Performance testing and comparative study of natural radioactivity in soil samples using high purity germanium (HPGe) detector

被引:8
作者
Bangou, Christian [1 ,3 ]
Otoo, Francis [1 ,2 ]
Darko, Emmanuel Ofori [1 ,2 ]
机构
[1] Univ Ghana, Sch Nucl & Allied Sci, POB AE1,Atom Campus, Accra, Ghana
[2] Ghana Atom Energy Commiss, Radiat Protect Inst, POB LG 80, Accra, Ghana
[3] Autoritenatl Radioprotect & SureteNucl ARSN, POB 7044, Ouagadougou, Burkina Faso
关键词
NORM; Indicators; RPI; ARSN; Laboratory;
D O I
10.1016/j.mex.2021.101397
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
High Purity germanium (HPGe) detectors are found to be suitable for nuclear techniques for measuring radionuclides with very good energy resolution. Inter-comparison exercise is an important tool for external quality control that enables determination of the accuracy and uncertainty of detector measurement system. In this work, a comparative study of natural radioactivity in soil samples was conducted between the laboratory of Autorite nationale de Radioprotection et de Surete Nucleaire (ARSN), Burkina Faso and the laboratory of Radiation Protection Institute of Ghana Atomic Energy Commission (RPI-GAEC),Ghana to ascertain the reliability and accuracy of measurements made in Burkina Faso. For this purpose: Some replicate soil samples, assumed as proficiency test samples, were analyzed on both the High Purity Germanium detector of ARSN and RPI. The statistical performance indicators of z-score, precision, trueness and relative bias were used for the evaluation. The limit for acceptable precision and the maximum acceptable bias for all the radionuclides were set to 20% and 15% respectively. (C) 2021 The Authors. Published by Elsevier B.V.
引用
收藏
页数:12
相关论文
共 16 条
  • [1] Barnekow U, 2019, IAEA TECHNICAL REPOR, V486
  • [2] Central West Region, 2018, COMM DEV PLAN UNPUB
  • [3] Darko E.O, 2009, RADIAT PROT DOSIM
  • [4] el Samad O., 2018, J RADIAT NUCL APPL, V3, P9, DOI [10.18576/jrna/030102, DOI 10.18576/JRNA/030102]
  • [5] Detection efficiency of NaI(Tl) detector based on the fabricated calibration of HPGe detector
    El-Gamal, Hany
    Negm, Hani
    Hasabelnaby, Mohamed
    [J]. JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) : 360 - 366
  • [6] Faanu A, 2010, ENVIRON MONIT ASSESS
  • [7] Faanu A, 2011, ASSESSMENT PUBLIC EX, DOI [10. 16194/j.cnki.31- 1059/g4.2011.07.016, DOI 10.16194/J.CNKI.31-1059/G4.2011.07.016]
  • [8] Faanu A, 2016, NATURAL RADIOACTIVIT
  • [9] Faanu A., 2016, RAD PROTECT ENV, V39, P155, DOI 10.4103/0972-0464.194962
  • [10] IAEA, 1989, TECHNICAL REPORTS SE