Tribocorrosion of passive metals in the mixed lubrication regime: theoretical model and application to metal-on-metal artificial hip joints

被引:37
作者
Cao, Shoufan [1 ]
Maldonado, Sandra Guadalupe [1 ]
Mischler, Stefano [1 ]
机构
[1] Ecole Polytech Fed Lausanne, Tribol & Interfacial Chem Grp, CH-1015 Lausanne, Switzerland
关键词
Tribocorrosion modeling; Mixed lubrication; Wear; Passive metals; Artificial hip joint; EXSITU SURFACE-ANALYSIS; SIMULATED BODY-FLUIDS; FE-CR ALLOYS; IN-VITRO; SULFURIC-ACID; WEAR; CORROSION; IMPLANTS; BEHAVIOR; COCRMO;
D O I
10.1016/j.wear.2014.12.003
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Wear is a key problem limiting the lifetime and performance of total hip arthroplasty. The synovial fluid, acting as lubricant, can alleviate the wear of implants but also introduces corrosion. The interaction of mechanical wear and electrochemical corrosion, nominally tribocorrosion, has been proposed as one of the crucial degradation mechanisms of implants and recently saw significant progress in mechanistic understanding and modeling. This study presents a predictive wear model based on a mechanistic approach of tribocorrosion and lubrication and taking into account clinical relevant parameters such as normal load, velocity and clearance. Predicted wear rate corresponds well to the running-in wear rate from simulator experimental results and allows to identify the dominating wear mechanism and to evaluate the influence of several parameters. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:55 / 63
页数:9
相关论文
共 38 条
[1]  
Azzi M, 2011, TRIBOCORROSION OF PASSIVE METALS AND COATINGS, P222
[2]   NATURE OF THE PASSIVE FILM ON FE-CR ALLOYS AS STUDIED BY O-18 SECONDARY ION MASS-SPECTROMETRY - REDUCTION OF THE PRIOR FILM AND STABILITY TO EXSITU SURFACE-ANALYSIS [J].
BARDWELL, JA ;
SPROULE, GI ;
MITCHELL, DF ;
MACDOUGALL, B ;
GRAHAM, MJ .
JOURNAL OF THE CHEMICAL SOCIETY-FARADAY TRANSACTIONS, 1991, 87 (07) :1011-1019
[3]   EXSITU SURFACE-ANALYSIS OF PASSIVE FILMS ON FE-CR ALLOYS - WHEN IS IT VALID [J].
BARDWELL, JA ;
SPROULE, GI ;
GRAHAM, MJ .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1993, 140 (01) :50-53
[4]   The pathways of dynamic recrystallization in all-metal hip joints [J].
Büscher, R ;
Fischer, A .
WEAR, 2005, 259 :887-897
[5]   Influence of microstructure of HC CoCrMo biomedical alloys on the corrosion and wear behaviour in simulated body fluids [J].
Casaban Julian, Leandre ;
Igual Munoz, Anna .
TRIBOLOGY INTERNATIONAL, 2011, 44 (03) :318-329
[6]   Comparison of in vitro with in vivo characteristics of wear particles from metal-metal hip implants [J].
Catelas, I ;
Medley, JB ;
Campbell, PA ;
Huk, OL ;
Bobyn, JD .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2004, 70B (02) :167-178
[7]  
Chan FW, 1999, CLIN ORTHOP RELAT R, P10
[8]   Tribo-electrochemical characterization of metallic biomaterials for total joint replacement [J].
Diomidis, N. ;
Mischler, S. ;
More, N. S. ;
Roy, Manish .
ACTA BIOMATERIALIA, 2012, 8 (02) :852-859
[9]   Tribological principles in metal-on-metal hip joint design [J].
Dowson, D .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H2) :161-171
[10]   A hip joint simulator study of the performance of metal-on-metal joints - Part II: Design [J].
Dowson, D ;
Hardaker, C ;
Flett, M ;
Isaac, GH .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :124-130