted Olecranon Fractures [J].
Buijze, Geert ;
Kloen, Peter .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2009, 91A (10) :2416-2420
[4]   Is tension band wiring technique the "gold standard" for the treatment of olecranon fractures? A long term functional outcome study [J].
Chalidis B.E. ;
Sachinis N.C. ;
Samoladas E.P. ;
Dimitriou C.G. ;
Pournaras J.D. .
Journal of Orthopaedic Surgery and Research, 3 (1)
[5]   Factors Associated With Reoperation After Fixation of Displaced Olecranon Fractures [J].
Claessen, Femke M. A. P. ;
Braun, Yvonne ;
Peters, Rinne M. ;
Dyer, George ;
Doornberg, Job N. ;
Ring, David .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2016, 474 (01) :193-200
[6]   Nonoperative Management of Displaced Olecranon Fractures in Low-Demand Elderly Patients [J].
Duckworth, Andrew D. ;
Bugler, Kate E. ;
Clement, Nicholas D. ;
Court-Brown, Charles M. ;
McQueen, Margaret M. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A (01) :67-72
[7]   The epidemiology of fractures of the proximal ulna [J].
Duckworth, Andrew D. ;
Clement, Nicholas D. ;
Aitken, Stuart A. ;
Court-Brown, Charles M. ;
McQueen, Margaret M. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2012, 43 (03) :343-346
[8]   Comparison of Olecranon Plate Fixation in Osteoporotic Bone: Do Current Technologies and Designs Make a Difference? [J].
Edwards, Scott G. ;
Martin, Benjamin D. ;
Fu, Rose H. ;
Gill, Joseph M. ;
Nezhad, Mani K. ;
Orr, Jeffrey A. ;
Ferrucci, Allen M. ;
Love, James M. ;
Booth, Richard ;
Singer, Andrea ;
Hsieh, Adam H. .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2011, 25 (05) :306-311
[9]   Biomechanical Assessment of Locking Plate Fixation of Comminuted Proximal Olecranon Fractures [J].
Fantry, Amanda ;
Sobel, Andrew ;
Capito, Nicholas ;
Hodax, Jonathan D. ;
Pidgeon, Tyler ;
Koruprolu, Sarath ;
Rich, Ryan ;
Fava, Joseph L. ;
Green, Andrew ;
Paxton, Scott .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2018, 32 (11) :E445-E450
[10]   Comminuted olecranon fractures: A comparison of plating methods [J].
Gordon, MJ ;
Budoff, JE ;
Yeh, ML ;
Luo, ZP ;
Noble, PC .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 2006, 15 (01) :94-99