A comparative Monte Carlo study on 6MV photon beam characteristics of Varian 21EX and Elekta SL-25 linacs

被引:0
作者
Mesbahi, A. [1 ]
Mehnati, P. [1 ]
Keshtkar, A. [1 ]
机构
[1] Tabriz Univ Med Sci, Dept Med Phys, Sch Med, Tabriz, Iran
来源
IRANIAN JOURNAL OF RADIATION RESEARCH | 2007年 / 5卷 / 01期
关键词
dosimetric features; Elekta SL-25; Varian; 21EX; Monte Carlo method;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background. Monte Carlo method (MC) has played an important role in design and optimization of medical linacs head and beam modeling. The purpose of this study was to compare photon beam features of two commercial linacs, Varian 21EX and Elekta SL-25 using MCNP4C MC code. Materials and Methods: The 6MV photon beams of Varian 21EX and Elekta Sl-25 linacs were simulated based on manufacturers provided information. Photon energy spectra and absolute absorbed dose values were calculated for field sizes of 10x10 and 20x20 cm(2). Also, contamination electron spectra for field size of 20x20 cm(2) were scored for both linacs. Results: Our results showed that the relative absorbed dose values and contamination electron spectrum were similar and comparable, but photon fluence and absolute absorbed dose values were 17% and 13% higher for Varian linac respectively for the field size of 10x10 cm(2). Conclusion: Despite the differences in head components of two commercial linacs, their relative depth dose values were very close to each other. The absolute dose per incident electron showed some discrepancy, as well. Thus, this study suggests the use of absolute absorbed dose values as an invaluable factor when different linacs head are compared using Monte Carlo Method.
引用
收藏
页码:23 / 30
页数:8
相关论文
共 23 条
[1]  
Briesmeister J.F., 2000, LA13709M LOS AL NAT
[2]   Simulation and measurement of air generated electron contamination in radiotherapy [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN ;
Carolan, M ;
Metcalfe, PE .
RADIATION MEASUREMENTS, 2000, 32 (02) :105-111
[3]   Energy spectra, angular spread, fluence profiles and dose distributions of 6 and 18 MV photon beams: results of Monte Carlo simulations for a Varian 2100EX accelerator [J].
Ding, GX .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (07) :1025-1046
[4]   Comparison of beam characteristics of a gold x-ray target and a tungsten replacement target [J].
Faddegon, B ;
Egley, B ;
Steinberg, T .
MEDICAL PHYSICS, 2004, 31 (01) :91-97
[5]   ANGULAR-DISTRIBUTION OF BREMSSTRAHLUNG FROM 15-MEV ELECTRONS INCIDENT ON THICK TARGETS OF BE, AL, AND PB [J].
FADDEGON, BA ;
ROSS, CK ;
ROGERS, DWO .
MEDICAL PHYSICS, 1991, 18 (04) :727-739
[6]   Monte Carlo dose calculations for a 6-MV photon beam in a thorax phantom [J].
Farajollahi A. ;
Mesbahi A. .
Radiation Medicine, 2006, 24 (4) :269-276
[7]  
MACKIE TR, 1984, USE COMPUTERS RAD TH
[8]   A Monte Carlo approach to electron contamination sources in the Saturne-25 and-41 [J].
Malataras, G ;
Kappas, C ;
Lovelock, DMJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (09) :2435-2446
[9]   Characterization of electron contamination in megavoltage photon beams [J].
Medina, AL ;
Teijeiro, A ;
Garcia, J ;
Esperon, J .
MEDICAL PHYSICS, 2005, 32 (05) :1281-1292
[10]   Development and commissioning of a Monte Carlo photon beam model for Varian Clinac 2100EX linear accelerator [J].
Mesbahi, A ;
Reilly, AJ ;
Thwaites, DI .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (06) :656-662