Finite Element Assessment of a Hybrid Proposal for Hip Stem, from a Standardized Base and Different Activities

被引:3
作者
Guzman, Manuel [1 ]
Durazo, Emmanuel [1 ]
Ortiz, Alejandro [1 ]
Sauceda, Israel [1 ]
Siqueiros, Miriam [1 ]
Gonzalez, Luis [1 ]
Jimenez, David [2 ]
机构
[1] Univ Autonoma Baja California, Fac Ingn, Mexicali 21280, Baja California, Mexico
[2] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
来源
APPLIED SCIENCES-BASEL | 2022年 / 12卷 / 16期
关键词
hip joint; aseptic loosening; standardized base; finite element method; hybrid implant; FEMORAL STEM; COMPOSITE-MATERIALS; STRESS; DESIGN; JOINT; PROSTHESIS; CEMENTLESS; POLYMER; BIOMATERIALS;
D O I
10.3390/app12167963
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Choosing a suitable prosthesis to restore the functionality of the hip joint is a complex problem. The stem geometries, materials, and type of hip damage are critical factors for avoiding potential issues (aseptic loosening, fracture, and natural wear and tear). Comparing the available stems to select the best option is not straightforward because of the various loads and boundary conditions used in the tests, making the process difficult to compare the advantages and disadvantages among them. This work proposes stem assessment using a standardized base (generated from a literature review and ISO standards) to compare the stem geometries and present a new hybrid design to improve performance using the best qualities of the implants reported in the literature review. Sixteen hip prostheses were evaluated with the finite element method (FEM) using the same boundary and loading conditions through multi-objective analysis (von Mises stress and strain). Consequently, a hybrid geometry proposal was obtained by assessing specific points through the stem length (medial and lateral region) to define the cross-section (trapezoidal) and the new profile. The new hybrid implant proposal presented a stress reduction of 9.6% when compared to the reference implant P2-T (the implant with the best behavior) in the most critical activity (activity 4) using a titanium alloy. A similar stress reduction of 9.98% was obtained using ASTM F2996-13 and ISO 7206-4:2010(E) standards.
引用
收藏
页数:23
相关论文
共 62 条
[1]  
Abd-Elhaleem M, 2018, INT C COMP CONTR EL, P1, DOI [10.1109/ICCCEEE.2018.8515764, DOI 10.1109/ICCCEEE.2018.8515764]
[2]  
Affatato S, 2015, BIOMECHANICS KNEE
[3]  
Affatato S., 2014, Perspectives in Total Hip Arthroplasty, P3, DOI DOI 10.1533/9781782420392.1.3
[4]   Advanced biomaterials in hip joint arthroplasty. A review on polymer and ceramics composites as alternative bearings [J].
Affatato, Saverio ;
Ruggiero, Alessandro ;
Merola, Massimiliano .
COMPOSITES PART B-ENGINEERING, 2015, 83 :276-283
[5]   Current and future biocompatibility aspects of biomaterials for hip prosthesis [J].
Aherwar, Amit ;
Singh, Amit K. ;
Patnaik, Amar .
AIMS BIOENGINEERING, 2016, 3 (01) :23-43
[6]   Changes in the stress in the femoral head neck junction after osteochondroplasty for hip impingement: A finite element study [J].
Alonso-Rasgado, Teresa ;
Jimenez-Cruz, David ;
Bailey, Colin G. ;
Mandal, Parthasarathi ;
Board, Tim .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (12) :1999-2006
[7]   Fully Porous 3D Printed Titanium Femoral Stem to Reduce Stress-Shielding Following Total Hip Arthroplasty [J].
Arabnejad, Sajad ;
Johnston, Burnett ;
Tanzer, Michael ;
Pasini, Damiano .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2017, 35 (08) :1774-1783
[8]   Investigating stress shielding spanned by biomimetic polymer-composite vs. metallic hip stem: A computational study using mechano-biochemical model [J].
Avval, Pouria Tavakkoli ;
Samiezadeh, Saeid ;
Klika, Vaclav ;
Bougherara, Habiba .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2015, 41 :56-67
[9]   Design process of cementless femoral stem using a nonlinear three dimensional finite element analysis [J].
Baharuddin, Mohd Yusof ;
Salleh, Sh-Hussain ;
Zulkifly, Ahmad Hafiz ;
Lee, Muhammad Hisyam ;
Noor, Alias Mohd ;
Harris, Arief Ruhullah A. ;
Majid, Norazman Abdul ;
Abd Kader, Ab Saman .
BMC MUSCULOSKELETAL DISORDERS, 2014, 15
[10]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871